Company insight
Component selection criteria in medical devices
For most medical device design engineers, selecting components from multiple vendors is a critical step in the design process. Whether the components and tubing are inside the device or part of a disposable tube set, there are many different criteria that should be considered in the selection process. This article by Qosina will provide guidelines on where to look when searching for the right components for the operations needed.
epending on the type of medical device, there can be quite a bit of tubing, connectors, pumps, fluid regulators and other fluid management sub-systems. Each section of tubing, connections and pumps completes a task for fluid or air flow inside of the device. These can be found in medical devices such as ventilators, in-vitro diagnostic devices, dialysis devices, blood pump systems, etc. Selecting the appropriate tubing and connector components can make a significant difference in device performance and manufacturability. Tubing is often necessary to keep fluid or air moving. Tubing selection factors include durometer, ID/OD and wall thickness, connection to pumps or tubing connectors and fluid paths within the medical device. For example, tubing must often be worked around other sub- systems within the device. This means the designer must be careful not to place tubing in a position where it could kink within the device. The durometer and thickness of the tubing can help reduce the potential for kinks. In addition, tubing needs to be selected based on rigidity, chemical resistance and fluid flow dynamics. Assembly of the tubing with the pumps and connectors within the device is critical to the production teams that need to consistently assemble the full device for performance to design specifications.
D
Connectors on the inside of the device are utilised to connect tubing from multiple subassemblies in order to support functional performance in the device. The connectors chosen must be easily assembled to the tubing but also ensure a robust connection, preventing leaks. Often, connectors used internally
in medical devices are either barbed (inserted in the tube), compression (holds the outside of the tube) or bond-in (glued/bonded to the outside of the tube). Connectors are chosen to pass fluid or air and to control the flow of fluid or air backward and forward. This may be handled by standard open-flow connectors, check valves and filters. Check valves control forward flow and limit back flow of fluid. Filters limit the type of fluid flowing and remove any particulate that should not continue to flow forward in the device.
Tubing in medical devices are vital for keeping fluid or air moving, and must be carefully maintained. for the application.
Medical device designers do extensive prototyping of subassemblies in laboratories. This way, tubing and connector options can be tested to ensure the best design for manufacturability and performance. Different types of pumps and flow control devices can be tested to make sure that tubing and connectors are chosen to optimise the best fluid control
Medical Device Developments /
www.nsmedicaldevices.com
Cost is typically not a major factor when selecting tubing and connectors for the inside operations of a medical device. This is because the volumes required are usually low, and the surrounding mechanical systems are significantly more expensive. The primary goal is to ensure that operations within the device are maximised to the specifications. Special attention is needed to keep the device easy to assemble and allow for quality checks to ensure consistent performance. Typically, tubing and connectors used outside of the medical device get a lot more attention. This is because these tube sets are the single-use side of the device and generate the most revenue for the medical OEM. External medical device tube sets are related to devices such as infusion pumps, suction/irrigation for surgery, blood transfusion, dialysis and ophthalmic surgery. These are only a
27
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80 |
Page 81 |
Page 82 |
Page 83 |
Page 84 |
Page 85 |
Page 86 |
Page 87 |
Page 88 |
Page 89 |
Page 90 |
Page 91 |
Page 92 |
Page 93 |
Page 94 |
Page 95 |
Page 96 |
Page 97 |
Page 98 |
Page 99 |
Page 100 |
Page 101 |
Page 102 |
Page 103 |
Page 104 |
Page 105 |
Page 106 |
Page 107 |
Page 108 |
Page 109 |
Page 110 |
Page 111 |
Page 112 |
Page 113 |
Page 114 |
Page 115 |
Page 116 |
Page 117 |
Page 118 |
Page 119 |
Page 120 |
Page 121 |
Page 122 |
Page 123 |
Page 124 |
Page 125 |
Page 126 |
Page 127 |
Page 128 |
Page 129 |
Page 130 |
Page 131 |
Page 132 |
Page 133 |
Page 134 |
Page 135 |
Page 136