Guide to USING PT100 SENSORS T


Ken Jefferies reports on sophisticated sensors for precision thermometry


he use of a platinum resistance thermometer probe in industrial


and scientific applications will normally result in good, accurate temperature sensing without the need for special cables (unlike thermocouples). To achieve the best possible accuracy and reliability, care is needed with installation of the probe, connections and choice of the host instrumentation.


wire wound version is fused into ceramic or glass. With regard to wire-wound


resistors, various methods of detector construction are employed to meet the requirements of differing applications. Te unsupported ‘bird cage’ construction is used for temperature standards, and the partially supported construction is used where a compromise is acceptable between primary standards and use in industrial applications. Other constructional methods include the totally supported construction, which can normally withstand vibration levels to 100g. Meanwhile, metal film Pt


Flat film Pt100 sensing resistors


Choice of wire-wound or flat film sensing resistor Te resistance element used in the probe is produced in one of two forms, either wire-wound or flat film. Metal film resistors consist of a platinum layer on a ceramic substrate; the coil of a


resistors take the form of a thin (1 micron) film of platinum on a ceramic substrate. Te film is laser trimmed to have a precise Ro value and then encapsulated in glass for protection. Terminating the resistance thermometer is a key issue to be considered. Fundamentally, every sensing resistor is a two- wire device. When terminating the resistor with extension wires during probe construction, a decision must be made as to whether a two-, three- or four- wire arrangement is required for measurement purposes. In the sensing resistor, the


Wire-wound sensing resistors


electrical resistance varies with temperature. Temperature is measured indirectly by reading the voltage drop across the resistor in the presence of a constant current flowing through it using Ohm’s Law: V = R.I Te connection between the thermometer assembly and the instrumentation is made with standard electrical cable with copper conductors in two, three or four core construction. Te cabling introduces electrical resistance that is placed in series with the resistance thermometer. Te two resistances are therefore


cumulative and could be interpreted as an increased temperature if the lead resistance is not allowed for. Instruments using three


or four wire configurations reduce or eliminate lead resistance effects. However, the wiring configuration of the thermometer must be compatible with the input to the associated instrument.


High accuracy options (tolerance classes) Assuming a three- or four-wire connection, and the use of a class B sensing resistor, a standard thermometer assembly will provide an accuracy of around 0.5°C between 0°C and 100°C. Considerable improvement on this figure can be achieved by various means including the use of closer tolerance sensors. Te principle of operation is the non-linear variation of the electrical resistance of the resistor with temperature. Te commonly used Pt100 has a resistance of 100.00Ohms at 0°C and 138.51Ohms at 100°C according to IEC751. Te use of virtually pure


rare metal (platinum) in the sensing resistor endows this type of temperature sensor with both high accuracy and long term stability; good quality instrumentation is therefore essential to realise these virtues. It is important to remember


that the overall accuracy of any measuring system (e.g. sensor, instrument, interconnection, application, etc) is compromised by the sum of the uncertainties in that system.


For more information ✔ at www.scientistlive.com/eurolab


Ken Jefferies is technical consultant to Labfacility. www.labfacility.com


www.scientistlive.com


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84  |  Page 85  |  Page 86  |  Page 87  |  Page 88  |  Page 89  |  Page 90  |  Page 91  |  Page 92  |  Page 93  |  Page 94  |  Page 95  |  Page 96  |  Page 97  |  Page 98  |  Page 99  |  Page 100