40 Measurement and Testing ISSUES WITH LIQUEFIED CALIBRATION GAS MIXTURES


Liquefied gas mixtures, or ‘liquid mixtures,’ are often used for calibrations where the application demands calibration with a ‘heavy’ or longer-chain hydrocarbon. Liquid mixtures have the advantage of closely matching the product being analysed (e.g. stream gas), while also offering increased efficiency by allowing an increased number of calibrations with the same cylinder, when compared to a gaseous mixture.


While liquid mixtures do have significant advantages, there are some disadvantages to consider.


Instability


Liquid mixtures are inherently unstable. They are best used quickly, as the longer they are left unused, the further they will move from the originally certified composition. This is due primarily to vapourisation and condensation. Vapourisation describes a liquid becoming a gas and condensation describes a gas becoming a liquid.


We described in a previous article (valve types) how the dual port/dip tube valve worked. A ‘headspace’ of gas (typically helium) is kept at a constant pressure. This headspace applies pressure on liquefied gases in order to keep them in the liquid phase. This is also known as an “overpressure.” The stability of the liquid mixture is directly related to the relationship between the gas contents and liquid contents in each cylinder.


The kinetic theory of matter describes how all matter is constantly in motion. The rate of which depends on the amount of energy it has: solids with the least amount of energy, and gases with the most. The amount of energy is directly related to the temperature the matter is in. Therefore, solids melt into liquids and evaporate into gases with increases in temperature.


In a gas mixture, the molecules of each component are constantly moving. This assists with homogeneity of the mixture by constantly mixing it – but it is also the reason for reactive compounds to decompose. They will eventually come into contact with another material and react into a new compound.


In a liquid mixture, the issue of reactive components meeting a reagent is still present but the main issue affecting stability is the transition of compounds between the liquid phase and the gas phases.


Vapourisation There are two types of vapourisation: evaporation, and boiling.


Evaporation can occur at any time where a molecule is near the surface of a liquid, has the energy to escape, and the gas into which it moves is not fully saturated. Evaporation occurs slowly so the effects may not be immediately apparent. Regardless, it will always occur inside the cylinder and can’t be avoided.


Boiling occurs when the vapour pressure of the liquid is equal or greater than the pressure exerted on it by its surroundings.


Boiling releases gases very quickly and must always be avoided. The vapour pressures of each component will determine the feasibility of the mixture, and the required ‘overpressure’ to keep them in the liquid phase.


Condensation


Condensation is the opposite of evaporation and can occur when the cylinder cools or the overpressure is increased. The previously evaporated gases now re-enter the liquid phase. This may sound like a good thing, but if the liquid mixture was used during the time that these components were in the gas phase, they will now re-enter the liquid phase at different ratios than originally intended. This will change the overall composition of the mixture.


The amount of evaporated gases the headspace can contain is relative to its saturation point. This point is determined by the temperature. Changes in temperature will change the saturation point thus allowing evaporation or condensation to occur.


Rules to follow to get the most out of your liquefied gas mixture


1. Keep the cylinder cool and the temperature stable. There is no advantage to heating a liquefied gas cylinder, unlike with a gaseous mixture where heat typically helps with stability and homogeneity of non-reactive components. Keep the temperature stable to prevent phase transition as much as possible.


2. Use the smallest size cylinder possible. As the contents of the cylinder are used the gaseous headspace increases in volume relative to the liquid. This means that there is greater a greater mount of space available for liquids components to evaporate into and more of the calibration mixture will be lost to the headspace. As all components evaporate at different rates, the longer the mixture is kept, the further it will move from the desired composition. By using a smaller cylinder, the cylinder contents will be consumed more quickly, reducing the amount of time for evaporation to occur.


3. Provide a constant overpressure. Rather than topping up the overpressure periodically, keep the overpressure permanently connected to a gas source of the appropriate pressure and gas type. This will prevent changes in saturation point of the headspace and will also maintain the temperature (when a gas is compressed, it heats up).


4. Keep a gaseous reference standard. Liquid mixtures are great for day-to-day verifications due to their efficiency. However, to ensure the accuracy of your instrument, you should always have a gaseous reference standard on hand for periodic calibrations and to keep track of the stability of your liquid standard.


Quick TOC Online Analyser for Pure Water


The QuickTOC purity online TOC analyser from LAR Process Analysers AG is specifically designed for pure water analysis. The sample is injected directly into the reactor using a special injection loop, thus avoiding sample contamination from any pollutants in the ambient air. At the heart of the TOC analyser is the reactor, where thermal oxidation at 1,200°C takes place. At this temperature, the complete oxidation of all organic compounds to carbon dioxide (CO2) is guaranteed – without any need of oxidation promoting catalysts. The formed CO2 will be measured by an NDIR (non dispersive infrared) detector resulting in the exact determination of TOC. The QuickTOCpurity determines the TOC content within 3 minutes.


The TOC analyser can measure up to 6 sample streams in one unit. Due to the patented calibration and validation technique, called QuickCalibration, operators can check the analyser automatically and remotely at any time. Hence, time consuming preparation, costly liquid standards and on-site validation measurements are avoided.


LAR’s QuickTOCpurity determines the TOC in less than three minutes and thus, allows the operators to take fast countermeasures in case of water pollution. For More Info, email:


43130pr@reply-direct.com Complex Analysis Made Simple


For complete and precise analysis of complex samples, a dual Thermal-Desorber Gas Chromatograph (TDGC) equipped with two Flame Ionisation Detectors (FIDs) and one Mass Spectrometer (MS) is one of the most efficient analytical tools. A Chromatography Data System is vital for efficient and reliable operation of any modern on-line and continuous chromatography acquisition system.


VistaMS offers a unique solution for simple, automatic identification and quantification of molecules from complex samples and data transfer of your MS analyses. This software, included in all Chromatotec’s MS systems, can intercompare and validate results obtained using multiple detectors. In ambient air systems, data from two FIDs and one MS are timestamped, saved and analysed to provide the most reliable and precise results.


email: For More Info, email: email:


Analytical results can be easily re-processed changing all parameters of the TDGC-FID/ MS system (retention times, response factors, volume sampled, etc.). Automatic communication with a NIST library allows for confirmation of identification of the user’s complex samples. Physical parameters of the TDGC and MS as well as all analytical results can be transferred using Modbus protocol.


For More Info, email: email:


For More Info, email: 43132pr@reply-direct.com


5. Use a constant pressure cylinder. Constant pressure or “piston” cylinders are designed with a piston inside, forming a seal against the internal walls. This piston provides a physical barrier between the ‘overpressure’ gas and liquid mixture, preventing evaporation and condensation from occurring. Unfortunately, a piston cylinder can’t prevent boiling. Care still needs to be taken when designing a mixture. Components which boil off will form bubbles within the liquid mixture, altering the mixture ratios and resulting in a poor or incorrect calibration. Piston cylinders are very expensive and very small in volume. This may negate the efficiency of the liquid mixture and makes a gaseous mixture more viable.


Liquid mixtures are an important element in the hydrocarbon analytical industry. They can be extremely useful if used in the correct manner. Using the techniques above will ensure the standards are beneficial in your analytical calibration program.


For More Info, email: email:


For More Info, email: email:


Author Details


Paul Fry, CAC GAS & Instrumentation Unit 3, 36 Holbeche Road, Arndell Park, NSW 2148 Tel: +61 (0)2 8676 6500 • Email: cac@cacgas.com.au Web: www.cacgas.com


43513pr@reply-direct.com


AUGUST / SEPTEMBER • WWW.PETRO-ONLINE.COM


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60