EQUIPMENT & ACCESSORIES CATALOG EDITION IV COOLING STAGES
Why cool?
Low vacuum (LV) or variable pressure (VP) modes are now standard on most scanning electron microscopes (SEMs). For this reason it has become important to control water evaporation from wet specimens. Cooling such specimens reduces the loss of water by evaporation, or - depending on chamber pressure - can prevent it altogether.
Saturated vapour pressure of water decreases considerably with temperature. At room temperature, water will very quickly evaporate - causing considerable damage to specimen composition and ultra-structure. This is due to high forces of surface tension at the drying front as it passing though the specimen. In most biological systems this will result in distortion and collapse of membranes and other structures.
At 300Pa, the specimen temperature needs to be less than -9.5°C, and at 85Pa less than -25°C to stop water evaporation. Therefore, by cooling a specimen to -25°C, chamber pressures up to 85Pa can be used with little or no water loss by evaporation. In this way, changes in specimen structure can be minimised. In addition, being able to operate at higher vacuum gives a better signal-to- noise ratio and clearer images.
zzzThe EMS Coolstage for SEM, LV or VP
OVERVIEW
The Coolstage is a Peltier-driven SEM cooling stage for scanning electron microscopy (SEM), low vacuum (LV) or variable pressure (VP) applications. The stage can be cooled to sub- zero temperatures for specimens that may be sensitive at ambient temperature, subject to beam damage, or may otherwise 'sublime' (lose water) at ambient temperatures.
There are three versions of coolstage - Standard, Enhanced, and Ultra - to cover differing specimen requirements.
FEATURES Standard Coolstage
z Temperature range -30°C to +50°C at 300Pa
z Self contained cooling - no additional external cooling water needed
z Temperature accuracy +/- 1.5°C or 2% - whichever is greater
z Minimal image drift
z Cooling and heating rates of up to 30°C per minute
z Keypad control - with simultaneous display of actual and target temperature
z Supplied with SEM chamber port feed- though - specify when ordering
z One-year warranty
Enhanced Coolstage z Temperature range -30°C to +160°C at 300Pa
z All other specifications as per Standard Coolstage
Ultra Coolstage
z Temperature range -50°C to +50°C at 300Pa
z All other specifications as per Standard Coolstage
PRODUCT DESCRIPTION
The Coolstage is a temperature-controlled specimen stage that can be fitted to any low vacuum (LV) or variable pressure (VP) scanning electron microscope (SEM).
The Standard Coolstage consists of a single stage Peltier device, onto which a thermally isolated specimen holder and dual temperature sensor is mounted. The Coolstage assembly is mounted onto the SEM stage using an adaptor plate specific to the microscope. Cooling pipes and electrical wires connect to the SEM feedthrough flange. External components are a recirculating water chiller and power supply case, and a compact keypad for digital temperature readout and control.
Compact, efficient cooling and temperature control
The temperature range of the Standard Coolstage is -30°C to +50°C at 300Pa. The specimen holder is water-cooled using a small, self-contained closed loop recirculating chiller that is normally positioned approx 2m from the SEM. A microprocessor controls and monitors the temperature of the cold stage. A small keypad is used to set the required temperature and display target and current temperatures.
2 Ultra Coolstage. Range: -50ºC to +50ºC at 300Pa Standard Coolstage. Range: -30°C to +50°C at 300Pa
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76