SPECTROSCOPY


T e backscatter elimination cup blocks backscatter from the sample holder.


INSTRUMENT T e Rigaku ZSX Primus and ZSX Primus IV instruments are fl oor- standing sequential wavelength dispersive X-ray fl uorescence (WDXRF) spectrometers, which have advantages in high spectral resolution and high sensitivity from light to heavy elements. T e instruments are designed to off er fl exibility and versatility for a wide range of applications. T e spectrometers referenced here are equipped with an Rh target X-ray tube. T e maximum tube power is 4kW for the spectrometers. Up to 10 analysing crystals can deliver rapid quantitative determination of major and minor atomic elements, from beryllium (Be) to uranium (U). Instrument control software


provides users with easy operation. T e fl owbar system fully supports setup of qualitative and quantitative analysis. T e ZSX Primus IV spectrometer was used for measurement in this example. T e measurement conditions are shown in Table 1.


Ci : certifi ed value of standard sample Cˆ i : calculated value of standard sample n : number of standard samples m : degree of freedom (2: linear, 3: quadratic).


TABLE 1. Table shows the measurement conditions


ANALYSIS RESULTS Calibration curves of boron and fl uorine in water using the Ultra Carry method were generated. T e calibration summary is shown below.


T e accuracy of calibration was calculated by the following formula:


Accuracy =


CONCLUSIONS T e results show that the Ultra Carry fi lter paper is applicable to boron and fl uorine analysis in water. T e LLD of boron and fl uorine in water is 8.4 ppm and 0.4 ppm respectively. Since the water solution is dried on Ultra Carry, samples can be measured under vacuum, enabling analysis of elements from boron to uranium without the need of helium gas. Ultra Carry also facilitates analysis of trace elements lower than 0.1 ppm, or at ppb level, in water solution. Since sample preparation is very simple, it is possible to reduce analytical error derived from sample handling and also shorten the total analysis time.


Michael Nelson is with Rigaku. www.rigaku.com


e


r


3D Raman image of a pharmaceutical ointment. 3D Raman Imaging Raman • AFM • SNOM • MADE IN GERMANY www.scientistlive.com 41 RISE


Turn ideas into discoveries


Let your discoveries lead the scientifi c future. Like no other system, WITec’s confocal 3D Raman microscopes allow for cutting-edge chemical imaging and correlative microscopy with AFM, SNOM, SEM or Profi lometry. Discuss your ideas with us at info@witec.de.


www.witec.de


s


p


e


e


d


s


e


n


is


n


it


o


iv


i


y t


s


lo


t


u


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72