24
A researcher may choose fluorescence spectroscopy to take advantage of the specificity of the technique. When a UV-Vis absorption spectrum of a mixture is analysed, all the molecules in the mixture that absorb light will contribute to the absorbance observed. With fluorescence spectroscopy, generally the excitation wavelength of a specific molecule can be used so that the emission profile that is observed is specific for that molecule.
The benefit of UV-Vis spectroscopy is in the low sample preparation requirement, particularly when analysing a pure sample, such as a purified protein. Many fluorescence assays exist for detecting protein concentration or amount, but these generally rely on the use of a fluorescent dye and can be inaccurate. UV-Vis spectroscopy is preferred for analysing pure proteins as it is a simple and highly accurate technique. UV-Vis spectroscopy can also measure higher concentration samples that might not be measurable using fluorescence due to quenching or inner filter effects.
Synergy: Combining UV-Vis and fluorescence
In many biological laboratories a UV-Vis spectrophotometer will be found next to a Fluorescence spectrophotometer as the techniques are complementary. The combination of the two techniques can provide greater insight for many biological applications and combining detection modes into a single instrument provides researchers with some efficiency gains especially with high throughput applications where microplate readers are used. Multimodal plate readers provide the ability to employ diverse labelling strategies around a particular application area. For example, in the realm of studying kinases, a hybrid microplate reader combing both UV-Vis absorbance and fluorescence (as well as other detection modalities) can be used.
While UV-Vis spectroscopy is ideal for quantitative analysis or characterising enzymes, fluorescence can provide additional information on how molecules interact. Two examples of this are polarisation assays providing fast and quantitative information on molecular interactions, and fluorescence resonance energy transfer (FRET) assays that also take advantage of fluorescent properties of molecules to analyse interactions.
If using the techniques together, the main application and benefit is in the optimisation of both methods. When developing methods, the combined use of UV-Vis and Fluorescence spectroscopy can ensure optimal results. For example, UV-Vis spectroscopy can be used to confirm an excitation wavelength or the concentration of a molecule before it is analysed by fluorescence spectroscopy so that useful data is obtained.
Using the techniques together can also benefit sample optimisation. For lower concentration samples or if the sample is scarce, fluorescence will be a more sensitive technique that requires less initial sample. However, not every molecule fluoresces, and some fluorescence assays can be cost-prohibitive so the simplicity and low sample preparation requirements of UV-Vis spectroscopy may be preferred.
Practical tips for researchers
Sample preparation matters UV-Vis:
• Ensure sample purity: Contaminants can interfere with absorption measurements. Purify your samples to minimise unwanted signals.
• Use appropriate cuvettes: Choose quartz cuvettes for UV measurements and glass cuvettes for visible range.
• Blank correction: Always measure a blank (solvent) to subtract its absorbance from sample measurements.
Fluorescence:
• Avoid auto-fluorescent materials: Some plastics, glass, or impurities emit fluorescence. Use low-fluorescence materials for cuvettes and containers.
• Dilution effects: High concentrations can lead to inner filter effects or quenching. Optimise sample concentrations for accurate results.
Instrument calibration and validation
UV-Vis: • Calibrate wavelength accuracy: Regularly check the instrument’s wavelength calibration using standard solutions (e.g., holmium oxide filter).
• Validate linearity: Measure known concentrations of standard solutions to verify linearity within the Beer-Lambert range [2].
Fluorescence:
• Check lamp intensity: Fluorescence lamps degrade over time. Monitor and replace them as needed.
• Validate sensitivity: Use known fluorophores (e.g., quinine sulphate) to validate sensitivity and linearity.
Spectral overlap and multicomponent analysis
UV-Vis:
• Be aware of overlapping spectra: Some molecules have similar absorption profiles. Deconvolute overlapping peaks using mathematical methods.
• Use derivative spectroscopy: Derivative spectra enhance peak resolution, making it easier to identify components.
Fluorescence:
• Spectral unmixing: When multiple fluorophores are present, use spectral unmixing algorithms to separate their contributions.
• Lifetime-based analysis: Fluorescence lifetime measurements provide additional information beyond intensity.
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80 |
Page 81 |
Page 82 |
Page 83 |
Page 84 |
Page 85 |
Page 86 |
Page 87 |
Page 88 |
Page 89 |
Page 90 |
Page 91 |
Page 92 |
Page 93 |
Page 94 |
Page 95 |
Page 96 |
Page 97 |
Page 98 |
Page 99 |
Page 100 |
Page 101 |
Page 102 |
Page 103 |
Page 104 |
Page 105 |
Page 106 |
Page 107 |
Page 108 |
Page 109 |
Page 110 |
Page 111 |
Page 112 |
Page 113 |
Page 114 |
Page 115 |
Page 116 |
Page 117 |
Page 118 |
Page 119 |
Page 120 |
Page 121 |
Page 122 |
Page 123 |
Page 124 |
Page 125 |
Page 126 |
Page 127 |
Page 128 |
Page 129 |
Page 130 |
Page 131 |
Page 132 |
Page 133 |
Page 134 |
Page 135 |
Page 136 |
Page 137 |
Page 138 |
Page 139 |
Page 140 |
Page 141 |
Page 142 |
Page 143 |
Page 144 |
Page 145 |
Page 146 |
Page 147 |
Page 148