48 CHROMATOGRAPHY


Stability even under tough conditions


Fredrik Lime & Cecilia Mazza discuss the efficient isolation of main compounds and impurities using pH.


A


s businesses strive to analyse and purify more samples during the same


timeframe, the turnaround time becomes an area of focus in the laboratory and in manufacturing of natural products and synthetic drugs.


With this mindset, there is also an increased expectation on the flexibility of the stationary phase’s operating conditions, column lifetime and the reproducibility of the results. In addition, when a company works on a new drug candidate, being able to use the same stationary phase and scale the methods accordingly throughout discovery, development and production is a unique opportunity to minimise uncertainties and to save project time.


Fig. 2. Chromatographic result of the purification of crude Tryptamine.


Consequently, laboratories participating in drug discovery and commercialisation projects favour the use of a stationary phase that is available in a wide range of particle sizes to improve productivity across the entire company.


For regular silica-based stationary phases, exposure to extreme pH, especially basic, has a negative impact on the chemical stability, and therefore column lifetime.


www.scientistlive.com


However, the silica/organosilane surface of the Kromasil Eternity platform offers a chemical stability with long-lasting stationary phase, even under tough pH conditions and higher temperatures.


Substances with ionisable groups can exhibit retention time differences under reversed-phase chromatography depending on the degree of ionisation of the compounds and their interactions with the stationary phase under given mobile phase composition. Hence, by changing pH, selectivity between ionisable substances can be altered so that resolution is optimised for a given separation and purification.


As in many cases pharmaceuticals are basic, they are ionised under low or neutral pH conditions, resulting in low retention, poor loadability and broad peaks. Being able to run at high pH means basic compounds become


Fig. 1. Chromatographic results for testing stationary phase loadability with diphenhydramine.


more retained and peaks are narrower, revealing higher chances for better resolution and loadability.


In the isolation of compounds for characterisation of substances, Phase I, II or III studies as well as production, loadability is a key parameter. Loadability of the base diphenhydramine was tested using EternityXT C18. Fig. 1. shows the chromatographic results for diphenhydramine at low and high pH. At low pH of 3.7, the molecule is ionised, leading to a large band broadening even at very low loadings.


Te same loading at high pH of 10.5 (in the upper right chromatogram) produces a sharp peak without any tendency to broaden as a function of concentration overload. To


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84  |  Page 85  |  Page 86  |  Page 87  |  Page 88  |  Page 89  |  Page 90  |  Page 91  |  Page 92