Laser-Pulsed Atom Probe Analysis Condition 439
Figure 7. Thefractionofmultipledetectioneventsasafunctionofcharge-stateratio of tungsten ions [W4+/(W2++W3+)] plottedusing APTdataacquiredfrom(a) LEAP 3000X HR (3000PE and 3000BT test series), (b) LEAP 5000 XS (5000PEA and 5000PRR test series). Here A and D indicate that the variable was changed in ascending and descending sequence, respectively. The detailed experimental conditions are listed in Table 2.
observed in this work, a larger specimen radius can improve mass resolution. This difference is attributed to the different thermal diffusivity of these two materials, which is the intrinsic factor controlling the cooling rate. The impact of the laser repetition rate on the mass resolution is also closely related to the thermal diffusivity of the specimen. For materials with good thermal diffusivity, after the illumina- tion with the ultra-fast laser pulse, the specimen can cool down within 1 ns (Vurpillot et al., 2009), but for the samples of poor thermal diffusivity, the cooling time will be longer. In this study, we did not observe a strong dependence of mass resolution on pulse repetition rate. The highest laser pulse repetition rate applied is 1MHz, where the time interval between two pulses is 1 μs, indicating that the time needed by the specimen to cool back to the base temperature is <1 μs. Actually, the cooling down time of the specimen can be roughly estimated from the length of the peak tails in the correctedTOF spectrum, which is around 6 ns, much shorter than 1 μs. If the cooling time is longer than the pulse interval, the effects of the laser pulse will accumulate and the specimen will not cool back to the base temperature. Con- sequently, the mass resolution will increasingly degrade, which is not observed here. In addition, as specimen geometry can also influence the cooling time, the impact of laser pulse repetition rate on mass resolution is also affected by specimen geometry. In contrast to the mass resolution, the NSR is not so
sensitive to pulse energy and specimen base temperature. Even though different analysis parameters and specimens were used in the 18 measurements of the 3000PE and the 3000BT series, the resulting NSR values only show small variations (Fig. 3a, 3c), calculated to be 0.14±0.042. The specimen radius and the laser pulse repetition rate have influences on the NSR. When the specimen radius is large and the pulse repetition rate is low, the NSR will increase. There are several origins of the background noise. First, residual-gas atoms, or other types of atoms from the chamber or contamination adsorbed on the specimen
surface, can be field evaporated by the pulse or the intense DC electric field and contribute to the background. Figure 3 shows the average pressure in the analysis chamber of each measurement. Normally it is in the range of 10−9 Pa (10−11 torr). Although decreasing the specimen base temperature can lead to a drop in the pressure, due to the condensation of some gas species such as CO and CO2, the background level is only slightly affected. Second, the micro-channel plate (MCP) of the detector will also generate noise. Third, some heavy ionswhich are too slow to reach the detector within an individual detection window, as well as ions with substantial delays in evaporation and largely shif- ted TOF, will also be treated as background noise in the mass spectrum. In this study, the contributions from the third aspect are thought to be limited and the vast majority of noise is supposed to come from residual gas and the MCP. These background ions are normally evenly distributed in time. With a higher laser repetition rate or detection efficiency, the time needed to collect the same amount of signals is shorter. Therefore, the amount of noise is reduced. Larger specimen radii require higher standing voltages and the TOF of noise ions becomes shorter, hence, more noise ions can reach the detector, resulting in a higher level of background. On the other hand, a larger specimen radius means a larger surface area for residual-gas atom absorption. Besides, DC evaporationmay also have a contribution to the tremendous increase in the NSR of the 100 kHz test in the 5000PRRD series. When a low pulse repetition rate and a specimen with large radius are adopted, the temperature rise of the specimen is smaller, leading to an increase in the evaporation field and, as a result, DC evaporation could take place. The accuracy of the resulting chemical composition is linked to the fraction of multiple events. Generally, there are
two main mechanisms giving rise to the multiple events. The first one is termed as correlated field evaporation. When one atom is field evaporated, atoms in the vicinity sites are prone to field evaporate successively due to the increase in the
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80 |
Page 81 |
Page 82 |
Page 83 |
Page 84 |
Page 85 |
Page 86 |
Page 87 |
Page 88 |
Page 89 |
Page 90 |
Page 91 |
Page 92 |
Page 93 |
Page 94 |
Page 95 |
Page 96 |
Page 97 |
Page 98 |
Page 99 |
Page 100 |
Page 101 |
Page 102 |
Page 103 |
Page 104 |
Page 105 |
Page 106 |
Page 107 |
Page 108 |
Page 109 |
Page 110 |
Page 111 |
Page 112 |
Page 113 |
Page 114 |
Page 115 |
Page 116 |
Page 117 |
Page 118 |
Page 119 |
Page 120 |
Page 121 |
Page 122 |
Page 123 |
Page 124 |
Page 125 |
Page 126 |
Page 127 |
Page 128 |
Page 129 |
Page 130 |
Page 131 |
Page 132 |
Page 133 |
Page 134 |
Page 135 |
Page 136 |
Page 137 |
Page 138 |
Page 139 |
Page 140 |
Page 141 |
Page 142 |
Page 143 |
Page 144 |
Page 145 |
Page 146 |
Page 147 |
Page 148 |
Page 149 |
Page 150 |
Page 151 |
Page 152 |
Page 153 |
Page 154 |
Page 155 |
Page 156 |
Page 157 |
Page 158 |
Page 159 |
Page 160 |
Page 161 |
Page 162 |
Page 163 |
Page 164 |
Page 165 |
Page 166 |
Page 167 |
Page 168 |
Page 169 |
Page 170 |
Page 171 |
Page 172 |
Page 173 |
Page 174 |
Page 175 |
Page 176 |
Page 177 |
Page 178 |
Page 179 |
Page 180 |
Page 181 |
Page 182 |
Page 183 |
Page 184 |
Page 185 |
Page 186 |
Page 187 |
Page 188 |
Page 189 |
Page 190 |
Page 191 |
Page 192 |
Page 193 |
Page 194 |
Page 195 |
Page 196 |
Page 197 |
Page 198 |
Page 199 |
Page 200 |
Page 201 |
Page 202 |
Page 203 |
Page 204 |
Page 205 |
Page 206 |
Page 207 |
Page 208 |
Page 209 |
Page 210 |
Page 211 |
Page 212 |
Page 213 |
Page 214 |
Page 215 |
Page 216 |
Page 217 |
Page 218 |
Page 219 |
Page 220 |
Page 221 |
Page 222 |
Page 223 |
Page 224 |
Page 225 |
Page 226 |
Page 227 |
Page 228 |
Page 229 |
Page 230 |
Page 231 |
Page 232 |
Page 233 |
Page 234 |
Page 235 |
Page 236 |
Page 237 |
Page 238 |
Page 239 |
Page 240 |
Page 241 |
Page 242 |
Page 243 |
Page 244 |
Page 245 |
Page 246 |
Page 247 |
Page 248 |
Page 249 |
Page 250 |
Page 251 |
Page 252 |
Page 253 |
Page 254 |
Page 255 |
Page 256 |
Page 257 |
Page 258 |
Page 259 |
Page 260 |
Page 261 |
Page 262 |
Page 263 |
Page 264 |
Page 265 |
Page 266 |
Page 267 |
Page 268 |
Page 269 |
Page 270 |
Page 271 |
Page 272 |
Page 273 |
Page 274 |
Page 275 |
Page 276 |
Page 277 |
Page 278 |
Page 279 |
Page 280 |
Page 281 |
Page 282 |
Page 283 |
Page 284