This page contains a Flash digital edition of a book.
Full-Field MXRF


Figure 2 : Elemental images collected from droplets while drying using an X-ray color camera in TXRF illumination geometry. The droplet consists of two 10 µL droplets and one 20 µL droplet: Mn (violet), Ni (red), and Cu+Sc (green). The X-ray source is located at the top of each picture (shaded regions are indicated by dotted lines). (a) Image taken after 7.5 min drying, (b) image taken after 11.75 min of drying, (c) image taken after 20.5 min, and (d) image of completely dried specimen after 27 min. Measurements were carried out at the Beamline/BESSYII instrument at Helmholtz-Zentrum Berlin.


Figure 4 : Schematic showing a sheet beam exciting a slice of specimen and an energy-dispersive detector detecting elemental signals. Reprinted from [ 5 ] with permission from the Royal Chemical Society.


channels. T e counts acquired for a specifi c element line can be extracted and displayed to show the spatial distribution of this element in an X-ray map. However, these array detectors do not achieve the high count rates of SDD detectors, for example, the SLCam ®


array camera count rates currently are


Figure 3 : Schematic showing multiple dimensions in the acquisition of an XRF spectrum at each pixel. Similar data cubes may be acquired in both scanning and full-fi eld microscopy modes. Reprinted with permission from Oliver Scharf.


detector, using arrays of capillary optics (>200 k capillaries) to guide more fl uorescence photons to the detector array, and improving event processing soſt ware.


Modern full-field array detectors like the SLcam® provide


spectral (energy) resolution of < 160 eV [ 4 ], comparable to single-chip silicon drift detectors (SDDs). At each pixel a full spectrum comprising an energy range of >10 keV is acquired. Figure 3 shows the typical data cube for a spectrum image with a 264×264 array and spectra collected over 1,024


40


only about 22 cps/pixel. The spatial resolution obtained with full-field setups is in the single-digit micrometer range and is comparable to spatial resolutions achieved in scanning mode at second-generation synchrotron facilities. Excitation and detection . In full-field X-ray microscopy, the sample is illuminated with a grazing incident beam, a total reflected beam, or a sheet-like beam [ 5 , 6 ]. A schematic of the setup using a sheet-like X-ray excitation beam from Radkte et al. [ 5 ] is shown in Figure 4 . A 3D image of an object can be obtained by translating the object through the sheet beam, where the in-depth image resolution is given by the thickness of the sheet beam. Elemental specific XRF data can be obtained from this excitation. Semiconductor array detectors such as charge coupled devices (CCDs) are used to achieve an elemental analysis as the object passes through the sheet beam. CCDs are in general energy-dispersive; however, to achieve energy (spectral) resolution comparable to SDDs, sophisticated devices like the silicon-based SLcam ® (developed by PNSensor GmbH, Munich, and IFG, Berlin, with other partners ) have to be used [ 4 ]. Such devices are often referred to as color X-ray cameras, or CXCs, because they provide energy resolution sufficient to discriminate


www.microscopy-today.com • 2015 May


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76