20 ANALYTICAL AND LABORATORY EQUIPMENT
Monochromator technology with filter-like performance
Tobias Pusterla reveals the next generation of monochromators and explores the advantages they bring to end users.
M
icroplate reader detection is based either on filters or
grating-based monochromators. Typically, filters grant higher detection sensitivity because of increased transmission of light, whereas monochromators offer higher flexibility as wavelength selection can be adjusted for different fluorophores.
Fig. 1 Schematic representation of a double monochromator system used in microplate readers.
Double monochromators represented the state-of-the-art for flexible microplate reader detection until 2013 when BMG Labtech launched the Clariostar reader with LVF monochromator technology. Despite its name, the LVF monochromator uses neither a concave grating nor does it mechanically break up the light. Instead, it uses special adjustable filters, so-called linear variable filters (LVF), to select wavelengths. LVFs are defined as filters whose spectral properties vary linearly from one end of the filter to the other. Te units consist of two LVFs that form the rising and falling edge of a filter. By moving the LVF slides relative to each other, the system separates light into distinct wavelengths and continuously adjustable bandwidths. Te Clariostar reader consists of two LVF monochromators, one for
Fig. 2. Schematic representation of the LVF monochromator technology used in the Clariostar microplate reader.
excitation and one for emission. In addition, a linear variable dichroic mirror slide is used to separate the excitation from the emission light and further reduce undesired wavelengths
LVF monochromators transmit more light than grating-based monochromators since, like filters, LVF slides allow for more than 80% light transmission. Tis enables filter-like sensitivity.
In grating-based solutions, bandwidths are limited to a
maximum of 20-25 nm. Te Clariostar allows continuously adjustable bandwidths up to 100nm. Te broader the bandwidth, the more light and the higher the sensitivity.
Although some microplate readers include both filters and monochromators, these cannot be combined as each uses a separate light path. In the Clariostar, LVF monochromators and filters share the same light path as the slides holding the LVFs also incorporate slots for fixed filters. LVF monochromators and filters can hence be combined in one measurement.
Tese LVF monochromators have several advantages over grating-based systems.
For more information ✔ at
www.scientistlive.com/eurolab
Tobias Pusterla is with BMG Labtech in Germany.
www.bmglabtech.com
www.scientistlive.com
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80 |
Page 81 |
Page 82 |
Page 83 |
Page 84 |
Page 85 |
Page 86 |
Page 87 |
Page 88 |
Page 89 |
Page 90 |
Page 91 |
Page 92 |
Page 93 |
Page 94 |
Page 95 |
Page 96 |
Page 97 |
Page 98 |
Page 99 |
Page 100 |
Page 101 |
Page 102 |
Page 103 |
Page 104 |
Page 105 |
Page 106 |
Page 107 |
Page 108 |
Page 109 |
Page 110 |
Page 111 |
Page 112 |
Page 113 |
Page 114 |
Page 115 |
Page 116 |
Page 117 |
Page 118 |
Page 119 |
Page 120