40
What If You Were Told to Create a Pipette Management SOP Within Three Weeks?
As many people do, you may have your pipettes inspected and calibrated regularly by an outside facility – once a year, or perhaps twice a year, depending on your workplace policies. But imagine what would happen to a year’s worth of data when, say, an annual calibration showed that you might have been using inaccurate pipettes all along!
Today, what with Good Laboratory Practice (GLP) and other quality assurance regulations/guidelines, researchers are in increasing need of sustainable ways to internally manage the accuracy of their pipettes. The problem is, however, that many of them would be at a loss as to what exactly they should put into actual practice. Wouldn’t you too?
Not to worry, though. As a solution, A&D provides two kinds of devices that you can use to introduce a simple, viable standard operating procedure (SOP) for managing pipettes.
Daily Check
The first of the two is the AD-1690 leak tester. Most pipette failures occur due to degraded lower parts like the o-ring, piston and tip holder compromising the pipette’s air tightness. By depressurising inside the pipette, the AD-1690 can determine whether an air leak exists in a matter of seconds, allowing you to quickly sort out troubled pipettes.
Leak Test Periodic Check
In addition to the above, you are advised to inspect the pipettes’ actual performance from time to time. For that purpose, a series of pipette accuracy testers (AD-4212B-PT, AD-4212A-PT and FX-300i-PT), or the BM series of micro analytical balances with their optional pipette accuracy testing kit (BM-014), are useful. The test procedure follows the gravimetric method as recommended by ISO8655.
With the special ‘WinCT-Pipette’ software, these testers perform mass-to-volume conversions and make accuracy and repeatability judgments based on the criteria you set all automatically for you. They are also equipped with an evaporation trap, which maintains high humidity inside and thereby prevents the evaporation of the test liquid that causes measurement errors.
For documentation requirements, the test results and other relevant information can be printed out in a report format with just one click.
Accuracy Test Dispensing Volume Correction
Now, supposing that some of your pipettes have been found inaccurate by your periodic check, what should you do? Send them for readjustment? Or try a confusing do-it-yourself readjustment method if that is made available by the manufacturer? The former is costly and the latter is not so easy. And both are time-consuming, aren’t they.
If you use the MPA series of electronic pipettes recently released by A&D, it is easy to readjust the dispensing volume by yourself, as all you have to do is simply input the correct value measured by a pipette accuracy tester into the pipette.
With the MPA series and the two above-mentioned devices, you will be assured of not only the quality of your pipettes but also the means to maintain and account for that quality whenever required.
Please contact A&D for more detailed information on each product.
29858pr@reply-direct.com
For everyday inspection, this leak test along with some quick appearance and function checkups will probably suffice for most users.
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80 |
Page 81 |
Page 82 |
Page 83 |
Page 84 |
Page 85 |
Page 86 |
Page 87 |
Page 88 |
Page 89 |
Page 90 |
Page 91 |
Page 92 |
Page 93 |
Page 94 |
Page 95 |
Page 96 |
Page 97 |
Page 98 |
Page 99 |
Page 100 |
Page 101 |
Page 102 |
Page 103 |
Page 104 |
Page 105 |
Page 106 |
Page 107 |
Page 108 |
Page 109 |
Page 110 |
Page 111 |
Page 112 |
Page 113 |
Page 114 |
Page 115 |
Page 116 |
Page 117 |
Page 118 |
Page 119 |
Page 120 |
Page 121 |
Page 122 |
Page 123 |
Page 124 |
Page 125 |
Page 126 |
Page 127 |
Page 128 |
Page 129 |
Page 130 |
Page 131 |
Page 132 |
Page 133 |
Page 134 |
Page 135 |
Page 136 |
Page 137 |
Page 138 |
Page 139 |
Page 140 |
Page 141 |
Page 142 |
Page 143 |
Page 144