EXTERNAL QUALITY ASSESSMENT
concentrations, with CV of 14% and 30% at 8 kPa; and 10% and >50% at 4 kPa for the haemolysate and aqueous materials respectively, consistent with improved saturation characteristics of the haemolysate material.
A wider bias, exceeding the allowable APS, was observed for the IRMA and NPT7 analysers in 2007, with wider inter-laboratory variation observed for the OPTi, ABL6 and Rapid Point 400 analysers in 2007 and the iSTAT, OPTi, and Prime+ in 2021. The latest pO2
accuracy study
(OX1124) was distributed to over 3,300 instruments in November 2024 and results compared with a concentration matched aqueous sample from BG1024. Figure 3 illustrates the intra-method variation (mean ± 2 SD) shown as an error bar for each analyser. The analytical performance specification is expressed as the Weqas TAE (Total Allowable Error) of
Analyser type
Conventional electrodes Multi-use cartridge Optical sensor
Single-use cartridge % POCT analysers Total
2001 505 14 13 11 7
543
2006 853 239 34 84
29.5 1210
2010 947 541 39
137 43.1 1664
Year
2016 541
1033 28
269 71.1 1871
2020 244
1440 16
338 88
2038 Table 1. Change in blood gas analyser type and growth of POCT analysers from 2001 to 2023.
±26.5% at 2.56 kPa and ±34.1% at 6.40 kPa for the pO2
accuracy and aqueous
samples respectively. In keeping with previous studies, there was a general improvement in method bias to the all method average for most analysers for the bovine haemolysate sample, compared with the aqueous samples. A significant positive bias compared to the all-method average was
observed for the Omni/Omni S/b221, Opti, i-Stat, and i-Stat Alinity method groups for the aqueous sample, whereas the data agreed well with the all-method average for the bovine haemolysate sample. Significant negative bias compared to the all-method average was observed for the ABL90 flex and Prime+ method groups for the aqueous sample, but whereas the ABL90 flex data agreed well with the all-method average for the bovine haemolysate sample, the negative bias remained for the Prime+ method. The overall CV for pO2
was lower for
the bovine haemolysate sample (overall CV 12.9%) compared to the aqueous sample (overall CV 48.9%). The overall CV includes components of both imprecision and method bias, with the large deviation in bias observed in the aqueous sample contributing to the wider overall CV. Although a higher CV was observed for some methods for the bovine haemolysate sample, the concentration of pO2
2023 127
1876 13
427 94.8 2443
in the two samples were not exactly concentration matched, with overall mean 2.56 kPa compared with 6.4 kPa for the aqueous sample. However, apart from the Prime+ method, most methods performed well at this very low pO2
concentration, confirming that pO2
methods are accurate across the clinical range required.
Conclusions Fig 3. a) pO2 40 results for blood gas distribution BG1024 (aqueous sample); b) pO2 distribution OX1124 (tonometered bovine haemolysate). results for pO2 accuracy
In conclusion, arterial blood gas analysis is a commonly used technique in a range of emergency and critical clinical situations, as well as in assessment of patients with chronic disease. Improvements in technology have expanded the use of POCT blood gas devices into multiple clinical areas, including intensive care units, emergency departments and ambulances. Assessment of patient oxygen status is an immediate concern for clinical teams, and EQA is a key component of ensuring that devices provide accurate results that are suitable for clinical use. Previous work has shown that aqueous samples used for blood gas analysis in EQA programmes allow participants to be compared with their peer group, but
SEPTEMBER 2025
WWW.PATHOLOGYINPRACTICE.COM
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80 |
Page 81 |
Page 82 |
Page 83 |
Page 84