Analytical Instrumentation
GC-VUV Method Provides a Diesel Tank Corrosion Screening Tool
VUV Analytics, Inc. recently developed a GC-VUV and static headspace method to monitor for low-level ethanol and methanol contamination of ultra-low sulfur diesel (ULSD) underground storage tanks. Ethanol is a common “oxygenate” added to automobile gasoline in the United States to enable more complete combustion and reduce air pollution. Typically, ethanol is not used in diesel fuel but may find its way into diesel storage tanks through cross-contamination with ethanol-containing gasoline (e.g., E10, which contains 10% ethanol). According to a recent EPA report, ethanol may support microbiologically-influenced corrosion (MIC) of diesel storage and delivery systems when low molecular weight organic acids, such as acetic acid, are formed during ethanol’s microbial degradation by certain bacteria1
contribute to MIC via formation of formic acid when diesel contains biodiesel. Corrosion has been reported to cause failures in valves, filters, seals, and other equipment which ultimately leads to underground storage tank leaks and groundwater pollution.
Detectability of polar alcohols at low levels on a gas chromatograph (GC) can be problematic due to its tendency to produce broad, tailing peaks on GC columns or be lost completely due to adsorption. Static headspace offers an elegant solution to the GC sample introduction problem by employing a non-volatile mineral oil diluent that allows diesel sample volumes of several hundred microliters. 200 µL of diesel is placed in 2 mL mineral oil in a 20-mL headspace vial, heat/shake equilibrated for 10 min in a Gerstel MPS2 autosampler, followed by injection of the headspace into a GC-VUV system to collect absorbance spectra for methanol and ethanol. Hydrophobic mineral oil diluent suppresses volatilisation of the diesel hydrocarbons that would otherwise overload the GC column, while simultaneously promoting ethanol transfer to the headspace for sampling and analysis.
GC-VUV absorbance data is three dimensional (time, absorbance, wavelength) and specific to compound chemical structure. Compounds that indirectly contribute to corrosion such as water and alcohols can be positively identified by VUV spectroscopy in complex matrices like ultra-low sulfur diesel (ULSD). In addition, ethanol and methanol spectra are easily differentiated. Gasoline hydrocarbon compounds of interest that co-elute during the GC run can be spectrally deconvolved for accurate quantitative analysis.
An example of the GC-VUV and static headspace method for ethanol and methanol analysis is shown in Figure 1. Methanol and ethanol diluted in DMSO to concentrations of 5, 10, 20, 50, 100, 200, 500, and 1000 ppm were spiked into 200 µL diesel. An Agilent 6890 GC was used with its injection port set to 250°C. A helium carrier gas flow rate of 4 mL / min was used throughout the experiment. The oven profile for the residual solvent analysis began at 35°C (held for 1 min), followed by an increase to 275°C at a rate of 30°C/min. A split ratio of 2.5 was utilized to help maximise sensitivity. A total GC runtime of 12 minutes was programmed for each sample analyzed. A VGA-100 GC detector was used in this experiment with the transfer line & flow cell temperatures set to 275°C. The makeup gas pressure used was approximately 0.36 psi. An acquisition range of 120 to 240 nm was selected with an acquisition speed of 4.5 spectra/sec. A GERSTEL Multi-Purpose Sampler (MPS) was used with the syringe temperature held constant at 90°C, an incubation temperature set to 80°C, and an incubation time set to 10 min. An injection volume of 250 µL was selected to ensure good peak shape while the agitator was set to 250 rpm with an injection speed of 200 µL/sec. Figure 1 demonstrates excellent dynamic quantitative range for methanol and ethanol between 5 – 1000 ppm in diesel. Inset shows an overlay of peaks corresponding to water, methanol, and ethanol at concentrations ranging from 5 – 100 ppm concentration spikes. The ability to simultaneously detect water opens the possibility of a more rigorous approach to water analysis in fuel stream samples.
References
(1) Investigation of Corrosion-Influencing Factors in Underground Storage Tanks with Diesel Service, U.S. Environmental Protection Agency Office of Underground Storage Tanks, EPA 510-R-16-001, July 2016.
For More Info, email: email:
For More Info, email: email:
44507pr@reply-direct.com vuvanalytics.com
For More Info, email: email:
For More Info, email: email:
VUV VE R I F I ED
7192ad@reply-direct.com ANNUAL BUYERS GUIDE 2018 •
WWW.PETRO-ONLINE.COM
25
. Methanol, a component of biodiesel production, is also suspected to
Figure 1: GC-VUV and static headspace GC-VUV method for methanol and ethanol detection in Diesel. A chromatographic overlay of 5, 10, 20, 50, 100, 200, 500, 1000 ppm methanol and ethanol spikes in 200 µL diesel is shown. An overlay of peaks corresponding to water, methanol, and ethanol at concentrations ranging from 5 – 100 ppm concentration spikes can be seen in the chromatogram zoom inset.
VUV Absorption Spectroscopy.
GC detection in a whole new light.
Making Vacuum Ultraviolet absorption spectroscopy available to the general scientifitic community for the first time. As in, ever.
Universal, yet selective detector with very sensitive spectral response
Unambiguous identification & quantitation of compounds, including isomers
Shorter GC runtimes
Easy deconvolution and quantitation of co-eluting analytes
No calibration required - 1st principle detection technique provides a predictable linear response
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80 |
Page 81 |
Page 82 |
Page 83 |
Page 84 |
Page 85 |
Page 86 |
Page 87 |
Page 88 |
Page 89 |
Page 90 |
Page 91 |
Page 92 |
Page 93 |
Page 94 |
Page 95 |
Page 96 |
Page 97 |
Page 98 |
Page 99 |
Page 100