Low-Maintenance Turbidimeter for Potable Water Applications


Higher demand for safe drinking water and tighter monitoring of contaminants in


potable water are in contrast to the current trend towards reducing maintenance / operating costs and remote controlled plants. Main maintenance tasks on a


regular base for turbidimeters are cleaning the optical windows and replacement of the light source as well as readjustment /


recalibration with a primary standard. A new approach with a non-contact, LED turbidimeter reduces maintenance efforts significantly and thereby contributes to lowering the cost of ownership.


State of the art turbidimeters for potable water analysis are based on the nephelometric measurement principle as described in the Standard Methods [1] and adapted by ISO (ISO 7027 [2]) or EPA (EPA 180.1 [3]). A ratio turbidimeter measures the ratio between scattered and transmitted light. If just the trans- mitted light is detected, it is a single beam turbidimeter.


The set up is quite simple but there are some disadvantages, especially in terms of maintenance effort.


Several tasks have to be con- ducted on a regular base for these types of turbidimeters to ensure its reliability and functionality.


Nephelometry


In a nephelometric set-up, turbidity is determined by the measurement of light, scattered at an angle of 90° to the incident beam.


Readings are from a calibration curve established with a primary standard (i.e. Formazine). Results are expressed in FNU (Formazine Nephelometric Units) or NTU (Nephelometric Turbidity Units).


The main maintenance tasks Cleaning of optical windows


The turbidimeters consist of a flow cell design, where the optical windows are in direct contact with the sample. It is obvious, that cleaning is sometimes necessary, which means maintenance hours, causing operating costs. Cleaning the optical windows often requires a readjustment of the turbidimeter’s sensitivity.


Replacement of the light source


A light bulb with a tungsten filament has to be replaced at least once a year which needs a readjustment or recalibration of the turbidimeter. Additionally, a decrease of the lamp intensity leads to wrong measurement values if not monitored correctly


Chamber drain Figure 1: Schematic set-up of a nephelometric turbidimeter


Settled particles in the measurement chamber have to be removed from time to time to avoid interference based on a carry-over effect.


The new SWAN AMI Turbiwell In the design of the new AMI Turbiwell, the focus was on low maintenance:


Hgh precision Secondary Standard for AMI Turbiwell


•Non-contact set-up: The optical windows are not in direct contact with the sample which means there is no fouling and as a consequence, no cleaning is required. The optical windows are heated to prevent condensation.


•Light source: Instead of a “light bulb” with a lifetime of around 8000 operating hours, a light emitting diode (LED) with a lifetime of approximately 100’000 operating hours is used. Therefore, a replacement is not necessary and readjustment or recalibration is redundant. Two different light sources are available which are either ISO 7027 compliant or accepted as an alternative method to EPA 180.1 [4]/[5]


Figure 3: Verification Kit


•Chamber drain: An automatic or manual chamber drain avoids a carry-over effect and contributes to reliable measurement value.


Figure 2: Set-up of the SWAN AMI Turbiwell


The light beam of the LED source falls on the water surface and is refracted.


At an angle of 90°, the detector measures the incoming, scattered light.


The transmitted beam intensity is not measured; therefore, the AMI Turbiwell is a single beam turbidimeter and not a ratio turbidimeter.


The special design of the chamber (barrier and beam trap) prevents measurement errors due to light reflection.


A dark current measurement is completed at least once a day to compensate the influence of temperature variations on the electronics


But how can a single beam turbidimeter compensate for fluctuations in the intensity of the light source?


Measuring principle AMI Turbiwell


“Normalisation” of Turbidity Signals


NTS=I90 / I0 NTS: Normalised turbidity signal


I90: Scattered light intensity I0


: Emitting light intensity


The absence of the transmitted beam intensity measurement and consequently the absence of the ratio for light source fluctuation compensation requires a new/different approach.


Compensation is achieved by normalizing the emitted light intensity. A beam splitter (semi-transparent mirror) within the light source directs a part of the emitted light to a photodiode which registers the emitted beam intensity. Building the ratio of the scattered light intensity to the emitted beam intensity eliminates the fluctuations in the light source.


This “normalisation”, the non-contact set-up, the long-life LED light


source and the regular chamber drain make a recalibration of the turbidimeter with a primary standard unnecessary.


If a system check is required due to local regulations, a dry verification using a high precision secondary standard can be conducted to validate the instrument’s performance.


Annual Buyers Guide 2010


IET


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84  |  Page 85  |  Page 86  |  Page 87  |  Page 88  |  Page 89  |  Page 90  |  Page 91  |  Page 92  |  Page 93  |  Page 94  |  Page 95  |  Page 96  |  Page 97  |  Page 98  |  Page 99  |  Page 100  |  Page 101  |  Page 102  |  Page 103  |  Page 104  |  Page 105  |  Page 106  |  Page 107  |  Page 108  |  Page 109  |  Page 110  |  Page 111  |  Page 112  |  Page 113  |  Page 114  |  Page 115  |  Page 116  |  Page 117  |  Page 118  |  Page 119  |  Page 120  |  Page 121  |  Page 122  |  Page 123  |  Page 124  |  Page 125  |  Page 126  |  Page 127  |  Page 128  |  Page 129  |  Page 130  |  Page 131  |  Page 132  |  Page 133  |  Page 134  |  Page 135  |  Page 136  |  Page 137  |  Page 138  |  Page 139  |  Page 140  |  Page 141  |  Page 142  |  Page 143  |  Page 144  |  Page 145  |  Page 146  |  Page 147  |  Page 148