Nanoscale Stoichiometric Analysis of a High-Temperature Superconductor 419 If all the peaks had the same shape and width, then the
composition would be independent of range width. The difference in composition measured with different range widths is a result of the asymmetry of the peaks, due to the thermal tails and the changes in peak widths corresponding to different ions. The peak width, in laser-assisted measure- ments, will depend on the field required to evaporate each ionic species from the sample apex (Gault et al., 2012). This effect causes the peaks to change across the reconstruction. The selected widths are therefore not tailored to a specific experimental condition, but are being applied to the average of a range conditions, i.e. the best performing ranging
approach is effectively the most robust across the changing field conditions. Because no information could be obtained for the evaporation field of every type of ion detected in the present mass spectra, as a compromise, full-width-half- maximum was selected for the following analysis. The mea- sured chemical composition from the whole atom probe run at the selected range width was Y7.4Ba16.2Cu23.8O52.4 at% and contained traces of ~0.17 at% Ce, whereby Ce was identified as the CeO2+ peak at 78 Da.
Thermal Tails
Samples which are thermally insulating when illuminated with a laser pulse will dissipate heat more slowly. As a result, the specimen remains at a higher temperature for a longer time, and hence the field evaporation process can continue for several nanoseconds after the application of each laser pulse (Bunton et al., 2007). This delayed evaporation man- ifests in the acquired time-of-flight spectrum as a “thermal tail” which extends from the main peak to longer times. When the time-of-flight spectrum is converted into a mass- to-charge spectrum, this effect is evident through tails which extend from the main peak to higher effective masses. During analysis of the spectrum, thermal tails can raise the effective background and hide smaller peaks, potentially causing the introduction of errors into any chemical quan- tification. Figure 4b clearly illustrates a reduction of thermal tail size in the whole spectrum when the laser energy is reduced. Figure 4c shows the small 65Cu2+ peak just after the large O2
+ peak, which at higher laser energies would be
hidden in the thermal tail. This demonstrates that, for the present samples, a reduction in laser energy from 0.6 to 0.05 nJ can lead to an effective improvement in mass resolution, and hence more confidence in the ranging of the mass spectrum. This effect can be explained as the reduction in laser
energy is expected to lead to a lower maximum temperature reached at the sample apex. The temperature change caused by the laser has been both experimentally estimated and modeled in previous studies on a variety of materials (Liu & Tsong, 1984), and is known to be dependent upon the laser energy and the individual specimen properties (Liu & Tsong, 1984; Kelly et al., 2014). If the maximum temperature is reduced, the time required for the surface to quench will also decrease. This, in turn, will cause ion evaporation to occur
within a narrower time range after each individual pulse, reducing the “thermal tail” effect in both the time-of-flight spectrum and therefore also in the final mass spectrum.
Effect of Laser Energy
could be correlated to its lower evaporation field when com- paredwith Cu orY(Gault et al., 2012). However, it is important to note that evaporation fields of elements have only been measured in their elemental state, they could be different in the case of compounds. Barium could be field-evaporating throughout the experiment in a way which is uncorrelated with the laser pulses. In this case the ions would leave the tip and reach the detector, however their time-of-flight could not be measured and they would therefore be unidentifiable, hidden in the mass spectrum background noise. This sugges- tion is further supported by the evidence presented in Figure 5b, which proves that at higher laser energies (which means lower applied voltage, therefore lower field) more Ba is detected. Further evidence supporting this theory is provided by the
The effect of altering laser energy on composition is shown in Figure 5a. Each data point represents a 600,000 ion sec- tions of the 17 million-ion data set. The entire Ba content of the analyzed volume was detected in the form of Ba2+ ions at 65–69 Da, which was expected based on the post-ionization theory developed by Haydock and Kingham (Kingham, 1982). However, this also means that the overlap with the thermal tail of the Cu+ ions at 63 and 65Da (Fig. 5b) can cause an overestimate of the Ba content of the sample. The compositions given in Figure 5c were therefore back- ground corrected by fitting an exponentially decaying func- tion to the thermal tails and subtracting them using a custom Matlab code developed by London et al. (2015). London et al. (2015) developed a model for fitting the peaks in their mass spectrum, by using an exponentially corrected Gaussian function for the peak shape, with an added exponentially decaying function on one side as the thermal tail, as descri- bed in further depth in their paper. At a laser energy below 0.25 nJ, the overall composition of the sample is approxi- mately constant, but a significant Ba deficiency is seen at every laser energy, as seen in Figure 5b. One suggested reason for the measured Ba deficiency
rise in background rate with decreasing laser energy. The background rate was measured in a region where no peaks or thermal tails were present, between 4–11 Da. At low laser energy (between 0.05 and0.15 nJ) the background rate was 8.1×10−5/ amu, whereas at high laser energy (between 0.45 and 0.6 nJ) it was 7.6×10−5/amu, when measured on a data set divided into bins of 0.05 Da. The background rate showed an effective increase by 6% at lower laser energy, which can be indicative of off-pulse evaporation by one or more chemical species.
