search.noResults

search.searching

saml.title
dataCollection.invalidEmail
note.createNoteMessage

search.noResults

search.searching

orderForm.title

orderForm.productCode
orderForm.description
orderForm.quantity
orderForm.itemPrice
orderForm.price
orderForm.totalPrice
orderForm.deliveryDetails.billingAddress
orderForm.deliveryDetails.deliveryAddress
orderForm.noItems
Air Monitoring 37 Table 2. Gradient conditions employed in the separation of 18 PFAS compounds


Time (mins) 0


0.1 8.5


10.5 10.6


A SCIEX QTRAP®


% B 5


20 95 95 5


ESI mode with a source temperature of 450°C and an IonsprayTM


6500+ system was used to detect the compounds of interest, running in negative source voltage of -4500 V.


Figure 1 shows an example chromatogram obtained by injecting a 1 µL sample at a concentration of 1000 ng/L [19].


When developing the method it was noted that there was a signifi cant amount of background contamination, which subsequently led to an intensive investigation to systematically determine the source and levels of contamination. A variety of possible sources were identifi ed. Initially, the sample


11,12,13


1.00e4 2.00e4 3.00e4 4.00e4 5.00e4 6.00e4 7.00e4 8.00e4 9.00e4 1.00e5


0 1.0 2.0 3.0 4.0 17,18


Figure 2 Extraction of trace residual PFAS present in a variety of common laboratory consuables


Figure 2. Extraction of trace residual PFAS present in a variety of common laboratory consuables


will also result in chromatographic peaks, since it will build up at the front of the column during the low elutropic phase of the gradient elution, and when the gradient reaches a critical concentration, the PFAS components will start to elute, resulting in inaccurate concentration determination of the individual PFAS.


8 9,10


19,20,21 23


2,3 1 4,5 6 7


14 15,16


22 24 25 5.0 Time, min 6.0 7.0 8.0 9.0 10.0


Figure 1 Example chromatogram obtain with 18 PFAS compounds and selected stable isotope labelled internal standards


Figure 1. Example chromatogram obtain with 18 PFAS compounds and selected stable isotope labelled internal standards


vials and caps were the focus, with a simple solvent extraction being applied to determine if there were any extractable PFAS present. Each possible source and alternatives were extracted using 300 µL of methanol. The resulting solution was injected as a 10 µL aliquot. For the early eluting compounds this resulted in some poor peak shapes which is the combined effect of injection of a larger volume of strongly eluotropic diluent, weakly retained analytes being more susceptible to system dispersion, as well as being present at low concentration levels and close to the instrument’s limit of detection. Due to the use of the qualifying extracted ion chromatograms, this facilitated the identifi cation/qualifi cation of contaminants without having to introduce another possible contamination source e.g. from a pre-concentration step and the use of a blow down evaporator.


The data, given in Table 2, is not quantitative, but the peak areas do indicate that there are detectable levels of PFAS. In this case two septa, both based on a polyimide silicone material compared to the polypropylene material used in the other septa, were identifi ed as having ADONA (3H-perfl uoro-3-[(3-methoxy-propoxy)propanoic acid]) present at detectable levels.


Table 3. Extraction of trace residuals present in caps, septum or vials demonstrates that some vial caps do contain PFAS components


1000 2000 3000 4000 5000 6000 7000 8000 9000 10000 11000


0


Vial Manufacturer Vial material


Cap Manufacturer Septum material ADONA


Other PFAS (18 cpds)


Blank run N/A N/A N/A N/A û û


1 2


PP PP 1


2


500 1000 1500 2000 2500 3000 3500 4000 4500


0 7.0


1000 2000 3000 4000 5000 6000 7000 8000


0


PP PP PI/Si PI/Si PP û û ü14930 ü2341 û û û


7.0


1000 2000 3000 4000 5000 6000 7000 8000 9000


0 6.6 6.8 7.2 7.4 7.6 7.8 8.0 Time, min û 7.0 7.2 7.4 Time, min 7.6 û û 7.8 8.0


1000 2000 3000 4000 5000 6000 7000 8000 9000


0 6.6 6.8 7.0 7.2 8.2 8.4 8.6 8.8 7.2 7.0 7.2 7.4 7.6 7.8 Time, min 3 7.4 7.6 7.8 Time, min 8.0 3 8.2


