16 ANALYTICAL AND LABORATORY EQUIPMENT


COMPUTER CONTROL interfacing


Gary Stevens explains what you need to know when automating lab IT systems


S


ooner or later that manual fluid control system you have been developing will need to be automated and the different options for doing so are numerous, to say the least. Whether your priorities are cost, simplicity, convenience or you simply haven’t decided yet, the starting place is establishing what will you be able to do yourself and the size of your budget. Te range of options for automating systems can be anything from buying a fully integrated laboratory automation instrument costing many thousands of pounds to buying a single-board computer and programming it from the ground up for the cost of the change in your pocket. In the middle ground there are various programmable logic controllers (PLCs) and computer interfaces with varying options for ancillary


modules to drive outputs or read inputs. If, like many research and


development professionals, you have a basic understanding of programming languages, then a simple computer interface can be the answer to manipulating the less demanding control systems. Simple on/off control of valves and pumps can be achieved at a modest cost and choosing how you want to program the device is a bonus, whether it is in Visual Basic, Visual C, C++ or Pascal. Should programming not be an option then there are a number of manufacturers that supply PLCs and computer interfaces using their own proprietary software. Note that some software is an additional cost to the hardware and must be purchased on a ‘per installation’ basis, which can prove to be expensive. Many of the said devices also need Input/Output modules to operate external devices. Somewhere along the line,


if a PLC or computer interface is to be used, connecting it to a computer is almost a certainty and the ability to use a variety of communications options


is useful. Te most common connection is USB and it is the easiest to set up, however, other options such as RS-232, RS-485 and other protocols such as MODBus and CANBus are now also commonplace. One very simple computer


interface is the CIB that has 16 I/O that can be configured as 16 Inputs, 16 Outputs or 8 Inputs and 8 Outputs. Te device is USB as standard but also has a RS-485 on board. Te one feature that sets this


apart from similar products is the method of sending data to the board. Once communication has been opened, all that is needed to switch an output on is to send an uppercase letter corresponding to an output, A for 1, B for 2, etc., and to turn it off, a lowercase letter is sent. Tis methodology enables a multitude of software applications to communicate with the board including terminal emulation programs using simple text-based commands. Te outputs from the CIB


are at 5V logic level and as such trigger external components such as valve drivers and relays to operate other devices. A range of drivers is available on a multi- board module if required.


Outputs from the CIB trigger external components to aid automation processes


www.scientistlive.com For more information ✔ at www.scientistlive.com/eurolab


The CIB is a simple computer interface with 16 I/O


Gary Stevens is managing director of NResearch. www.nresearch.com


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84