Environmental Laboratory 17 Table 1. HS-GC-FID operating conditions for GRO determination in water
TRACE 1310 GC Parameters Inlet Module and Mode: Split Ratio:
Septum Purge Mode, Flow (mL/min):
Carrier Gas, Carrier Mode, Pressure (kPa):
Oven Temperature Program Temperature 1 (°C): Hold Time (min): Temperature 2 (°C): Rate (°C/min):
Hold Time 2 (min):
FID Temperature (°C): Air Flow (mL/min): Flow (mL/min): Flow (mL/min):
H2 N2
Acquisition Rate (Hz):
SSL, split 20:1
Constant, 5 He, constant pressure, 150
50 1
220 15 5
300 350 35 40 25
TriPlus 500 HS Autosampler Parameters Incubation Temp. (°C): Incubation Time (min): Vial Shaking:
85 30
Fast Vial Pressurization Mode:
Vial Pressure (kPa) (Auxiliary Gas Nitrogen):
Vial Pressure Equilibration Time (min):
Loop Size (mL):
Pressure 200
1 1
Loop/Sample Path Temp. (°C): 105 Loop Filling Pressure (kPa):
150
Loop Equilibration Time (min): 1 Needle Purge Flow Level: Injection Mode:
5 Injection Time (min):
Standard 1
concentration of a substance that can be measured and reported with 99% confidence that the analyte concentration is greater than zero.2
According to the Wisconsin method3 for
GRO determination, the required limit of quantitation (LOQ) is 100 µg/L (ppb) or less for water samples and the method blank should not exceed a concentration of 50 µg/L (ppb). The method detection limit (MDL) was assessed analyzing n = 7 blank tap water samples (5 mL) and n = 7 tap water samples spiked at the concentration of 12.5 µg/L (ppb). MDL and LOQ were then calculated applying Equations 1 and 2, respectively.
The recovery was calculated using Equation 3 and was in the range 80% to 120%, with an average value of 105%. MDL, LOQ, and percent recovery results for the spiked samples are reported in Table 2. None of the investigated compounds could be detected in the tap water samples as shown in Figure 2.
(Equation 1) MDL = t(n-1,1-α=0.99)
Where:
Figure 2. Comparison between chromatograms obtained analyzing n = 7 real tap water samples (unspiked) and a tap water sample spiked at 12.5 µg/L (ppb) (red trace). None of the investigated gasoline organics could be detected in the unspiked tap water samples.
t = Student’s t-value appropriate for the single-tailed 99th percentile t statistic and a standard deviation estimate with n-1 degrees of freedom, for
n = 7 injections: t = 3.143 S = standard deviation of the replicate analysis
(Equation 2) * S
LOQ = 10 * S Where:
S = standard deviation of the replicate analysis
Table 2. Calculated MDL, LOQ, and % recovery for n = 7 tap water samples spiked at a concentration level of 12.5 µg/L (ppb)
Gasoline Range Organics
Methyl tert-butyl ether (MTBE) Benzene Toluene
Ethylbenzene
m-Xylene, p-Xylene o-Xylene
1,3,5-Trimethylbenzene 1,2,4-Trimethylbenzene Naphthalene Average
Spiked Conc. (µg/L)
12.5 12.5 12.5 12.5 12.5 12.5 12.5 12.5 12.5
Average Measured Conc. (µg/L, n = 7)
11.5 12.8 13.7 12.8 12.8 12.4 14.4 13.3 14.6 13.1
Calculated MDL (µg/L)
1.4 1.2 1.7 1.3 0.8 0.8 1.7 1.7 2.2 1.4
Calculated LOQ (µg/L)
4.4 3.9 5.5 4.0 2.7 2.6 5.5 5.3 7.1 4.6
Average Recovery (%, n = 7)
92
103 110 102 103 100 115 107 117 105
(Equation 3) Average %R = (Cave
Where: Cave
Csp
Figure 1. Calibration curves for GRO obtained by injecting seven concentration levels (6.25 to 10,000 µg/L). R2 relative standard deviations (% RSD) are shown. Each calibration level was prepared and analyzed in triplicate (n = 3).
as well as response factors
/Csp
) * 100%
= average concentration of the spiked samples = initial spike concentration
To assess the method accuracy (%) in tap water samples containing raw gasoline matrix, 30 µL of the sample stock solution (prepared as described in the sample preparation section) were diluted into two flasks previously filled with 30 mL of tap water and fortified with the standard solution at a concentration of 1000 µg/L (ppb) and 10,000 µg/L (ppb). A blank matrix solution was prepared by adding 30 µL of sample stock solution to 30 mL tap water. Then, 5 mL of each fortified solution and the blank matrix were transferred into 10 mL headspace vials (n = 5) and analyzed to assess the recovery. The average recoveries (%) for the spiked matrix samples were calculated using Equation 3 and confirmed to be within 80–120% of the spiked levels with an average value of 96.5% (Table 3). Chromeleon CDS matrix correction feature allowed for automated subtraction of the background leading to a precise quantitation of the spiked samples.
Precision System repeatability was assessed using n = 10 consecutive injections of tap water samples spiked at a concentration of 50 µg/L (ppb) and n = 10 tap water samples spiked with the 5% raw gasoline solution. Peak area %RSDs obtained for both
www.envirotech-online.com AET October / November 2019
MTBE
Benzene
Toluene
Ethylbenzene m-Xylene, p-Xylene o-Xylene
1,3,5-Trimethylbenzene 1,2,4-Trimethylbenzene
Naphthalene
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44