FEATURE GAS SENSING NO EXCUSES, BUMP IT
ATEX, IEC and British HSE standards all call for the bump test and calibration of any portable instrument used for worker health and safety protection in a potentially hazardous environment. That means before every use. Here, Shawcity explains the importance of the often over-looked bump testing of gas detection equipment
Many environmental conditions can degrade sensor response. LEL sensors can be inhibited or poisoned by silicon compounds (some lubricants), sulfur compounds (e.g., H2
S), lead (Pb)
compounds, acids and compounds that generate acids on combustion, like chlorinated hydrocarbons. Electrochemical sensors are affected
oday’s gas detection technology offers better protection than ever before but one often-overlooked issue is the regular bump testing of instruments and, crucially, how many users are unaware of the importance of this simple procedure or unsure how to do it. In simple terms, the usual
T
recommendation is that gas detection instruments are serviced and calibrated at least once a year, depending on the technology and the manufacturer. However, it is less well known that ATEX, IEC and British HSE standards all call for the bump test and calibration of any portable instrument used for worker health and safety protection in a potentially hazardous environment. That means before every use. The bump test is a quick check to ensure the instrument’s sensors and alarms are all functioning correctly before deploying the instrument. Just because a sensor is giving a zero reading, it does not necessarily mean that the sensor is working properly. The analogy one could use is a car which has been serviced and has passed its MOT. Although the recommended annual checks have been done there are numerous reasons why, through day-to-day use, the vehicle might become unsafe to drive before the next service, due to unpredictable wear and tear or damage.
18 JUNE 2018 | INSTRUMENTATION
by high levels of organic vapours and corrosive gases. PID lamps can lose output by: simple ageing; becoming coated when exposed to excessive dust, high-boiling compounds, phosphine, organic amines or silazanes; or by contact with liquids. Sensors will also lose response if the sample gas does not adequately reach it because of leaks in the sample train or contamination of the inlet dust filters. Other elements which can affect performance include heavy knocks or drops, damage, frequent exposure to extreme temperatures, humidity and high particulate levels. All of these factors illustrate why bump testing and calibration checks are necessary to regularly monitor performance and accuracy. The ISEA (International Safety
Equipment Association) and ATEX recommend the following general procedure: daily “Bump Test”; a daily calibration check; and full calibration if either two previous checks fail.
A DAILY “BUMP TEST” A daily bump test should be performed before the instrument is put into use. Gas is applied just long enough to make the sensor(s) alarm, to ensure that the gas is getting to the sensor and that both the sensor and alarm are working.
A DAILY CALIBRATION CHECK A daily calibration check should be performed before the instrument is put into use. Calibration gas is applied to
the sensor to see if it responds within predefined error limits, typically within 10-20 per cent of the expected value, as set by the manufacturer’s recommendation or user’s company policy. In critical cases where a permit to work
is given as a result of a gas test, ATEX recommends that a bump test or calibration check also be conducted at the end of the day’s use to ensure that the instrument was functioning properly throughout the entire monitoring period.
Everyone should have absolute confidence that any instrument on which somebody’s life could
depend should be fully functioning, each and every time it is used."
FULL CALIBRATION Full calibration should be performed if either the bump test or calibration check fail. A full calibration should also be conducted at regular intervals as recommended by the manufacturer, user’s company policy, or regulatory agency. Shawcity’s managing
director, Neil O’Regan, is chair of the BSIF Measurement and
Instrumentation Special Interest Group, a collective
with members representing various manufacturers and
distributors. The group’s current focus is a ‘No Excuses, Bump It’ campaign, with the aim of educating the industry on how to effectively maintain instruments and answer FAQs on bump testing and calibration. O’Regan comments: “Shawcity has been at the heart of gas detection and monitoring in the UK for over forty years and one of the key issues we consistently urge our customers and the wider industry to consider is the vital importance of bump testing and calibrating any gas detection instrument to ensure their accuracy is tested regularly.” O’Regan adds: “Everyone should have
absolute confidence that any instrument on which somebody’s life could depend should be fully functioning, each and every time it is used.”
Shawcity
www.shawcity.co.uk
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54