MISTAKES AND PROBLEMS IN GAS SAMPLING AND CONDITIONING
Irrespective of the current political statements and changes in the US the world understands that a reduction of emissions is essential in order to limit the climate change within an acceptable level. The fi rst step to achieve this target is an acceptance and approval of (continuous) emission monitoring systems. Continuous Emission Monitoring Systems (CEMS), from the operators’ point of view, is often a very unloved topic: As the measurements are solely made to comply with legal requirements and local laws it is seen as a cost intensive investment with no direct returns on investment. This differentiates CEM Systems from Process Analytical Technology (PAT) which is business case driven and allows the operators to improve the outcome of the process. Some of the operators might use the abbreviation CEM for “cry every morning”, to sum up the challenges these kind of systems face.
In order to improve the acceptance of CEM systems two major challenges have to be achieved:
1. Low Total Cost of Ownership which includes the initial investment as well as the operating costs
2. Accuracy and long-term stability
The surprising fact is that the sampling and conditioning system is the key component which defi nes success or failure in accomplishing this target, rather than the analyser itself. Tendencies to measure the gas directly in the process via (often optical) insitu technologies are more and more declining in favour of the traditional extractive measurement. The reason for this trend is on one hand the limitation that most insitu analysers only allow operators to measure one component whilst, on the other hand, they do not
offer the fl exibility of an extractive system: In cases where additional
components or lower measurement ranges and limits have to be monitored, the simple change or addition of the analyser component is able to solve the issue.
Nevertheless, sampling and conditioning components in extractive monitoring can even cause problems or falsifi cation of the sample itself. This article is about the identifi cation of potential weak points for the different sampling and conditioning components.
Gas Sampling
When taking a gas sample from a process or stack it is necessary to remove dust and other particles at the same time to ensure a sample temperature above the water vapour resp. acid dew point. If not, condensation will occur, the fi lter will be blocked and water soluble measuring gas components like SO2 H2
, NO2 S or NH3 will be washed out uncontrollably. Two major areas
of the gas sampling probe are extremely crucial: the probe and the probe housing itself as well as the connection from the probe to the heated sample line. As probes are often installed at the stack and often at a very high and windy position, the ambient temperatures can be lower than the ambient temperature at the ground. The heating element as well as effective insulation of the probe body are important in order to keep the temperature level above the condensation point. The accurate homogeneous probe temperature is essential as is the connection to the secondary sampling system, which represents another crucial point: the interface to the heated sample line. If the junction from the probe
AET October / November 2017
www.envirotech-online.com
to the sample line has no additional insulation, cold spots can occur and cause condensational problems such as blocking lines and loss of water soluble gas components. Blocked lines cause a breakdown of the measurement and are very diffi cult to identify and rectify. Loss of gas components will falsify the measurement result. In addition, when trasporting the sample gas to the analyser house it is very important to prevent condensation with every millimeter. Therefore the heating cable, within the heated sample line, has to be positioned up to the very end of the line. Heat capacity and insulation also need to be calculated suffi ciently according to the possible minimum ambient temperatures.
, HCL,
These described factors represent the need for a seamless connection of the Gas Sampling Probe and the heated sample line. AGT-PSG is the only company which manufactures all components from one source: With the biggest fi lter surface in the PSG Basic and PSG Plus Probes and the widest range of heated sample lines, the German supplier offers package solution which integrate seamlessly.
Gas Conditioning
When sample gas has passed through the heated probe and line – the most crucial process step follows: sample gas drying by cooler. The reason for this step is to protect the analyser against condensate and minimisation of water vapour affecting performance. Condensate will damage the sensitive and expensive analyser and water vapour cross sensitivities create inaccurate
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36