Event Multiplicity
Subsequent to each laser (or high voltage) pulse, the micro- scope acquires signals fromthe detector for a duration dictated by the pulses repetition rate, in what is referred to as an event. During an event, if two or more ions are detected, then all ions
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80 |
Page 81 |
Page 82 |
Page 83 |
Page 84 |
Page 85 |
Page 86 |
Page 87 |
Page 88 |
Page 89 |
Page 90 |
Page 91 |
Page 92 |
Page 93 |
Page 94 |
Page 95 |
Page 96 |
Page 97 |
Page 98 |
Page 99 |
Page 100 |
Page 101 |
Page 102 |
Page 103 |
Page 104 |
Page 105 |
Page 106 |
Page 107 |
Page 108 |
Page 109 |
Page 110 |
Page 111 |
Page 112 |
Page 113 |
Page 114 |
Page 115 |
Page 116 |
Page 117 |
Page 118 |
Page 119 |
Page 120 |
Page 121 |
Page 122 |
Page 123 |
Page 124 |
Page 125 |
Page 126 |
Page 127 |
Page 128 |
Page 129 |
Page 130 |
Page 131 |
Page 132 |
Page 133 |
Page 134 |
Page 135 |
Page 136 |
Page 137 |
Page 138 |
Page 139 |
Page 140 |
Page 141 |
Page 142 |
Page 143 |
Page 144 |
Page 145 |
Page 146 |
Page 147 |
Page 148 |
Page 149 |
Page 150 |
Page 151 |
Page 152 |
Page 153 |
Page 154 |
Page 155 |
Page 156 |
Page 157 |
Page 158 |
Page 159 |
Page 160 |
Page 161 |
Page 162 |
Page 163 |
Page 164 |
Page 165 |
Page 166 |
Page 167 |
Page 168 |
Page 169 |
Page 170 |
Page 171 |
Page 172 |
Page 173 |
Page 174 |
Page 175 |
Page 176 |
Page 177 |
Page 178 |
Page 179 |
Page 180 |
Page 181 |
Page 182 |
Page 183 |
Page 184 |
Page 185 |
Page 186 |
Page 187 |
Page 188 |
Page 189 |
Page 190 |
Page 191 |
Page 192 |
Page 193 |
Page 194 |
Page 195 |
Page 196 |
Page 197 |
Page 198 |
Page 199 |
Page 200 |
Page 201 |
Page 202 |
Page 203 |
Page 204 |
Page 205 |
Page 206 |
Page 207 |
Page 208 |
Page 209 |
Page 210 |
Page 211 |
Page 212 |
Page 213 |
Page 214 |
Page 215 |
Page 216 |
Page 217 |
Page 218 |
Page 219 |
Page 220 |
Page 221 |
Page 222 |
Page 223 |
Page 224 |
Page 225 |
Page 226 |
Page 227 |
Page 228 |
Page 229 |
Page 230 |
Page 231 |
Page 232 |
Page 233 |
Page 234 |
Page 235 |
Page 236 |
Page 237 |
Page 238 |
Page 239 |
Page 240 |
Page 241 |
Page 242 |
Page 243 |
Page 244 |
Page 245 |
Page 246 |
Page 247 |
Page 248 |
Page 249 |
Page 250 |
Page 251 |
Page 252 |
Page 253 |
Page 254 |
Page 255 |
Page 256 |
Page 257 |
Page 258 |
Page 259 |
Page 260 |
Page 261 |
Page 262 |
Page 263 |
Page 264 |
Page 265 |
Page 266 |
Page 267 |
Page 268 |
Page 269 |
Page 270 |
Page 271 |
Page 272 |
Page 273 |
Page 274 |
Page 275 |
Page 276 |
Page 277 |
Page 278 |
Page 279 |
Page 280 |
Page 281 |
Page 282 |
Page 283 |
Page 284