CHROMATOGRAPHY


to monitor around the world. Teir widespread use and environmental persistence make them truly a global issue. Concerns over possible health risks are driving environmental scientists to look for these compounds everywhere from penguins in the south to polar bears in the north, and many points in between, and they are discovering the extent of PFAS contamination is literally pole to pole. However, one other concerning place where at least some of these compounds are present is inside the very instruments used for PFAS analysis of environmental samples. Background PFAS contamination can interfere with low-level quantification requirements, which are usually in the range of parts per trillion (ppt) or below. Instrument-related background


WHY USE A DELAY COLUMN? P


oly- and perfluoroalkyl substances, or PFAS, are rapidly emerging as some of the most important environmental contaminants


Fig. 1. Installation position of the PFAS delay column


contamination can stem from LC components that contain fluoropolymers (e.g. PTFE), which can leach into mobile phase solutions. Using tubing and containers without fluorochemicals and switching instrument components to PEEK or stainless steel mitigates some of the risk of background contamination, but it does not eliminate the impact of interferences on trace-level analysis, so another solution is required.


In a standard set-up, instrument-related


Fig. 2. System-related PFAS contamination is present in a blank sample analyzed without a delay column installed


PFAS interferences move through the system with the mobile phase and collect at the head of the analytical column between sample injections. However, adding a second column, called a “delay column,” into the mobile phase flow path immediately before the injector is an effective way catch the background PFAS before they reach the analytical column (Fig. 1). In this scenario, when the sample is injected, only PFAS in the sample will focus at the head of the analytical column. Ten, during the course of a gradient analysis, PFAS from the sample will begin to move through the analytical column like normal. Te background PFAS trapped in the delay column during equilibration will be eluted and pass through the analytical column as well. However, the background PFAS will elute later than the sample PFAS because they entered the analytical column after the injected sample due to additional retention by the delay column prior to reaching the injector. In addition, the background PFAS never focus on the head of the analytical column because they are present in the flow path in continuum rather than as a discrete injection. Te net effect is that by using a delay column users can differentiate instrument-related PFAS background contamination from target PFAS in a sample. Fig. 2 is an example of the kind of


background contamination that interferes with accurate PFAS analysis. Te system


www.scientistlive.com 27


Mike Chang and Shun-Hsin Liang describe how to eliminate the impact of instrument-related PFAS interferences by using a delay column


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72