This page contains a Flash digital edition of a book.
Performance of a Silicon Drift Detector


DT at 0.5-ms TC versus ~61-percent DT at 3.2- to 12.8-ms TC). Te indicated input count rate saturates at high probe currents (~80 nA) and saturation depends on the average detected X-ray energy (for example, ~500 kc/s for Mn at 20 kV versus >750 kc/s for Al at 10 kV). Additionally, the indicated input count rate is not proportional to the probe current at high input count rates. It appears that the throughput and indicated input count rate problems are related to problems in the detector and/or processing electronics. Tough the manufacturer is aware of these problems and is currently working on their next generation of signal processing electronics, a prototype was not available for testing at the time of writing. Tough the manufacturer has incor-


Figure 9: BSE image and elemental maps (100 × 128 pixels) for Al, Cr, and Re from CMSX4 superalloy specimen at 15 and 5 kV (top and bottom row, respectively). See text for experimental details.


spatial resolution (the estimated lateral spatial resolution for the elemental maps is ~100 nm at 5 kV).


Discussion Te silicon driſt detector (SDD) tested here provides


both high geometric collection efficiency and high count rate capability for X-ray microanalysis and elemental mapping (spectrum imaging). Testing of this SDD on the JEOL 6500F SEM indicated input count rates ≥500 kc/s for a pure manganese standard. Pulse pile-up effects increased with increasing input count rate and decreasing time constant (which allowed higher count rates). However, it should be noted that the use of a pure Mn standard (or likewise another pure element specimen) is a stringent test for pulse pile-up effects as ~60 percent of the X-rays detected are in the Mn K and Mn L peaks (46 percent K, L 12 percent L). Te intensity of a sum peak is proportional to the count rate for the two peaks being summed. For a given input count rate and specimens with multiple elements present, the total counts are divided among more peaks. Tis reduces the count rate for each of the peaks, thereby reducing the intensity of the sum peaks. Of course, there will be additional sum peaks from the multiple elements. Both the energy resolution and peak energy increased


slightly with input count rate. However, if the analyst uses optimized counting conditions (that is, input counts rates giving less than or equal to the maximum throughput), the increase in FWHM or peak position will be ≤10 eV, respectively. With the current generation of detector/processing electronics, there is a measurable degradation in the energy resolution as the time constant is decreased. It appears that the larger SDD chip is more prone to such effects, but no detailed testing of other SDDs with smaller active areas was undertaken as part of this work. Te throughput depends on time constant; a maximum


throughput of ~215 kc/s is achieved at 0.5-ms TC. Te DT at maximum throughput decreases at short TCs (~49-percent


46


porated a routine for correcting pulse pile-up effects into their latest version of soſtware, that routine does not appear to be correctly optimized for this large SDD (best case removal was ~60 percent of


the counts in the 2× sum peak). Te remaining sum peak was not symmetric. Te current routine simply deletes the counts arising from pulse pile-up effects. Ideally the pile-up counts should be returned to their correct (original) energy.


Conclusion From the results presented above, it is clear that there


are numerous advantages of silicon driſt detectors (SDDs) for SEM-based microanalysis and elemental mapping. Te larger geometric collection effiency, high count rate, and high throughput capabilities and operation without liquid nitrogen offer benefits to the analyst. Peak shiſt with increasing input is minimal for usable input count rates (that is, <70-percent DT). As the detector using an older, large diameter SDD chip, the energy resolution does not match more current SDDs that can resolve <130 eV under optimum conditions.


Acknowledgments Research was supported by the Oak Ridge National


Laboratory’s SHaRE User Facility, which is sponsored by the Scientific User Facilities Division, Office of Basic Energy Sciences, U.S. Department of Energy.


References [1] DE Newbury, Scanning 27 (2005) 227–39. [2] EA Kenik, Microsc Microanal 14 (Supp. 2) (2008) 1172CD.


[3] EA Kenik, Microsc Microanal 15 (Supp. 2), (2009) 560CD. [4] Private communication, R Anderhalt. [5] TS Elam, R Anderhalt, A Sandborg, J Nicolosi, and D Redfern, Microsc Microanal 14 (Supp. 2) (2008) 1260CD.


[6] RP Gardner and L Wielopolski, Nuclear Instruments and Methods 140 (1977) 289–96.


[7] PR Statham, X-Ray Spectrom 6(2) (1977) 94–103. [8] Q Bristow and RG Harrison, Nucl Geophys 5(No.1/2) (1991) 141–86.


www.microscopy-today.com • 2011 May


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76