INNOVATIVE CALIBRATION STRATEGIES FOR


ROUTINE QUALITY ASSURANCE AND QUALITY CONTROL OF ANALYZERS FOR REACTIVE TRACE GASES, WITHOUT THE NEED FOR PRIMARY STANDARDS


Reliable, sub-ppb trace gas measurements are essential for a range of industrial process and air quality monitoring purposes.


Picarro’s Cavity Ring-Down analyzers (CRDS) have become the preferred technology for stable and continuous, high-precision trace gas measurements since the commercial release in 2005. Currently, they are used for a range of gas monitoring purposes, such as: fence line emissions near industrial facilities (HF, H2 H2


S, EtO), air quality inside sterilized isolator spaces (H2


S, EtO), process control to improve yield (HF, SO2 HCl), and many other applications.


H2 O2 , NH3 ),


occupational exposure limits to toxic gases, stack emissions monitoring to comply with regulations and emissions (HF, H2


, H2 CO, S,


A key feature of CRDS analyzers is their unprecedented stability and ease of use. This is thanks to the low calibration requirements, which is accomplished by the patented wavelength monitor. wavelength monitor tracks one or more well-defi ned spectral reference lines (most commonly: water), which guarantees the unmatched low drift. Therefore, the performance of Picarro’s industrial trace gas analyzers only needs to be validated on an infrequent basis (e.g., during a yearly performance validation).


Commonly, a calibration of crucial gas measurement equipment is performed with commercially available and certifi ed gas standards, specifi cally to determine the linearity and the zero value of the analyzer. Importantly, sourcing accurate standards for hazardous, corrosive, and reactive trace gases is challenging and often technically impossible. To overcome this limitation of such standards, Picarro established a three-fold approach: 1. Factory calibration with a golden analyzer, 2. Validation of the linearity of an analyzer using a surrogate gas approach, and 3. Accurate determination of the zero value of an analyzer.


1. Golden analyzer approach


The golden analyzer is a carefully calibrated inhouse reference instrument with exceptionally low drift that is used as a transfer standard to cross-calibrate production units. The golden analyzer


2. Surrogate gas validation


To offer a straightforward calibration strategy for reactive and hazardous gases, Picarro has developed the method of surrogate gas validation for gases such as ammonia (NH3 chloride (HCl), hydrogen peroxide (H2 and formaldehyde (H2


), hydrogen O2 ), hydrogen fl uoride (HF), CO) (see Table 1). The principle is that two


adjacent adsorption lines of individual species behave in similar manners. As such, the accuracy and linearity of the analyzer can be validated using surrogate gases. Picarro selected these surrogate gases based on the spectroscopy for each analyzer. Key criteria for this selection process were: A. non-reactive gas, B. Commercially available certifi ed cylinders, and C. Adsorption line adjacent to the primary gas.


For most analyzers, Picarro recommends that the linearity is verifi ed by measuring three different surrogate gas concentrations, in addition to the zero-point measurement (see point 3 ‘accurate determination of zero value’). In practice, reactive trace gas analyzers from Picarro are assessed with a span accuracy of 5% or 10% (measured on the surrogate gas), refl ecting the specifi ed accuracy of the certifi ed surrogate gas standards.


3. Accurate determination of zero value


As mentioned, the drift of Picarro’s CRDS analyzers is constrained by the patented wavelength monitor which tracks one or more well-defi ned spectral reference lines. This impedes highly


Table 1: Hazardous, corrosive, and reactive primary gases, and the corresponding surrogate gases and concentrations that can be used to validate the performance of the Picarro CRDS analyzers. Note that the concentrations of the surrogate gases are usually higher than their natural abundances because the spectral features of the surrogate gases tend to be weaker.


Model G2103 G2108 PI2114 G2205 G2307


Primary Gas Ammonia (NH3 )


Hydrogen chloride (HCl) Hydrogen peroxide (H2O2


Hydrogen Fluoride (HF) CO)


Formaldehyde H2 ) Surrogate Gas


Methane (CH4 Methane (CH4 Oxygen (O2


) Methane (CH4 )


Carbon dioxide (CO2 )


) )


Surrogate Gas Concentrations 0, 200, 1 000, 10 000 ppm 0, 7, 50, 100 ppm 0, 7, 50, 100 ppm 0, 20.94% (ambient) 0, 7, 50, 100 ppm


CO,


is regularly checked against a non-reactive proxy gas (see point 2 ‘surrogate gas validation’) and trusted, rare primary gas standards. It is important to note that the accuracy of the golden analyzers is not limited by the measurement precision of the trace gas standards, but rather the limited accuracy of available primary standards.


undesirable monotonical drift and thereby guarantees unmatched long-term stability. For most applications, the zero value only needs to be determined infrequently (e.g., together with the yearly span validation). An example of an exception to this is the measurement of ethylene oxide for ultra-low background monitoring purposes, where more frequent zero referencing delivers the desired low ppt performance.


One last aspect to keep in mind is that, depending on the trace gas of interest, accurate determination of the zero value is complicated because ultra-high purity (UHP) zero air may contain ppt to ppb amounts of the target gas. For this purpose, Picarro tested a range of scrubbing agents to assure convenient, reliable, and accurate determination of the zero value within an acceptable specifi cation (e.g., phosphoric impregnated activated charcoal (PAIAC) is recommended for ammonia, activated charcoal and coconut shell for hydrogen chloride and hydrogen fl uoride, manganese greensand for hydrogen peroxide, and DrieRite for formaldehyde).


For more information on best practices for QA/QC of reactive trace gas measurements, please download the Rapid Analyzer Validation Using a Traceable Surrogate Gas Approach paper.


Author Contact Details Dr. Magdalena Hofmann, Senior Application Scientist, Picarro • Web: www.picarro.com


IET SEPTEMBER 2022


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84  |  Page 85  |  Page 86  |  Page 87  |  Page 88  |  Page 89  |  Page 90  |  Page 91  |  Page 92