Analytical Instrumentation


Hydrogen Carrier Gas for Total Petrolium Hydrocarbon Analysis


Ed Connor, GC Product Specialist, Peak Scientific Fountain Crescent, Inchinnan Business Park, Inchinnan, PA4 9RE, Scotland, UK Tel: +44 141 812 8100 • Email: marketing@peakscientific.com • Web: www.peakscientific.com


Many labs are facing helium supply difficulties and rapid price increases as a result of the helium shortage. Lab managers are therefore looking for alternative carrier gases and for many GC applications, hydrogen provides an excellent alternative to helium. Using a hydrogen generator can supply gas to a whole lab and at the same time remove the danger of storing hydrogen in the lab. Laboratories that switch gas supply from helium to hydrogen can see a number of benefits including reduced overheads, faster throughput and even improved separations.


Carrier gases


Nitrogen, hydrogen and helium have differing relative carrier efficiencies along a velocity gradient (figure 1). Nitrogen offers good carrier efficiency at low velocities, but at high velocities, loses carrier efficiency. For this reason, nitrogen is not usually favoured for applications where helium or hydrogen can be used. Hydrogen offers very good efficiency at medium linear velocities and out- performs helium, in terms of carrier efficiency at very high velocities.


In a response to growing demand for hydrogen for carrier gas, GC manufacturers have recently produced a number of systems or instruments that improve efficiency of helium use or allow a switch to hydrogen for carrier gas. Agilent have developed an EPC device for the GC-7890B which switches carrier gas from helium to nitrogen carrier gas when the system is idle. Agilent, Bruker and Shimadzu have all developed GC/MS systems that require no hardware changes in order to use hydrogen or helium carrier gas and Thermo have tested all of their GC systems for hydrogen carrier gas compatibility.


Hydrogen is already used as a flame gas for FID detectors, one of the most commonly used GC detectors, however, the perceived health and safety risk of hydrogen means that many laboratories are now prohibited from having hydrogen cylinders in the laboratory. The safety concerns regarding placement of hydrogen cylinders in the laboratory stems from the large volume they contain (a 50 litre cylinder contains around 10,000 litres of pressurised hydrogen) and the dangers associated with moving heavy cylinders around. The high volume of hydrogen present in cylinders is a potential danger since a volume of 4% hydrogen in air equals the minimum threshold (lower explosion limit) for hydrogen ignition1


. Hydrogen can also undergo auto-ignition if it is rapidly released into air2. Possibly the biggest problem facing users of


hydrogen cylinders is cylinder to cylinder variation in hydrogen quality. Changing empty cylinders can interrupt GC operation, introduce contaminants into the system as well as inconveniencing the user. Cylinders are also large and can take up a considerable amount of lab space.


Gas generators, on the other hand have a relatively small footprint, contain a very low volume of hydrogen at any one time (less than 1000cc) and are able to produce gas on-demand to simultaneously supply carrier and flame gas to a number of GCs. The purity of gas produced by a hydrogen generator is often of higher purity than cylinder hydrogen. In addition to these features, hydrogen generators typically have forced air ventilation to prevent mixing of hydrogen and oxygen within the generator and an alarm system in case of low or high pressure. These features along with in-oven leak detectors supplied by hydrogen generator suppliers and emergency inlet shut-down features of GCs mean that laboratories should be quickly alerted in the event of a hydrogen leak.


Benefits of hydrogen carrier gas


The benefits of switching to hydrogen carrier gas from helium can include improved column separation efficiency over a wider flow range, improved sample throughput and lower running costs. Increased sample throughput without loss of resolution is an attractive feature of using hydrogen for carrier gas that should appeal to contract labs for which reduced overheads and faster chromatography will increase profits.


Hydrogen is often the best alternative to helium for carrier gas, although it can react with analytes under certain conditions. Therefore, chromatographers should be careful when using chlorinated solvents with hydrogen carrier gas because of the risk of hydrochloric acid (HCl) formation, which can affect column efficiency through formation of pores in the stationary phase. However, creative use of the split/splitless inlet pneumatics can allow rapid transfer of analytes onto the column, by minimising formation of HCl in the inlet3


. In some MS applications,


the reactivity of hydrogen can be used to the advantage of the analyst where protonation of ion fragments can be used in compound identification.


Figure 1: Van deemter curve showing carrier efficiencies of nitrogen, helium and hydrogen.


Figure 2. 12 compound mixture run using (2a) helium carrier gas, (2b) hydrogen cylinder carrier gas and (2c) hydrogen generator carrier gas


AUGUST / SEPTEMBER 2013 • WWW.PETRO-ONLINE.COM


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52