MICROPLATE READERS 21 END OF LIFE, LIVE THE Andrea Krumm explains how to monitor cell death in real-time C


ellular microplate assays are popular tools in life science. Tey enable studying cellular events that cannot be analysed in isolated cell components. Monitoring such a cellular process over time expands the output of an experiment as it allows comparing timing and extent of a biological reaction and as it prevents to miss the time point of reaction. Te microplate format is advantageous since various conditions can be analysed in parallel. Recording cellular reactions in real-time


requires specific characteristics of assays as well as of the measurement equipment. Obviously, a suitable microplate reader needs to be able to read the detection mode determined by the assay. For cellular assays these are predominantly fluorescence intensity, luminescence, absorbance, time- resolved fluorescence or combinations thereof. Te opportunity to measure from either top or bottom of the microplate is desired to measure adherent cells and may circumvent


disturbances of the measurement by cell culture medium components or condensation. An absolute requirement is the possibility to control atmosphere and temperature within the measurement chamber of the microplate reader. To keep human cell lines viable for the duration of the experiment, they need


to be incubated at 37°C and at a CO2 concentration of 5% to maintain the physiologic pH provided by a bicarbonate buffer. Since the oxygen available in human tissues is typically below ambient, it is further recommended to run real-time cellular assays at defined oxygen conditions and thereby mimic physiologic conditions. Te Clariostar multimode plate reader with atmospheric control unit (ACU) was


The Clariostar plate reader with ACU


designed to serve as an ideal platform for real-time cell based assays. Hence, it regulates temperature and atmosphere according to researcher’s needs and adds temperature and gas control data to the measurement results. Te choice between top or bottom measurement is done by a simple click in the plate reader’s software.


Assay components must not be cytotoxic and should not interfere with the analysed process itself. Te method must be stable for the time of analysis and at temperatures of 37°C, the temperature human and other mammalian cells are mostly cultured at. If the cellular reaction is reported by a change in fluorescence, the fluorophore needs to be resistant to bleaching and ideally emits light rather in the red range of the visible spectrum in order to be distinguishable from cellular autofluorescence.


An example for a cellular assay measured


in real-time is the monitoring of two types of cell death: apoptosis and necrosis. Although cell death is an important, normal consequence of multicellular homeostasis it also has an unfortunate role in many human diseases. Discovery of targeted cytotoxic and cytoprotective agents may therefore have important implications in treating cancer, neurodegeneration and a litany of other diseases. To fully appreciate the dysregulation in a disease model, and responses to experimental stimuli, it is important establish the mechanism of action of cell death and define response kinetics. Traditional experimental approaches for addressing mode of cell death are often time consuming, labour-intensive, costly and inadequate. A novel assay (RealTime-Glo Annexin


for the cells. Te combination allows walking away once the test is set up and ensures to record important cellular responses regardless


V Apoptosis and Necrosis Assay, Promega) provides the simplicity of an add-mix- measure assay, in a live-cell, real-time format. Te Clariostar with ACU provides the sensitive detection of duplexed luminescent and fluorescent signals, all while maintaining the appropriate [O2


] and [CO2 www.scientistlive.com


] environment


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84