PP 3


8.4 1 3


500 1000 1500 2000 2500 3000 3500 4000 4500


0 7.0


PP PP 1


1000 2000 3000 4000 5000 6000 7000 8000


0 7.0 7.2 7.4 7.6 7.8 8.0 Time, min Background PFAS 7.2 7.4 7.6 8.0 8.2 8.4


1000 2000 3000 4000 5000 6000 7000 8000 9000 10000 11000


0 7.0 7.2 7.4 7.6 Time, min 7.8


The approach to help alleviate this is to use a delay column which is placed in line before the injector, thus any PFAS components that are in the system will build up on the delay column preferentially before eluting onto the main analytical system. Once the gradient composition reaches a critical amount the PFAS components that have been retained on the delay column will be eluted onto the analytical column. The delay column is chosen so that it will introduce a retention time difference between the system derived PFAS and sample derived PFAS components. Figure 3 shows an example chromatogram; in this scenario the system is contaminated with a range of PFAS. It can be seen in the chromatogram that there are two peaks present. The fi rst peak relates to the PFAS in the sample and the second broader peak relates to the PFAS in the system. The poor shape of the chromatography is caused by the PFAS reagent going through two columns and also the build-up effect that the PFAS has as more solvent is passed through the column. The separation of the system PFAS components and the sample derived components means that accurate quantifi cation of analytes within the sample becomes feasible. However, it should be stated that this does not preclude inaccurate quantifi cation of the PFAS levels due to other sources of contamination within the sample itself, and as has already been shown, PFAS is ubiquitous within a laboratory environment. If this approach was not employed, then the system PFAS would elute at the same time as the sample PFAS, due to the nature of the focussing effect when using gradient chromatography. This would obviously result in the inaccurate quantifi cation of the amount of the individual analytes in the sample. As mentioned previously, the ubiquitous nature of PFAS means that system contamination can come from a variety of sources from the solvents to the tubing, whereas sample contamination can come from a myriad of sources contained within workfl ow materials & consumables, sample prep workfl ows etc.


Background PFAS


0e0 1e4 2e4 3e4 4e4


8.0 8.2 8.4 Background PFAS


0e0 1e4 2e4 3e4 4e4


7.8 Time, min 8.0 8.2 8.4 Background PFAS 8.2 8.4 8.6 8.8


0e0 1e4 2e4 3e4 4e4


0e0 1e4 2e4 3e4 4e4


0 7.4 Time, min 7.6 7.8 8.0


Blank run No delay column


0.5 ng/mL PFAS sample No delay column


Blank run Delay column installed


0.5 ng/mL PFAS sample Delay column installed


2 4


PFOA system background


PFOA from system co-elutes with sample PFOA


tR of Sample PFOA


Using delay column, system PFOA is delayed


Sample PFOA


System PFOA


6 Time, min 8


System PFOA is separated from sample PFOA


10


10000 15000 20000 25000 30000 35000


5000 0


Nitrile Glove Sticky tape White tape Laboratory Tissue Laboratory Tissue 2


Additional to the testing of the vials and caps, the analysis also looked at other possible sources of contamination as highlighted in Figure 2 [20]. The extraction process employed was also qualitative for this set of experiments and involved taking a small proportion of the sample and inserting into a test tube, adding 2 mL of methanol and shaking for a short period of time and then injecting 10 µL of the resulting solution into the LC-MS using the method specifi ed previously. It can be seen that there are a range of PFAS compounds that have been identifi ed. The profi les are different for the different sample types. It is interesting to note that the two sources of paper tissue that were being analysed were the same brand but bought on different dates, suggesting that the manufacturing process may involve variable amounts of PFAS material that is added due to the recycling of the paper, or the possibility of the production line having contamination issues.


ls present in caps, septum or vials demonstrates that some vials do contain PFAS components


As well as looking for possible contamination within general laboratory and its consumables, PFAS can also be found in solvents and in tubing associated with the instrumentation [21]. PFAS originating from the mobile phases or components on the HPLC system prior to the autosampler


Figure 3. Examples of how the PFAS delay column effectively separates system PFAS from sample PFAS


Figure 3 Examples of how the PFAS delay column effectively separates system PFAS from sample PFAS


Conclusion


The unique nature of per- and polyfl uoroalkyl substances in providing inert materials that have high thermal insulation properties and high levels of hydrophobicity have resulted in them becoming ubiquitous across society and within the analytical laboratory. This study has highlighted that PFAS can be found in a variety of sources from nitrile gloves to paper tissues and even in some septa used in sample vials. It is also feasible for background PFAS compounds to originate from the HPLC system, specifi cally PEEK tubing, however the use of a delay column and carefully chosen method conditions will ensure that any contamination of the HPLC system or solvents can be separated from sample PFAS components meaning that accurate quantifi cation can be achieved.


READ, SHARE or COMMENTon this article at: envirotech-online.com/article


Intensity, cps


PFOA


PFDA


PFOS


PFNA


Peak area Intensity, cps


NEtFOSAA PFBS PFDA


PFHpA PFHxA PFNA PFOA PFOS PFTA


PFTrDA PFUnA


HFPO-DA


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84  |  Page 85  |  Page 86  |  Page 87  |  Page 88  |  Page 89  |  Page 90  |  Page 91  |  Page 92  |  Page 93  |  Page 94  |  Page 95  |  Page 96  |  Page 97  |  Page 98  |  Page 99  |  Page 100  |  Page 101  |  Page 102  |  Page 103  |  Page 104  |  Page 105  |  Page 106  |  Page 107  |  Page 108