This page contains a Flash digital edition of a book.
DETERMINATION OF SULFUR COMPOUNDS continued


Table 1 – Instrument configuration and operating conditions S-Pro Select GC


Permeation oven 35 °C Helium dilution gas


Permeation devices


Dilution gas flow rate = 200–3600 mL/min H2


S wafer device—permeation rate = 809 ng/ min at 40 °C


COS wafer device—permeation rate = 1586 ng/min at 40 °C


MeSH wafer device—permeation rate = 441 ng/min at 40 °C


Automated injection system


Four-port selection valve


Six-port gas sampling valve (GSV) with 1-mL, Sulfinert-coated sample loop Automated, air-actuated valves All lines Sulfinert coated Valve oven temperature = 110 °C


Volatiles interface 200 °C Split mode


Split ratio = 10:1 Sulfinert coated


GC column


Agilent J&W (Santa Clara, Calif.) Select Low Sulfur column 60 m × 0.32 mm i.d.


Helium carrier gas = 1.2 mL/min Oven program


40 °C for 10 min 30 °C/min to 180 °C Hold for 10.5 min* Total run time = 25.2 min


Sulfur detection PFPD


2-mm combustor, BG-12 filter, R1924 PMT (photomultiplier tube)


/air ratio tuned for optimum sulfur emission 6–24 msec sulfur gate (linear mode) 1–3 msec hydrocarbon gate Square root on


Detector base temperature = 250 °C H2


*Oven hold time may be shortened depending on the target com- pounds.


The difficult separation of H2S from ethylene/ethane was achieved using


the Select Low Sulfur column. An H2


S peak typically elutes between the ethylene and ethane peaks for


many matrix concentration levels found in gaseous fuels. Some coelu- tion of H2


S may occur as the matrix concentrations increase above 50%


ethylene. COS was well separated from the propylene peak. These separa- tions are important because they demonstrate that the column is flexible


AMERICAN LABORATORY 22


Figure 3 – Natural gas sample 2 (obtained from DCG Partnership, Pearland, Texas).


enough to be used in several matrices, with no need to make significant changes to the chromatographic conditions. Ultimately, the matrix con- centration ranges can be balanced with the required sulfur sensitivity and GC split ratios to optimize performance for various types of analyses. Calibration was performed with the FlexStream gas standards generator (Figure 4). Linearity was established with a correlation coefficient of 0.999.


Repeatability study A repeatability study was performed by analyzing 20 replicate injections


of the same sulfur standards used for the calibration, with a flow rate of 1000 mL/min through the permeation oven.


Concentrations and %RSD were as follows: • H2


S at 0.376 ppmv—%RSD of 3.88%


• COS at 0.418 ppmv—%RSD of 3.34% • MeSH of 0.145 ppmv—%RSD of 5.84%


MARCH 2016


Figure 2 – Natural gas sample 1 (obtained from DCG Partnership, Pearland, Texas).


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60