Figure 3: A conceptual block diagram of an NMR spectrometer where an RF pulse stimulates a sample inside the RF coil to emit an RF signal related to the element and molecular structure of the sample.
Another digitizer option that may be of value in this type of application is the digital pulse generator (DPG). The DPG option enables the digitizer to output up to four rectangular pulses with user-set periods, phases, and widths. The timing resolution of the pulses is based on the sampling clock. For instance, a DPG option installed in an M5i.33xx-x16 digitizer can output up to four pulse streams with a timing resolution of just 3.2 ns. The pulse generator operates in conjunction with the digitizer so that both operate simultaneously. One concern in handling long acquisition records is streaming this data to a host computer for further analysis and archiving.
The M5i digitizers utilize a 16-lane Gen3 PCIe bus capable of transferring data at rates up to 12.8 GB/s. This exceptional speed allows single-channel data acquired at a sampling rate of 6.4 GS/s to be streamed, in a FIFO process, directly to the computer with no data loss. Even faster sampling rates can be streamed without data loss using a new 8-bit transfer mode. The mode supports data streaming at acquisition rates of up to 10 GS/s from a single channel.
This example looked at a single T/R switch operation. Some NMRs use multiple RF coils arranged in phase arrays, and measurements on these systems may require more channels.
This, again, can be addressed using the Star-Hub option.
CONCLUSION
The Spectrum Instrumentation M5i.33xx-x16 digitizer family is well-matched to pulsed RF applications found in NMR, MRI and radar. They offer GHz bandwidths supported by up to 10 GS/s sampling rates with long acquisition record lengths. Optional capabilities include synchronizing multiple channels and even generating programmable pulse outputs.
Spectrum Instrumentation
spectrum-instrumentation.com
Figure 4: Analysis of a T/R switch RF input (upper left grid), gate signal (upper center grid), and the T/R switch output (upper right). 18 March 2024 Instrumentation Monthly
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80