search.noResults

search.searching

note.createNoteMessage

search.noResults

search.searching

orderForm.title

orderForm.productCode
orderForm.description
orderForm.quantity
orderForm.itemPrice
orderForm.price
orderForm.totalPrice
orderForm.deliveryDetails.billingAddress
orderForm.deliveryDetails.deliveryAddress
orderForm.noItems
20 February / March 2016


which can clog fuel systems. Watercol 1460 was evaluated for the suitability to perform the analysis of water in gasoline. In addition to producing a sharp water peak, its selectivity results in elution of the water peak free from most other constituents present in gasoline. Plus, its maximum temperature of 260°C allows for timely elution of the heaviest gasoline components.


Figure 4: Water Calibration Curve (0.05-1%) on WatercolTM Conditions


column: WatercolTM 1460 1460, 30 m x 0.25 mm I.D., 0.20 µm (29695-U);


two identical columns installed in the same inlet using a 2 hole ferrule, each going to a separate detector.


oven temp: 45°C (4 min), 5°C/min to 75°C, 10°C/min to 220°C (10 min) inj. temp.: 250°C detector: FID, 250°C detector: TCD, 200°C carrier gas: helium, 20 cm/sec (measured in each column at an oven temperature of 135°C) injection: 1 µL, 50:1 split


liner: 4 mm I.D., split/splitless type, wool packed single taper FocusLiner™ design sample: 5 standards, water at 0.05, 0.1, 0.25, 0.5, and 1% (v/v), each in ethanol


Two identical columns were installed in a single inlet using a 2-hole ferrule. Each column was connected to a separate detector to permit the simultaneous detection of gasoline components as well as water. Triplicate samples of unleaded gasoline were analysed, both “as is” and spiked with water at 0.25% (v/v). Figure 5 shows representative chromatograms. Reproducibility was very good (RSD <5%) for the spiked gasoline replicates. The calibration curve shown in Figure 4 was used for quantitation. The average response factor calculated from each point of the curve had an RSD value of <20%, thus it was used to calculate the amount of water in the gasoline samples. The average concentration found in the unspiked gasoline replicates was subsequently subtracted from the average concentration measured in the spiked replicates. The blank corrected method accuracy was determined to be 79% at the spike level used.


Aroma Fusel Alcohols in Tequila


Figure 5: Spiked Gasoline Conditions


sample: unleaded gasoline, spiked with water at 0.25% (v/v) All other condition the same as Figure 4


Fusel alcohols are by-products of the fermentation process. Four of these (n-propanol, isobutanol, active amyl alcohol, and isoamyl alcohol) are important to alcoholic beverage manufacturers due to their contributions to the aroma characteristics. A tequila sample was analysed for the presence of aroma fusel alcohols using an FID and TCD detectors. The FID yielded greater responses for the alcohols, whereas the TCD produced a response for water. As shown in Figure 6, all four aroma fusel alcohols were detected in the tequila sample. Good peak shapes and resolution were obtained, allowing for distinct detection of these compounds. In addition, the selectivity of the phase resulted in elution of both ethanol and water free from all other analytes.


Conclusions


Ionic liquid GC phases based on new selective combinations of cations and


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60