17 Results
The linearity of the HPLC system was tested using analytical standards of Ibuprofen and Paracetamol at a range of 10-400mg/L. The specifi c wavelengths of both Ibuprofen and Paracetamol were utilised by the Diode Array Detector. Figures 2 show the calibration curves of both target analytes, at the maximum absorbance wavelengths.
The SCION HPLC exhibited excellent linearity over the concentration range of 10-400mg/L for both target compounds, each with an R2 value of 0.9994. Following the calibration standards, the sample containing the Ibuprofen and Paracetamol tablet extracts was analysed. Figure 3 shows the stacked chromatograms of both target compounds at their specifi c wavelengths.
Figure 4. Absorbance spectra comparison of both Ibuprofen and Paracetamol.
Figure 3. Stacked chromatograms of target compounds at their specifi c absorbance wavelengths.
Ibuprofen has a specifi c absorbance wavelength of 237nm whereas Paracetamol has a maximum absorbance at 254nm. Utilising the multi-wavelength capability of the Diode Array Detector it was possible to detect both target compounds in a single injection, in under 12 minutes. The concentration of both target compounds was determined using the respective calibration curves. The quantitative value for Ibuprofen was 34.11mg/L whilst the Paracetamol was calculated at 39.24mg/L.
For further identifi cation and confi dence in results, the absorbance spectra of the target compounds were also compared with a spectral library. Figure 4 shows the absorbance spectra of both Ibuprofen and Paracetamol.
CompassCDS software was used for the spectral library comparisons, in which an in house reference library was used. Figure 4 shows the overlay of the absorbance spectra generated during both the sample analysis and the reference standard, in which both Ibuprofen and Paracetamol exhibited a 100% match, thus confi rming the positive identifi cation of both target compounds (as the spectra of the sample is overlaid with the reference and the match is 100%; the reference spectra is not visible). Repeatability of the system was performed with six consecutive injections of the 10mg/L calibration standard. Figure 5 shows the overlay chromatograms for both Ibuprofen and Paracetamol.
Excellent repeatability of the 10mg/L analytical standard was observed, over six consecutive injections; with an RSD of 0.27% for Ibuprofen at 237nm and 0.18% for Paracetamol at 254nm. Retention times were also stable and reproducible with RSD values of 0.06% and 0.04%, respectively. The excellent repeatability and linearity observed demonstrates the robustness of the SCION HPLC system.
Conclusion
SCION Instruments developed a fast, accurate method for the simultaneous identifi cation and quantifi cation of combination drugs by HPLC-DAD. This method is critical to numerous industries including pharmaceutical and toxicology. With a single injection and utilising the multi-wavelength capabilities of the Diode Array Detector it was possible to clearly separate Ibuprofen and Paracetamol, two commonly used over the counter medications, in under 12 minutes. Compound identifi cation was further confi rmed through absorbance spectral comparisons using CompassCDS. Excellent linearity and repeatability of the SCION HPLC-DAD was demonstrated with low RSD values and high R2 values.
Figure 5. Overlay of 10mg/L analytical standard chromatograms (n=6).
Read, Share and Comment on this Article, visit:
www.labmate-online.com/article Unique HPLC Columns for Extremely Polar Samples
Cogent Diamond Hydride™ HPLC Columns from Microsolv have a silica-hydride surface similar to other Cogent TYPE-C™ phases, but the uniqueness of Diamond Hydride™ is that they have a very small amount of carbon on the surface that adjusts the hydrophobicity to an important level.
It has been the choice of scientists working in metabolomics, bio-analytical or with LCMS for compounds such as amino acids, organic acids, carbohydrates, and very polar small molecules. Non-polar and polar peptides have been separated in the same run.
An important alternative to HILIC, these columns perform extremely well in Aqueous Normal Phase which uses Acetonitrile and Water but equilibrates very rapidly and the data produced is very precise. Available in Guard, Analytical or Prep sizes with 100A, 4um or 2.2um particle sizes. More information online:
ilmt.co/PL/Ezzq
53049pr@reply-direct.com
WWW.LABMATE-ONLINE.COM
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80 |
Page 81 |
Page 82 |
Page 83 |
Page 84