Mass Spectrometry & Spectroscopy 17 Metrohm Offers Complete Solutions for Oil Condition Monitoring


Monitoring the quality of hydraulic and lube oils is critical to safeguard the availability and prevent premature wear of expensive machinery. Metrohm offers complete solutions for the fast and accurate analysis of key parameters such as moisture, TAN and TBN, remaining antioxidant content, and many others. These parameters can be determined by Karl Fischer titration, titration, near-infrared spectroscopy, and voltammetry, respectively – established techniques that are widely used by QC laboratories in the petrochemical sector and by contract laboratories offering analytical services to the power industry, manufacturers of construction equipment, compressors, or turbines, to name but a few.


Moisture content: automated thermal sample preparation for coulometric Karl Fischer titration


This method is the preferred solution for samples that release the contained moisture only at high temperatures, that are difficult to dissolve, or react with the KF reagent. With this method, the sample matrix is heated up to release the contained moisture, which is then determined by coulometric Karl Fischer titration in a separate reaction vessel: Hence, there are no matrix effects, there is no contamination of the titration cell, and accurate and reproducible results are guaranteed. A blog article explains the features and benefits of this method in detail: ilmt.co/PL/X1Zk


Remaining antioxidant content: voltammetric determination


Testing of in-service lubricants for their remaining antioxidant content is critical for prolonging the uptime of capital equipment as well as reducing running costs and repair expenses. Voltammetry is a fast and established technique for testing the remaining antioxidant content in industrial lubricants. The principle of this method is that the aromatic amines and hindered phenols (which are used as primary antioxidants) in the lubricant sample are extracted into an electrolyte. They are then measured directly in this extraction solution. Voltammetric determination is possible because the hindered phenols as well as aromatic amines contain a functional group that can be electrochemically oxidised. More info about this application can be found in this white paper: ilmt.co/PL/8NnB


Total Acid Number (TAN) in used engine oil: thermometric titration as per ASTM D8045


Heavily contaminated samples, e.g., used engine or turbine oils, can quickly blog a potentiometric sensor making frequent cleaning and conditioning a must. For such media, thermometric titration can be a better solution. Thermometric titration is the «rugged sibling» of potentiometric titration and can be used for any titration that triggers a change in the temperature of the sample solution during the titration reaction.


How does thermometric titration work? Instead of an electrochemical sensor, thermometric titration follows the titration reaction utilising a thermometer with a very short response time and high resolution. When all the analyte has been reacted with the sample, this sensor registers a minute difference in the rate of temperature change indicating the endpoint of the endo- or exothermic titration reaction. This sensor does not require any maintenance, nor does it need any conditioning, it can be stored dry, and contamination from solid matter in the sample is not an issue since there is no electrode diaphragm. Used oils contaminated with particles are a typical kind of sample that can be analysed by thermometric titration. More info about thermometric titration and more specifically Acid Number as per ASTM D8045 can be found in this blog article: ilmt.co/PL/Zen5


Multiparameter analysis of lubricants by near-infrared spectroscopy (NIRS)


For lubricant analysis, determination of Acid Number (ASTM D664), viscosity (ASTM D445), moisture content (ASTM D6304), and colour number (ASTM D1500) require the use of multiple analytical technologies and, in part, considerable volumes of solvents and reagents.


NIRS is a non-destructive analytical technology where the sample is investigated as is with no prior sample preparation and chemicals required for analysis. It is faster too than competing technologies and provides results for several physical and chemical QC parameters within a minute. Thus, Acid Number, Base Number, kinematic viscosity, viscosity index, density, colour number, and moisture number can all be determined simultaneously by NIRS, as is described in this blog article.


More information online: ilmt.co/PL/Venn 60499pr@reply-direct.com


ADVERTORIAL


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56