This page contains a Flash digital edition of a book.
16 ANALYTICAL AND LABORATORY EQUIPMENT


Smart instrument engineering D


Darren Broom explains how to extend the operational range of automated dynamic vapour sorption analysers.


ynamic vapour sorption (DVS) typically involves the measurement


of moisture uptake at near ambient temperature in the relative humidity (RH) range 2 to 98%. Applications include the determination of transport rates in food and packaging, and amorphous phase detection and quantification, hydrate analysis and hygroscopicity studies of pharmaceuticals. In some fields, there are significant advantages to the measurement of moisture sorption outside the above regime; however, this adds additional complications.


In DVS, water vapour is usually delivered to the sample by mixing wet (saturated) and dry gas streams to a varying extent. At low RHs the problem is the control of the flow rate. DVS analysers invariably operate using thermal mass flow controllers (MFCs), which provide very stable control but


have limitations both in absolute accuracy and the control of low flow rates, typically below 2% of their full-scale range. Tis corresponds to 2% RH if the flow ranges are matched. MFCs of different ranges can be used but this in turn limits the achievable RH range unless additional flow streams are added, which increases instrument cost and complexity.


Te difficulties at high RH originate from the problem of bulk condensation. A high level of temperature control is required to prevent this occurring near 100% RH because relatively small changes in the temperature can result in significant shifts in the saturation vapour pressure – or the partial pressure at which bulk condensation will occur. See the inset of Fig. 1, which shows the large shift at 50°C caused by an error of ± 1°C. Tis sensitivity poses significant challenges to designers of DVS instruments.


With regard to temperature, the saturated stream is typically generated using a reservoir held at the same temperature as the sample. Tis helps maintain a stable, isothermal environment but places limitations on the achievable measurement range. Te minimum is typically 5°C because temperatures lower than 0°C will result in freezing while the boiling point of water defines the maximum.


Overcoming challenges Each of these difficulties, however, can be overcome by appropriate instrument engineering. For high temperature measurements, for example, the sample can be independently heated. Tis approach is used in the Hiden Isochema IGAsorp-HT, a DVS analyser that can perform measurements up to 300°C (see Fig. 1). Tis configuration also allows control below 1% RH at ambient temperature.


Fig. 1. Water sorption isotherms for chabazite at elevated temperatures and (inset) the sensitivity of RH to small changes in temperature, illustrated by the example of a ± 1°C error at high RH at 50°C.


www.scientistlive.com


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84  |  Page 85  |  Page 86  |  Page 87  |  Page 88  |  Page 89  |  Page 90  |  Page 91  |  Page 92