search.noResults

search.searching

note.createNoteMessage

search.noResults

search.searching

orderForm.title

orderForm.productCode
orderForm.description
orderForm.quantity
orderForm.itemPrice
orderForm.price
orderForm.totalPrice
orderForm.deliveryDetails.billingAddress
orderForm.deliveryDetails.deliveryAddress
orderForm.noItems
1084 Dai Tang et al.


Figure 2. Measured beam current and stage current (electron beam incident on Fe matrix), at 10 and 20kV accelerating voltages, and different spot sizes, for a tungsten filament instrument (labeled “ASPEX”) and a Schottky field-emission instrument (XL30).


inclusions. The noise-to-signal ratio (N/S) in the BSE image is inversely proportional to the square root of the total number of BSE detected (n):


N S = ffiffiffinp


n = 1 n=CηIbτ; pn : ffiffiffi (1)


The backscattered signal detected from a given beam location is given by:


(2)


where C is a constant (determined by the size, position and gain of the BSE detector); η the BSE yield; Ib the beam cur- rent; and τ the dwell time per pixel. The beam current (Ib) was measured with a Faraday cup


and the stage current (Is) was measured with the beam incident on polished steel. Measured beam and stage currents are presented in Figure 2, showing the effect of spot size and accelerating voltage for the tungsten filament ASPEX instrument and the Schottky field-emission instru- ment XL30. (“Spot size” is a relative term used by the parti- cular instrument supplier; for the ASPEX instrument spot size is given as a percentage, and for the XL30 instrument as a numerical value from 2 to 7.) The general trends are that the beamcurrent is larger for larger spot sizes, and also depends on accelerating voltage. For the tungsten filament ASPEX, the beam current at 10kV (for the generally used 40% spot size) is ~30% of the beam current at 20kV (and the same spot size). For the field emission XL30, the difference is smaller: the beam current at 10kV (spot size 5; typically used in automated inclusion microanalyses) is ~60% of the beam current at 20kV (spot size 5). The smaller beam current at 10kV would require a


longer pixel dwell time to achieve the same noise-to-signal ratio as at 20 kV. The effect of pixel dwell time on the noise- to-signal ratio was measured as the standard deviation of BSE line scans on steel, setting the same brightness and


Figure 3. Effect of pixel dwell time on the backscattered electron (BSE) image noise-to-signal ratios for imaging at 10 and 20kV for (a) ASPEX instrument (spot size 40%) and (b) XL30 instrument (spot size 5); the effect of accelerating voltage is to change the beam current and BSE detector gain, as shown by normalizing the noise-to-signal ratio with respect to these two factors [(c) and (d); beam current Ib and relative detector gain G are listed in Table 1].


contrast levels. BSE contrast was set using Fe and Al as standards, adjusting the brightness difference (on the dimen- sionless 0–255 brightness scale reported by the instruments) to 100. Measured values (Figs. 3a and 3b) show that the noise- to-signal ratio is approximately proportional to the reciprocal of the square root of pixel dwell time, as expected. For mea- surements on the XL30 (INCA) instrument, extrapolation of the measured values to large dwell times—zero values of 1/τ0.5—indicate non-zero noise-to-signal ratios; this residual noisewas likely caused by some roughness on the surface of the steel sample used for these measurements. For a given pixel dwell time, noise-to-signal ratios were


larger at 10kV than at 20 kV, reflecting the lower beam current at 10 kV, and also the effect of the accelerating voltage on the BSE detector gain. For the solid-state BSE detectors used, the detector gain (G) is approximately proportional to the energy of BSE: that is, the detector gain is expected to be twice as large for imaging at 20kV than at 10 kV, at the same beam current (Goldstein et al., 2003). The beam current Ib (for one spot size as example) and relative BSE detector gain G (20kV relative to 10kV) are listed in Table 1. For monitoring possible changes in instrument condition, it was convenient to measure the stage current with the beam incident perpendicularly on polished steel; the measured stage current was converted to beam current using the ratios in Table 1. (The beam currents at corresponding spot sizes shown in Fig. 2 and Table 1 differ, because of such changes in instrument conditions.) If differences in beam current (Ib) and detector gain (G)


fully explained the effect of accelerating voltage on BSE noise, the noise-to-signal ratio would be inversely propor- tional to (GIb)0.5, for a given pixel dwell time. This was tested by plotting the product of the noise-to-signal ratio and


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84  |  Page 85  |  Page 86  |  Page 87  |  Page 88  |  Page 89  |  Page 90  |  Page 91  |  Page 92  |  Page 93  |  Page 94  |  Page 95  |  Page 96  |  Page 97  |  Page 98  |  Page 99  |  Page 100  |  Page 101  |  Page 102  |  Page 103  |  Page 104  |  Page 105  |  Page 106  |  Page 107  |  Page 108  |  Page 109  |  Page 110  |  Page 111  |  Page 112  |  Page 113  |  Page 114  |  Page 115  |  Page 116  |  Page 117  |  Page 118  |  Page 119  |  Page 120  |  Page 121  |  Page 122  |  Page 123  |  Page 124  |  Page 125  |  Page 126  |  Page 127  |  Page 128  |  Page 129  |  Page 130  |  Page 131  |  Page 132  |  Page 133  |  Page 134  |  Page 135  |  Page 136  |  Page 137  |  Page 138  |  Page 139  |  Page 140  |  Page 141  |  Page 142  |  Page 143  |  Page 144  |  Page 145  |  Page 146  |  Page 147  |  Page 148  |  Page 149  |  Page 150  |  Page 151  |  Page 152  |  Page 153  |  Page 154  |  Page 155  |  Page 156  |  Page 157  |  Page 158  |  Page 159  |  Page 160  |  Page 161  |  Page 162  |  Page 163  |  Page 164  |  Page 165  |  Page 166  |  Page 167  |  Page 168  |  Page 169  |  Page 170  |  Page 171  |  Page 172  |  Page 173  |  Page 174  |  Page 175  |  Page 176  |  Page 177  |  Page 178  |  Page 179  |  Page 180