12 August / September 2019
Figure 2: LC-MS method : 5-95% MeCN in 1.5 min and holding at 95% for 0.5 min, where the organic solvent is MeCN and the aqueous solvent is 10 mM ammonium carbonate pH 9; flow rate is 1.2 ml/min; temperature is 50 °C; stationary phase is the Waters XBridge C18 column with dimensions 50 mm x 2.1 mm i.d. (3.5 µm).
Analytical SFC-MS method : 5-55% methanol with ammonia (20 mM) in 1.5 min and holding at 55% for 0.5 min; flow rate is 3 ml/min; temperature is 55 °C; back pressure is 105 bar; stationary phase is the Naphthyl column (sample A) and the 1-AA column (samples B and C), with dimensions 100 mm x 3.0 mm i.d. (5.0 µm).
Preparative SFC-MS method: tailored gradient program # 5 (see table 1) for samples A, B and C; methanol with ammonia (40 mM) is used as the SFC modifier; flow rate is 100 ml/min; temperature is 55 °C; back pressure is 120 bar; column type is as stated analytically, with dimensions 150 mm x 30.0 mm i.d. (5.0 µm).
quantities (less than 200 mg) are normally processed and fractions dried within 90 minutes. Although SFC methods are much shorter compared to HPLC, the average loading per injection is generally lower. For the medicinal chemist, the overall speed benefit of SFC stems from the collection of target compound in small volumes of methanol, and the much-reduced time for solvent evaporation compared to reverse phase HPLC. It is important to note however, the different level of oversight required across the two techniques to achieve the robustness required in a walk-up environment. In our experience, the success of open access preparative SFC-MS depends on the availability of an analytical chemist or super-user, to partner with submitters and to help monitor the performance of instruments across the working day.
Acknowledgement
Waters Corporation for their open access software enhancements.
Figure 2 illustrates the benefit of screening purification samples by SFC, to compliment traditional reverse phase LC-MS. Sample ‘A’ delivers one peak by HPLC, but separates into three components by SFC. Sample ‘B’ is an example of a more polar reaction mixture. Poor retention and separation is achieved by HPLC, whereas adequate retention and very good separation is achieved by SFC. In this example, favourable solubility and separation enabled the injection of 830 mg of material. Sample ‘C’ illustrates an example where both HPLC and SFC separate
two regio-isomers. SFC is the obvious choice for purification due to the resolution of impurities from the target compound.
Conclusion
SFC-MS is now embedded as a primary tool for front line purification of reaction mixtures. For samples in the region of 200 mg to 1.5 g the median cycle time, from dissolution of the crude material to the isolation of pure solid, is 3 hours. Smaller
References
1. Rossé, G. (Ed.), Berger, T., Francotte, E., et al. (2018). Supercritical Fluid Chromatography. Volume 1. Berlin, Boston: De Gruyter.
2. Rossé, G. (Ed.), Lipka, E., Speybrouck, D., et al. (2018). Supercritical Fluid Chromatography. Volume 2. Berlin, Boston: De Gruyter.
3. Aurigemma, C., Farrell, W., J. Chromatography A, 1217 (2010) 6110-6114.
Widest Range of Innovative SFC Columns
Princeton Chromatography, Inc has been producing industry leading SFC columns and bulk chromatographic media for twenty years. With the largest and most diverse offering of stationary phases on the market for SFC, the company have something to fit nearly every achiral application. From revolutionary 2-Ethylpyridine phase pioneered in 1998, to the most current and cutting edge chemistries, Princeton Chromatography can provide selectivity to solve your toughest problems. The HA Series line of columns has given rise to the newest and currently most popular column in our inventory. HA- Morpholine has emerged as a strong performer on screening systems at several major R&D sites around the world. With direct scalability to prep, HA-Morpholine has been responsible for the purification of thousands of compounds since its release. Other HA Series phases of note are HA-Dipyridyl and HA-Diethylamino, both with unique
selectivity. In addition to the HA Series, we are proud to offer PPU (Propylpyridylurea), which is unique to Princeton Chromatography and in use on many SFC systems worldwide.
More information online:
ilmt.co/PL/64wE
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68