This page contains a Flash digital edition of a book.
Bioanalytical Challenge Kumar, King, Clark & Gorovits


Table 2. Unconjugated antibody interference in intact antibody–drug conjugate assay. Spiked unconjugated antibody conc. (μg/ml) Accuracy for LQC (%RE)


0


2.5 5.0 25


HQC: High QC; LQC: Low QC.


in the recovery of ADC QC samples when spiked with Ab (data not shown). Accuracy is defined as the closeness of agree-


ment between the observed concentration (E) and the nominal concentration (T) and is expressed as percent relative error (%RE).


%RE `j T


= ET 100# -


The assay performance was evaluated in rodent


serum for an ADC containing lysine-based con- ventional conjugation chemistry. The assay for- mat on a Mesoscale Discovery (MSD) platform involved capture of intact ADC via its antibody framework using target protein as the capture reagent and detection using antibodies against the small molecule drug. High (1000 ng/ml) and low (25 ng/ml) QC ADC samples spiked with varying concentrations of Ab (0–25 μg/ml) were quan- titated against the ADC reference standard. The range of quantitation (ROQ) for standard curve was 10–1300 ng/ml in 100% matrix;


• Due to highly potent cytotoxic activity of small molecule component of ADCs, the potential tox- icity concerns are usually higher for ADCs. As a result, compared with large molecule therapeutics, relatively lower dose ranges may be selected for nonclinical and clinical studies for ADCs. LBAs for ADC bioanalysis thus may require high assay sensitivities (in the low ng/ml range) than typically expected for large molecule therapeutics [18–21];


• The selection of an appropriate biological matrix also presents unique challenge during ADC bio- analysis. The preferred matrix for bioanalysis of small molecule therapeutics using LC–MS plat- form and large molecule therapeutics using LBA platform are typically plasma and serum, respec- tively. However, since ADCs contain both small and large molecule therapeutics, the selection of optimal matrix for ADC bioanalysis becomes debatable.


1608 Bioanalysis (2015) 7(13) Depending on the linker lability and ADC sta-


bility, the processing of serum may cause deconju- gation and release of cytotoxic small molecule drug that may interfere in quantitation of small molecule analyte of some ADCs. The plasma matrix is thus preferred for LC–MS-based quantitation of the antibody-conjugated small molecule drug and/or the released unconjugated small molecule drug and its metabolites [9]. But the plasma processing is also associated with its own set of challenges such as the selection of appropriate anticoagulant from the list of available anticoagulants such as EDTA, heparin or sodium citrate; the appropriate volume of blood collected in collection tubes to ensure optimum blood/anticoagulant


ratio; and the freeze–thaw stability of plasma [22,23];


• Though not commonly observed, LBA perfor- mance may be impacted by the binding of known matrix components to either


large molecule or


small molecule component of multicomponent ADC [24]. Such interferences may reduce desired sensitivity of the assay.


Table 3 shows an example for the impact of


known matrix component interference on the assay performance in serum samples from rodents and nonhuman primates (NHP). Though similar assay format and similar experimental conditions were employed for quantitation of intact ADC in both species, the assay performance reflected by signifi- cantly reduced recovery of QC samples at the lower end of


the ROQ was impacted in NHP serum


samples. Further analysis revealed that the spe- cific binding of a known matrix protein to one of the component of multicomponent ADC in NHP serum caused the observed decrease in recovery of QC samples (data not shown).


The assay performance was evaluated in two indi-


vidual species, rodents and NHP, for an ADC with lysine-based conventional conjugation chemistry. The assay format on an MSD platform involved capture of intact ADC via its antibody framework using target protein as the capture reagent and


future science group


-19 2.0 -11 -65


Accuracy for HQC (%RE) -22 -18 -27 -79


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84  |  Page 85  |  Page 86  |  Page 87  |  Page 88  |  Page 89  |  Page 90  |  Page 91  |  Page 92  |  Page 93  |  Page 94  |  Page 95  |  Page 96  |  Page 97  |  Page 98  |  Page 99  |  Page 100  |  Page 101  |  Page 102  |  Page 103  |  Page 104  |  Page 105  |  Page 106  |  Page 107  |  Page 108  |  Page 109  |  Page 110  |  Page 111  |  Page 112  |  Page 113  |  Page 114  |  Page 115  |  Page 116  |  Page 117  |  Page 118  |  Page 119  |  Page 120  |  Page 121  |  Page 122  |  Page 123  |  Page 124  |  Page 125  |  Page 126  |  Page 127  |  Page 128  |  Page 129  |  Page 130  |  Page 131  |  Page 132  |  Page 133  |  Page 134  |  Page 135  |  Page 136  |  Page 137  |  Page 138  |  Page 139  |  Page 140  |  Page 141  |  Page 142  |  Page 143  |  Page 144  |  Page 145  |  Page 146  |  Page 147  |  Page 148  |  Page 149  |  Page 150  |  Page 151  |  Page 152  |  Page 153  |  Page 154