8 Analyse new and used Motor and Hydraulic oils


FTIR absorption spectroscopy is an ideal analytical technique to characterise new, used and diluted motor oils, as well various other samples. Service oils, like Castrol’s Magnatec engine oil, are designed to minimise wear-and-tear in engines. After a while, the oils pick up elemental metals, soot and other contaminants which turn the fresh yellow oil into a dark and worn black. IR spectral analysis can help determine which contaminants are present, as well as what kind of chemical degradation occurs in the oil.


Traditionally, a vertically mounted cell was always the standard method of loading viscous or sticky fluids, which posed many problems including sample introduction and clean up. However, The PearlTM


liquid transmission accessory is designed to make the IR spectral measurement of liquids quick, easy and highly repeatable. Equipment and Samples The PearlTM accessory (P/N GS31000-B) was used in conjunction with fully-wedged ZnSe windows in the Oyster Cell.


Three different pathlengths were used; 25, 50 and 100 microns. The oil samples tested included a fresh sample of Castrol Magnatec, as well as a used sample after 18 months of use in a petrol engine. Fresh Renolin CL37 and 68 oils were also analysed. The spectra were recorded with a 1 cm-1


resolution, using The PearlTM coupled with an FTIR spectrometer.


The Results Figure 1 shows three spectra for a fresh sample of Castrol Magnatec motor oil, recorded using three different pathlengths: 25, 50 and 100 µm. The absorbance increases with increasing pathlengths in the Oyster Cell, as can be expected. It is worth noting that the oil samples had not been diluted, so they were at their maximum concentration during analysis. Despite this, the maximum absorbance in fingerprint region (500 – 1500 cm-1


at 2.5 for the 100 µm pathlength. This demonstrates that no dilution or sample preparation is needed to record the transmission spectrum of oil samples with 25, 50 or 100 µm pathlength windows. Figure 2 compares the spectra of fresh and used Castrol Magnatec oil samples using a 25 µm pathlength.


Table1: Pathlengths & part numbers used.


) peaks


Figure 1: IR spectra of fresh Castrol Magnatec.


Figure 2: A comparison of new & used Castrol Magnatecs.


Very little sample, < 5 ml, was needed to quickly differentiate the oils. The used sample shows evidence that the oil has been chemically altered and that contaminants have been introduced. Around 1150 cm-1


, the hallmark signs of sulphation are visible, while the region of 1600 – 1700 cm-1 Conversely, there is a loss of relative signal intensity at 1500 and 1460 cm-1 .


This is marked by blue circles and indicates there is less of the original oil in the used sample. Figure 3 compares three different fresh oil samples measured using a 25 µm pathlength, in black, red and blue, respectively:


• Castrol Magnetec engine oil • Renolin CL 37 hydraulic oil • CL 68 hydraulic oil


Clearly, the oils share a similar aliphatic band at 1470 cm-1 , but their other molecular signatures are different.


Conclusion FTIR spectroscopy is a well-known tool for the characterisation of various samples. This note highlights that viscous oil samples can be analysed simply and quickly using three different pathlengths in The PearlTM Very little sample was needed to differentiate the oil samples.


Moreover, the used samples showed evidence that contaminants had been introduced over time through use and contact with machinery.


shows evidence of oxidation and nitration in the oil.


. Figure 3: A comparison of the three oils.


The subtle spectral signatures between different oils, whether fresh or used shed light on the chemistry of the samples. A loss of signal intensity at certain frequencies indicated that the original oil sample had been chemically degraded through use in a car engine. The fact that the maximum absorbance in the fingerprint region (500 – 1500 cm-1


transmission spectra. More in-depth studies to determine how quickly the oils degrade with use can easily be carried out using The Specac PearlTM


Email sales@specac.com or visit www.specac.com for a free demonstration or more information. . 40966pr@reply-direct.com


) peaked at 2.5 for the 100 µm pathlength shows that a few drops of pure sample is enough to record top quality


LAB ASIA - NOVEMBER/DECEMBER 2016


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44