Electronics Industry Awards – Finalist Supplement


How to test the surface resistivity of semiconductor materials


Many important parameters of electronic devices are closely related to the resistivity and the uniformity of its distribution, such as the reverse saturation current of the diode, the saturation voltage drop of the transistor and the magnification , etc., which are directly related to the resistivity of silicon single crystal. Therefore, the resistivity test of devices has become an important process in chip processing, and its uniformity control and accurate measurement are directly related to whether power devices with better performance can be manufactured in the future.


U


nlike using a multimeter to measure conventional conductor resistance, the measurement of semiconductor silicon single crystal resistivity


and other metal thin film resistivity in the microelectronics field belongs to the micro- zone thin layer resistance test, which needs to use a tiny signal power supply and high precision measurement equipment, including the test wiring method, it also needs to use the four-wire connection method to enhance the accuracy of the measurement results, known in the industry as the four-probe test method.


What is the four-probe test method?


The four-probe technique can be tested on wafers and thin-layer resistances, such as silicon substrate wafers, milled wafers, epitaxial wafers, ion implanted wafers, annealed silicon wafers, metal films and layers. Probe analysis can be used to detect the uniformity of thin layer resistance across the chip surface and thus determine the problems in the ion implantation wafer and the injection process. The four-probe method can be divided into straight four-probe method and square four-probe method according to the measurement shape.


1 Straight four-probe method The principle of the linear four-probe test method is to press four probes with a needle distance of 1 mm on the flat surface of the sample at the same time, use a constant current source to power a small current to the two outer probes, then use a high-precision digital multimeter to measure the voltage on the two middle probes, and finally according to the theoretical formula to calculate the resistivity of the sample, resistivity calculation formula: represents the probe spacing.


, S 40 September 2023


The straight-line four-probe method can measure inhomogeneities in areas more than three times probe spacing.


Test equipment: probe station    


IT2806 High Precision Source Measure Unit: Combining the capabilities of six devices in one (Voltage Source, Current Source, 6 ½ Digital Multimeter (DCV, DCI, ohms), Battery Simulator, electronic load and Pulse Generator). In the resistivity test, the IT2806 high-precision source measure unit can be switched to the constant current source mode, and the small voltage drop between the two middle probes can be measured at the same time as the output current, and the free PV2800 host PC software can be used to automatically obtain the results of resistivity measurements.


Advantages of using IT2800 series  to measure resistivity: Easy testing: ITECH IT2800 series high precision source measure unit (SMU) comes standard with the free PC PV2800 software and an optional probe stage with different linear probe spacing, which can directly produce resistivity test results using the software’s built-in formula. High test accuracy: up to 100nV/10fA resolution, current measurement accuracy up to 0.1%+50pA, voltage measurement accuracy up to 0.015%+300uV; provide forward/reverse current continuous change test to improve test accuracy.


2 Square four-probe method  The Vanderbilt method is suitable for


Components in Electronics www.cieonline.co.uk.uk


testing flat, uniform thickness, arbitrary- shaped sample materials that do not contain any isolated holes. Compared with the linear four-probe method, there is no requirement for the sample shape. During the test, the four probe contact points must be located at the edge of the sample, and the test wiring method also provides constant current at two of the probe points, and measures voltage at the other two points. Eight measurements are taken around the sample and these readings are combined mathematically to determine the average resistivity of the sample. Detailed


test methods can be found in ASTM standard F76.


Summary: IT2800 series high-precision source measure unit can accurately realize the resistivity test of semiconductor sheet resistance, thus providing data basis for the improvement of semiconductor manufacturing process. IT2800 source measure unit (SMU) provides professional testing solutions for the industry due to its high-precision measurement and abundant advantages of probe station fixtures.


www.itechate.com


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80