1086 Dai Tang et al.
Figure 6. Measured effect of beam current on achievable differ- ence in brightness of backscattered electron (BSE) image between steel matrix and CaO–Al2O3–MgO inclusion (for maximum contrast setting); ASPEX at 10 kV.
on the accelerating voltage through its effect on the size of the interaction volume. The simulated line scans in Figure 5 show the broader transition area between the steel matrix brightness and inclusion brightness at the higher accelerating voltage—as expected from the larger interaction volume. Spatial resolution also depends on how narrowly the electron beam is focused. Beam diameter is affected by changing the “spot size” (which also directly affects beamcurrent and hence signal intensity, as discussed above). The spatial resolution was manually measured (via line scans through the centers of inclusions; see Fig. 7a) as the distance over which the brightness changed from 25 to 75% of the difference between an inclusion and the steel matrix. A disadvantage of performing resolution measurements on actual inclusions is the unknown shape and size of the inclusions below the polished surface, which would affect the resolution measurement somewhat (as illustrated by the simulations shown in Fig. 5). However, the advantage is that steel samples containing oxide inclusions are readily avail- able, and such measurements are directly applicable to the practical purpose of inclusion detection and analysis. The measured spatial resolution for different beam
currents is given in Figures 7b and 7c, for the two instru- ments at 10 and 20 kV. In general, the field-emission instrument has a better resolution than the tungsten fila- ment instrument as expected. For both instruments, spatial resolution is worse at the higher accelerating voltage (20 kV) (reflecting the larger interaction volume). At 20 kV, spatial resolution is nearly independent of the beam current. At 10kV, spatial resolution is better at smaller beam currents.
BSE Images: False Positives and Negatives in Inclusion Detection False positives represent erroneous detection of noninclu- sion features (Type I errors); false negatives occur when target inclusions are not detected (Type II errors). False
Figure 7. Effect of accelerating voltage and beam current (chan- ged by adjusting spot size) on spatial resolution for backscattered electron (BSE) imaging of CaO–Al2O3–MgO inclusions, at 10 and 20 kV. a: Schematic of the procedure to measure spatial resolution; (b) measured spatial resolution (ASPEX); (c) measured spatial resolution (XL30).
tial resolution) would influence the detection of inclusions through causing false positives or false negatives. This effect
positives would increase analysis time—since incorrectly detected “features” would be analyzed by EDX; false nega- tives would cause the true inclusion concentration to be under-analyzed. The quality of the BSE image (noise, contrast, and spa-
was simulated (using MATLAB), based on BSE images of five inclusions (with apparent sizes 2–4 μm2) exposed on the surface of steel (image size 512 ×512 pixels, pixel size 0.3 ×0.3 μm; see Fig. 8). For different dwell times, noise was added based on measurements for the ASPEX instrument (Fig. 3a). “Features” were detected when pixels were darker than the threshold (filtering out features of two pixels or smaller, as is typically done during actual analyses). The results show that dwell times smaller than 4 µs (Fig. 9a; threshold of 160) caused many false positives. Increasing the threshold level (as would be necessary to detect brighter inclusions such as sulfides) also increased false positives (Fig. 9b). Practical measurements demonstrated the effect of
false positives and false negatives: analyses were repeated on the same area of a single steel sample for different thresholds (Table 2), with everything else being equal (ASPEX instrument, 20kV accelerating voltage and spot size 40%; contrast difference between Fe and Al increased to 130 to allow more thresholds to be tested). Table 2 lists the number
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80 |
Page 81 |
Page 82 |
Page 83 |
Page 84 |
Page 85 |
Page 86 |
Page 87 |
Page 88 |
Page 89 |
Page 90 |
Page 91 |
Page 92 |
Page 93 |
Page 94 |
Page 95 |
Page 96 |
Page 97 |
Page 98 |
Page 99 |
Page 100 |
Page 101 |
Page 102 |
Page 103 |
Page 104 |
Page 105 |
Page 106 |
Page 107 |
Page 108 |
Page 109 |
Page 110 |
Page 111 |
Page 112 |
Page 113 |
Page 114 |
Page 115 |
Page 116 |
Page 117 |
Page 118 |
Page 119 |
Page 120 |
Page 121 |
Page 122 |
Page 123 |
Page 124 |
Page 125 |
Page 126 |
Page 127 |
Page 128 |
Page 129 |
Page 130 |
Page 131 |
Page 132 |
Page 133 |
Page 134 |
Page 135 |
Page 136 |
Page 137 |
Page 138 |
Page 139 |
Page 140 |
Page 141 |
Page 142 |
Page 143 |
Page 144 |
Page 145 |
Page 146 |
Page 147 |
Page 148 |
Page 149 |
Page 150 |
Page 151 |
Page 152 |
Page 153 |
Page 154 |
Page 155 |
Page 156 |
Page 157 |
Page 158 |
Page 159 |
Page 160 |
Page 161 |
Page 162 |
Page 163 |
Page 164 |
Page 165 |
Page 166 |
Page 167 |
Page 168 |
Page 169 |
Page 170 |
Page 171 |
Page 172 |
Page 173 |
Page 174 |
Page 175 |
Page 176 |
Page 177 |
Page 178 |
Page 179 |
Page 180