Stand-alone closed circulators providing powerful, accurately controlled cooling for analytical techniques and instruments where an optimum operating temperature is required. In addition, RC350G, RC400G and RC1400G can be used as heated circulators.
u Precise temperature control u Low coolant consumption*
u Settable high and low alarm levels with lamp and buzzer
u Over/under temperature cut-outs
u Flow fail device cuts power if no liquid in system
u Digital temperature display
*Use: Water for range +5 to +60°C Water/glycol mixture for –10 to +5°C Recirculating Chillers
As specified. Mounted on lockable wheels. For 220- 240V 50Hz single phase supplies
CL110-05 Model RC350G CL110-15 Model RC400G CL110-25 Model RC1400G CL110-30 Model RC3000G
Catalogue No Model
Temperature range Heater power
Stability at 20°C (DIN58966) (using water)
Pump head pressure at 1 litre/minute
Dimensions D Weight
Inlet/outlet connections Reservoir capacity
Operational ambient temperature range
EMC emissions Supply requirements
CL110-05 CL110-15 CL110-25 CL110-30 RC3000G†
RC350G
Typical cooling power at 20°C W 350 kW 0.75
±°C bar Maximum liquid flowrate litres/minute 15 1.60 mm 600
Wmm 370 Hmm 510
kg 42 litres RC400G RC1400G
°C –5 to +60 –10 to +60 –10 to +60 –10 to +60 400 0.75
1300 1.50
0.25*
0.25 12
0.62 600
370 510
42
mm 9.5 diameter — all models 1.7
1.7 °C +5 to +35 — all models
class BBA B 220-240V 50Hz single phase supplies — all models
†Note: The RC3000G has no heater and therefore is designed for cooling applications only. It can control to +60°C where the temperature of the exothermic reaction or process is above +60°C; this is achieved by switching the cooling on and off.
*With 10 litres of water in the system. **With 25 litres of water in the system.
Accessories Bypass RC BYP
Ensures that the flow through the chiller is always at least 1 litre/minute so that the chillers’ flow-fail device does not engage. This maintains temperature control and system integrity if narrow tubing or small cooling cells are used in an external circuit.
CL110-85 RC BYP Pressure gauge RC PR
Indicates output pressure from the chiller.
CL110-89 RC PR CL110-93 PRES Priming reservoir
0.25 15
0.62 630
380 590
53 2.5
3000 — †
0.5** 15
1.60 840
490 640
88 1.1
CL110-05, CL110-15 to CL110-30 are similar & 0808 200 5227 e-mail: sales@thelabwarehouse.com fax orders: 020 7729 2657