focus on Mass Spectrometry & Simon Hudson, HFL Sport Science Ltd, Fordham, Cambs
There have been numerous reports in the news over the last few years of new drugs of abuse becoming available on the UK market. These have included mephedrone, benzylpiperazine and the so-called synthetic cannabinoids found in ‘Spice’ type herbal high products. These have all led to revisions of the misuse of drugs act in an attempt to control their distribution and use. A slightly older problem is the importation of anabolic steroids. In the year 2008/9 there were 802 seizures of anabolic steroids in the UK [1]. Of these, 259 were by the Borders Agency and the remainder by the Police forces. The misuse of drugs act now includes most known anabolic steroids and a generic wording to include most ‘designer’ anabolic steroids.
In contrast to previous decades where drugs of abuse were produced domestically or smuggled in to the country, many now arrive in bogusly labeled consignments destined for private addresses in the UK. The UK borders agency frequently intercepts shipments of significant size containing unknown white powders. These often do not contain the ‘traditional’ drugs of abuse that are readily identified by technology available at ports of entry to the UK. In such situations, seizures are sent for further analysis at laboratories, typically utilising GCMS, in an attempt to make an identification. This approach is often successful where the drug is amenable to GCMS analysis and a reference mass spectrum exists in a library or publication. There are occasions however where the drug is not suitable for analysis by GCMS or no reference spectrum is available. In such circumstances the use of high resolution LCMS can aid in the identification process.
In recent months, we have been forwarded samples of powders and oils seized by the UK borders agency that could not be identified by GCMS and mass spectral comparison.
Most have been from shipments of several kilograms of material. We have used LCMS consisting of a ThermoFisher Accela UPLC sytem interfaced to an LTQ Orbitrap to identify several anabolic steroids, all of which are controlled under the misuse of drugs act. We have also identified other steroids using GCMS and in-house databases.
The powders received for analysis were found to be anabolic steroids, all ofwhich are listed as Class C drugs under the Misuse of Drugs Act. The precise origin of these powders was unknown, but it is likely that they came from China. They were all destined for private addresses in the UK.
The use of high resolution LCMS proved to be very useful in the identification of these compounds and was certainly complementary to GCMS in that some compounds, for example stanozolol and trenbolone were best analysed by LCMS whilst others were detected only by GCMS. Where reference spectra did not exist, as with drostanolone propionate, the combination of both techniques led to identification based on the elemental composition derived from accurate mass measurement and the similarity of the GCMS derived EI mass spectrum to a reference spectrum of another compound.
The identification of customs seizures of this type can often be limited by databases and this is where the use of accurate mass can assist in giving additional information. As was shown with this work, accurate mass determination, followed by the assignment of probable elemental composition, often gives a good starting point in the identification process.
Experimental
Sample Preparation Solutions of the materials were prepared in methanol.
For LCMS analysis, a sub aliquot of the solution was taken and diluted with water to produce a sample for injection with a concentration of 100µg/ml.
For GCMS analysis, 10µl of the initial solution was evaporated to dryness in a GC vial. To the residue, 200µl of derivatising solution (MSTFA:ethanethiol:ammonium iodide 1ml:6µl:3mg) was added. The vial was capped and heated for 2 hours at 80˚C before injecting on a GCMS system.
LCMS
The samples were analysed on a Thermo Fisher Accela UPLC system interfaced to a Thermo Fisher LTQ Orbitrap Discovery.
The chromatographic separation was performed using a Waters Symmetry C8 3.5µm x 2mm x 100mm column linked to a Phenomenex Security Guard C5 (4x2mm) pre-column.
The Orbitrap was calibrated for positive ion analysis using the standard Thermo Fisher Orbitrap calibration solutions and procedures. The instrument was operated in positive ion mode with a resolution setting of 30,000 with an internal lock mass of 113.03455 derived from uracil in the mobile phase. Data was acquired in full scan mode over a mass range of 100-650 Da.
10µl of sample was injected for analysis. GCMS
The samples were analysed on an Agilent 6890 GC interfaced to an Agilent 5973 MSD. The chromatographic separation was performed on a 25 meter x 0.25mm i.d. x 0.25µm film BPX5 capillary column from SGE.
The mass spectrometer was tuned and calibrated using perfluorotributylamine and data was acquired in full scan mode over a mass range of 50 to 700Da.
1µl of sample was injected on the system
Results Trenbolone enanthate
Spectroscopy
The Identification of Illegal Anabolic Steroids in Customs Seizures by GCMS and High Resolution LCMS
Monoisotopic [M+H]+ = 383.2580
Figure 1. Accurate mass LCMS of trenbolone enanthate showing TIC and mass spectrum.
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60