63
a change in the detector. If the compound changes in its form then the absorbance coefficient will also change which means the sensitivity of the assay will be affected. In the extreme example, the absorbance could be reduced by so much that the compound is no longer detected within the UV detector. The increase in biopharmaceutical compounds has highlighted this as an ever-increasing issue, since proteins can be very susceptible to degradation when exposed to a broad range of the electromagnetic spectrum.
applicability to a wide range of compounds. There are two formats that chromatographers use, one is a variable or fixed wavelength design and the other type is a diode array. The actual difference in terms of the operation is minimal, and the two types of detector are often used interchangeably. For the fixed wavelength detector, a UV/Vis light source is filtered to give the required wavelength, which results in the sample only being exposed to a single wavelength of light. With a diode array detection system, the sample is exposed to a full range of wavelengths, which are then captured by the array detector (Figure 2). This does not seem too significant, but for compounds which are sensitive to light this can be a significant difference resulting in the compound undergoing
The HelpDesk has discussed the importance of considering the effects of changing the instrumentation on the assay performance previously, and it is pertinent to reiterate this discussion. It is also important to consider that it is not just the pumps that can influence the performance of the assay and that the detector can also affect the very molecules that the chromatographer is trying to analyse. If changing the detector results in degradation of the analyte, then the sensitivity may alter, and in worse case situation complete loss of signal may be observed. As with all disciplines it is important not to make assumptions, and this is the case with assay development. It is important that the chromatographer understands how the analyte will be affected in each of the various stages of the analytical process to ensure that a robust assay is developed and that there are minimal issues when this assay is transferred.
References
1. L.R. Snyder, J.J. Kirkland, and J.W. Dolan, Introduction to Modern Liquid Chromatography, 3rd. ed. (Wiley, Hoboken, New Jersey, 2010)
Streamline your GC Analysis with Precision
For labs performing GC analysis, Peak Scientific have created a true standalone gas delivery solution, the Precision Series of hydrogen, nitrogen and zero air gas generators. These generators can bring workflow benefits and convenience to labs, particularly when compared with gas cylinder supply.
Many labs are currently using helium cylinders to provide carrier gas for their GC, primarily because it has been the traditional method of supply. However, by switching carrier gas from helium to hydrogen, labs are getting a more convenient and cost effective gas supply for their GC. With a consistent, on demand supply of hydrogen gas from a gas generator, labs need not be concerned with the hassle and cost of gas cylinder deliveries or the price increases associated with dwindling reserves of helium.
Peak Scientific’s Precision Hydrogen gas generators are available in Trace and Standard variants delivering high purity hydrogen gas for GC. The Precision Hydrogen Trace generators come in flow rates ranging from 250cc per minute to 500cc per minute and can be used to deliver both carrier and detector gas whilst the Precision Hydrogen Standard generators deliver from 100cc per minute up to 450cc per minute and are suitable only for detector gas.
With numerous flow rate versions of all Precision gas generator models labs can build a gas solution tailored to their specific GC set up. Engineered with lab space in mind, the Precision Series generators can all be stacked with up to four units at a time, requiring a limited amount of lab floor-space.
For peace of mind in your GC gas supply, you need Precision. For more information visit
www.peakscientific.com
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68