Safety
To demonstrate is sensitivity, some performance tests have been done. Analysis of 8 Sulphur compounds from external certifi ed standards have been done. Below are shown the results obtained for the stability tests and linearity tests after 20 consecutive analysis performed on the following components:
All tests performed are part of a protocol document pertaining to analyser validation and are defi ned by a European third part laboratory. To validate the instrument, it is required to perform 20 analysis and to get the results in compliance with ISO 5725-2. The samples are generated from different standards.
Hydrogen sulphide
Methyl Mercaptan (MM or MTM) Ethyl Mercaptan (EM or ETM) Dimethyl Sulphide (DMS)
(iso) 2-Propyl Mercaptan (IPM) ter Butyl Mercaptan (TBM) (N) 1-Propyl Mercaptan (NPM) TetraHydroThiophene (THT)
Table 1: List of molecules analysed during the test
Table 2 here above summarise the different results obtained per compound. Concentration (mg/m3
)
Mean SD
Relative Error (%) Repeatability (%)
Reference concentration
H2S MM EM IPM TBM THT DMS STD 3,16
9,06
0,011 1,50 0,72 3,11
(+/-4%) Table 2: Concentrations obtained for H2 6,02
0,21 2,38 6,01
8,05
2,06 1,20 8,22
5,18
0,96 1,21 5,13
27,20
0,031 0,072 0,048 0,031 0,146 0,84 0,68 9,14
0,51 1,07
(+/-4%) (+/-4%) (+/-4%) (+/-4%)
27,06 (+/-4%)
6,04
0,021 0,19 0,71 6,03
(+/-10%) S, MM, EM, IPM, TBM, THT and DMS STD over 20 measurements
Table 3 represents the criteria for compliance with ISO 19739 norm and the obtained values using the energyMEDOR®
instrument. Repeatability (%)
Performance criteria
H2 S
MTM (or MM) ETM (or EM) IPM
TBM THT
3 2 4
10 7 4
Obtained value 0,72
0,68 2,38 1,20 1,21 1,07
Relative reproducibility (%)
Performance criteria
25 10 30 20 25 20
Obtained value 1,50
0,84 0,21 2,06 0,96 0,51
Table 3: Comparison between performance criteria in compliance with ISO 19739 and obtained values.
The relative reproducibility and repeatability values are much better than performance criteria values as defi ned in the ISO 19739 norm. All tested compounds have a linear response (to SSD) in the trial conditions (i.e. range of 0 – 5 mg/m3
for most of sulphur compounds and THT range 0-25 mg/m3
with a R2 > 0.995 for all compounds. So the conclusion was that the performance criteria are in compliance with the norm.
An example of a chromatogram obtained with the energyMEDOR® around 1mg/m3
is shown in Figure 1 below. for the measurement of H2 S at ) Conclusion
There is a clear global tendency for analysing Sulphur compounds with speciation in the oil and gas fi eld but also on other markets, with a demand under expansion in industries like shale gas, Biogas, odour and CO2
in beverage. This should remain important in coming years.
Certifi cations are also a key issue for the company as it is important to continue to invest in these processes to validate the technologies for specifi c fi elds and applications. Chromatotec® Ex d applications and sampling module development allow to extend the scope of its MEDOR®
evolution.
investments in solution.
The diversifi cations of the processes and the need for effi ciency improvements result in customers looking for customised and performant solutions. These specifi c needs are at the base of Chromatotec®
Chromatotec® ’s energyMEDOR® appears to be the best alternative for these applications since it only
needs a small amount of nitrogen or zero air to operate its Sulphur Specifi c Detector. This certifi ed solution has been designed to operate in zone I and II and can be used with 230V, 115V and 24V power supply making it quite unique on such type of market.
Author Contact Details Figure 1: Typical chromatogram obtained with energyMEDOR
Franck Amiet – president and CEO & Jean-Philippe Amiet – M. Sc. Export Project Manager Mickaël Gezat – Project Manager, Chromatotec • 15 rue d’artiguelongue, Saint-Antoine, 33240 Val de Virvée France • Tel: +33 (0)5 57 94 06 26 • Email:
info@chromatotec.com • Web:
www.chromatotec.com
Figure 2: Analysis of H2 S, TOS and TS in two minutes H2 S CH3-SH
CH3CH2-SH CH3-S-CH3
(CH3)2-CH-SH (CH3)3-C-SH
CH3CH2CH2-SH C4H8S
At the end of each chromatogram, a validation of results is done by injecting and analysing the DMS from the internal calibration device (permeation tube).
Another application for natural gas market is related to the Integrity Monitoring. During the extraction of raw natural gas and following processing, Midstream companies are required to track the level of H2
S and TS (Total Sulphur = H2 S +
Total Organic Sulphur (TOS)). If the concentration measured exceed the required levels, the Midstream companies must shut down the gathering line until the measurement are showing the required concentration levels again.
S from the other sulphur species (shown in Figure 2).
For these analysis, the sample is fi rst loaded into a loop and then injected in an analytical column to separate H2
To expand the applications of the MEDOR® CHROMATOTEC®
, a specifi c sampling module has been developed by to inject extremely low volume (from 0.1 to 1 µL) into the autoGC. LPG analysis
(Propane and butane), which is directly connected in the sampling module inlet. Then it is vaporised and injected into the MEDOR®
thanks to the very low volume injected the MEDOR®
directly. Another application is for high concentration up to low %: instrument is not disturbed.
Other applications for heavier liquid hydrocarbon like condensate or gasoline are also available; as soon as they can be vaporised in the heated sampling module it is possible to analyse them with MEDOR®
.
53
All of our articles are online! To view and download them, visit:
www.petro-online.com
ANNUAL BUYERS GUIDE 2018 •
WWW.PETRO-ONLINE.COM
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80 |
Page 81 |
Page 82 |
Page 83 |
Page 84 |
Page 85 |
Page 86 |
Page 87 |
Page 88 |
Page 89 |
Page 90 |
Page 91 |
Page 92 |
Page 93 |
Page 94 |
Page 95 |
Page 96 |
Page 97 |
Page 98 |
Page 99 |
Page 100