COMBINATION OF OFCEAS SPECTROSCOPY AND LOW PRESSURE SAMPLING, AN ADVANTAGE FOR LOW CONCENTRATION MEASUREMENTS FOR CEMS
OFCEAS or Optical Feedback Cavity Enhanced Absorption Spectroscopy, is a gas analysis technology developed by the University Joseph Fourier. OFCEAS essentially differs from the older technology enhanced-cavity TDLAS by its feedback principle: a part of the emitted radiation is returned from the chamber to the laser, enabling the tuning of the laser and the cavity, creating a resonance phenomenon. The measuring cavity in which the sample is analysed is equipped with high refl ectivity mirrors (>99,99%), providing an optical path up to 20 km, in a small 40cm cell. The consequence of this phenomenon is the identifi cation of intense absorption peaks with narrow spectral width. The system presents very high measurement stability: there is no zero and no span drift and no need for regular calibrations.
LPS or Low Pressure Sampling, is a patented sampling technique. The pressure is maintained under 100mbar absolute from the sampling point to the exhaust of the analyser. Given that the transfer time is reduced, the dew point is decreased below ambient temperature, the interferences between absorption picks is minimised.
The combination of OFCEAS and LPS was implemented in gas analysers. At the beginning the goal was to use it for research purposes by the University Joseph Fourier, to measure very low concentration of gases in ambient air. But quickly the technology attracted the interest of the industry, and the technology is now used worldwide in CEMs and process applications.
For Continuous Emission Monitoring, OFCEAS and LPS provide many advantages.
OFCEAS allows advanced levels of detection, and measurement range can be from parts per billion to percentage. It can measure standard gas species (O2
, NOx , SO2 , CO, CO2 , H2 with growing interest in emission (HCHO, H2 S, HF, HCl, CH4
O), but also gases , NH3)
even at low concentrations. The technology is also self-referenced, and not deviating system, so there is no need of regular calibration. A direct measurement of all gas species is made every 100msec, without calculation treatment.
LPS (Low Pressure Sampling) below 100mbar absolute, reduces the dew point in the sampling which is a critical point for wet gases like in CEMS conditions. For CEMS, lines don’t need to have a regular heating at 180°C, LPS will allow usage of lines traced at 40°C to 80°C, and will avoid problems of cold spots in winter. The low pressure also has an impact on the absorption spectrums. The lower the pressure is, the thinner are the absorption picks, which allows differentiation of gas species in a same spectrum. The cross-interferences becomes very limited, and trace measurement becomes possible even with high concentration of CO, CO2 hydrocarbons, and H2
, O. Finally, the low pressure uses limited
amount of sampled gas, around 20L/h for CEMs. A low fl ow of emission gas sampled also means a low amount of particles
IET JANUARY / FEBRUARY 2022
sampled. The accumulation of dust in the fi lters of the analyser is reduced by ten, and the cost of spare parts is reduced.
The applications for continuous emissions monitoring are numerous, but the main advantages foreseen are:
- Waste incineration: Waste incineration requires the monitoring of certain compounds. In particular: HF, HCl, NH3
, and NOx .
And the emission limits for those compounds are expected to be lowered, according to the direction given by the BREF. OFCEAS is able to measure these gases in ppb concentrations, way below current thresholds. Which might not be the case for older technologies (NDIR, FTIR). OFCEAS might in this case be the only current technology able to reach those low certifi ed ranges in the near future.
- Gas turbines & wood panels: OFCEAS is also able to measure formaldehyde in fl ue gas, as low as 1ppb limit of detection. Formaldehyde measurement is required by the EPA on combustion turbines, to demonstrate that formaldehyde emissions do not exceed 91ppb, dry basis, corrected to 15% O2
.
The most common solution to measure formaldehyde that low has been to use a modifi ed FTIR with extended cavity. OFCEAS could be a more affordable alternative, due to its 20km path length. The wood panel industry is also interested to control formaldehyde emissions, and could also be interested to implement it.
OFCEAS could be also of interest for gas turbines when it comes to low CO measurements, as its limit of detection for CO could be as low as 1ppb.
- Refi nery and petrochemical plants: analysers combining OFCEAS and LPS are easily implementable in ATEX enclosures. The use of 80°C sampling lines, instead of 180°C standard lines, also reduces the problem of temperature classes in explosive areas. OFCEAS is also able to measure O2
with a mid-IR laser, when
FTIR cannot and would require a secondary zirconia sensor, which is not easily implementable in explosive areas due to its high temperature. Therefore, an ATEX integration of the
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56