few of the applications that utilise single- use tube sets.


Selecting tubing is important because the tubing must be durable, resistant to chemicals (internal and external of the tubing), compatible with the connectors that will be used and not prone to any kinks that might occur as the tubing is bent in different positions. If the connections to the tubing will be barbs, tubing should be selected to ensure a secure connection to the barb but allow for easy assembly in production. If you have the right tubing selected for performance and flow, the selection of the appropriate barb to go into the tubing on the connector side is the next step. If the connectors will be bonded in, then the tubing selected must be of a bondable material.


Whether bond-in or barbed, tubing connectors are selected based on the application. Most of the time, a tube set will have an end connection on both ends of the tubing. For example, in an ophthalmic surgery application, one end of the tubing might be a luer connection to the panel of the machine, while the other end is a customised connector to connect to the surgical instrument. In an IV tube set, one end of the connection could be a bag spike to an IV bag, while the other end is a luer to connect to a needle to the patient.


There are many different connectors that may be selected within the tube set to complete the application and full performance of the medical device. Flow control connections are often selected based on the application. Flow control connections could be check valves, filters, drip chambers and flow restrictors. External to the tubing, tube clamps can be used to control or stop flow in the tube set. Often, the connectors are chosen based on their performance characteristics to the specification, but in most cases there are multiple options for connectors from multiple sources. Potentially more important than the internal medical device tubing and connectors is that the external tubing and connectors should be ordered from a reliable source to ensure assurance of supply so that delivery issues with components is minimised. To summarise, the four significant criteria in selecting tubing and components are performance in


28


the application, reliable delivery and assurance of supply, quality of the component, and competitive pricing based on expected annual volumes. Working with various suppliers of tubing and components is an important part of the design process. Some suppliers are the manufacturers of the components, where others are not. If the market for the medical device is a high-volume supply of components, then working directly with the manufacturer is a good solution. These companies usually have minimum volume requirements and lead times that must be considered. In most medical applications, it is important to choose a manufacturer that performs their plastic moulding in a cleanroom environment. It is beneficial to work with a


distributor/supplier who can supply all the necessary components that are needed on the tube set. Rather than working with a variety of suppliers, the designer and buyer can work with one company to deliver all the connectors needed. Qosina is the most well-known company in this category, and in some cases, Qosina owns the component mould and design. Qosina carries thousands of stock components and will sell in small volumes, as they specialise in down-packaging performed in a 1,700ft2


ISO-8 class 100,000 cleanroom. There are multiple factors that should be evaluated when choosing suppliers for the internal or external tube set parts: ■


■


confirmation if supplier maintains a quantity management system review of supplier processes for


Selecting tubing is crucial in ensuring that, amongst other things, it does not kink if the tubing is bent. component quality issues or replacement


■ ■ ■ ■


moulding and packaging the components or tubing in a cleanroom environment


ability to provide samples for prototypes with reasonable minimum order quantities


ability to provide drawings (preferably 3D) for all components


ability to provide all material specifications to satisfy regulatory standards whether the standards are US or international


■ ■ ■


confirmation of supplier lead times on the parts specified and alignment with project needs


review volume pricing to determine best possible prices based on annual blanket orders


confirmation of part availability with no plans to obsolete the part you have specified


Choosing the right tubing and components for your medical device design can seem like a simple activity, but attention to the details as described in this article will help ensure the device is successful in the market with minimal supporting design changes after product release. The most crucial criteria is to select components that perform the best in the design for the intended application, but keep in mind that the vendor supplying the component can have a major impact on the lasting success of the medical device. ●


www.qosina.com Medical Device Developments / www.nsmedicaldevices.com


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84  |  Page 85  |  Page 86  |  Page 87  |  Page 88  |  Page 89  |  Page 90  |  Page 91  |  Page 92  |  Page 93  |  Page 94  |  Page 95  |  Page 96  |  Page 97  |  Page 98  |  Page 99  |  Page 100  |  Page 101  |  Page 102  |  Page 103  |  Page 104  |  Page 105  |  Page 106  |  Page 107  |  Page 108  |  Page 109  |  Page 110  |  Page 111  |  Page 112  |  Page 113  |  Page 114  |  Page 115  |  Page 116  |  Page 117  |  Page 118  |  Page 119  |  Page 120  |  Page 121  |  Page 122  |  Page 123  |  Page 124  |  Page 125  |  Page 126  |  Page 127  |  Page 128  |  Page 129  |  Page 130  |  Page 131  |  Page 132  |  Page 133  |  Page 134  |  Page 135  |  Page 136