Materials


materials, manufacturing processes, sterilisation, packaging, patient contact and clinical performance. “A material that looks ideal from an engineering perspective may introduce significant chemical characterisation, toxicological or clinical evidence requirements,” he adds. “If those are not identified until verification or submission, the manufacturer may face additional testing, redesign or significant delays. The modern approach is to bring regulatory, biological safety and quality considerations into development much earlier.”


This instinctively sounds highly pragmatic. Biocompatibility, extractables and leachables, long- term safety evidence and other inputs affect how quickly a device can move through development. Bad choices can create expensive consequences down the line, so companies no longer select materials primarily for performance but also for their regulatory implications.


Materials strategy is moving earlier in device development rather than remaining a downstream engineering task. However, as companies choose materials with established regulatory and clinical histories rather than those that may be technically optimal, this shift may have downsides.


Shifts in the regulatory landscape Developers must deal with the changing reality of the approval process. In the US, the FDA has issued guidance and draft guidance containing more detailed recommendations on chemical characterisation as one component of biocompatibility assessment. These developments sit alongside the agency’s partial recognition of ISO 10993-1:2025 in May 2026.


This can be viewed as a shift away from asking only what material a device is made from and whether it passes biocompatibility tests, towards analysing which chemicals could reach the patient from the finished device and at what levels, how reliably they have been identified and what exposure could mean toxicologically. The FDA’s September 2023 final guidance on ISO 10993-1 emphasises evaluating the finished device, including its materials, manufacturing processes, sterilisation and relevant residues. In May 2026, the agency also partially recognised ISO 10993-1:2025, the standard’s sixth edition.


The FDA does not clear or approve individual materials. Its assessment concerns the finished device and therefore considers the material, manufacturing processes, sterilisation and residuals from manufacturing aids. Its September 2024 draft guidance, Chemical Analysis for Biocompatibility Assessment of Medical Devices, sets out recommended approaches for conducting and reporting chemical analysis studies. As draft guidance, it represents the FDA’s current thinking rather than a binding requirement.


www.medicaldevice-developments.com


Biocompatibility assessments increasingly use chemical characterisation and toxicological risk assessment alongside, or to inform the need for, biological endpoint testing. Historically, a manufacturer might have approached biocompatibility by selecting tests associated with the device’s contact type and duration, but ISO 10993-1 frames biological evaluation as a risk- management process rather than a fixed checklist. In Europe, the EU Medical Device Regulation requires manufacturers to maintain risk management, technical documentation, clinical evaluation and post-market surveillance processes throughout the device life cycle, including relevant information about materials and biological risks. Under MDR Annex I, manufacturers must establish, implement, document and maintain a risk-management system that is systematically updated, so the choice of polymer, alloy, adhesive or coating does not disappear from the design history once the product launches.


“Developers and entrepreneurs better understand the regulatory system and plan in advance for what is needed to get approval.”


Professor Buddy Ratner


Furthermore, Annex II requires the technical documentation to include relevant verification and validation information, including device biocompatibility and physical, chemical and microbiological characterisation. The regulation requires evidence demonstrating that the finished device conforms to applicable safety and performance requirements, including relevant evidence on biocompatibility and material characterisation.


Entering the age of ‘design for approval’


Material decisions are moving beyond which polymer or metal will perform best. Instead, materials selection has become a regulatory and market-access decision.


“Everyone is considering it earlier, as there is a better understanding of what the US FDA and other regulators require,” says Buddy Ratner, professor of bioengineering and chemical engineering at the University of Washington. “The FDA has not necessarily intensified requirements, but developers and entrepreneurs better understand the regulatory system and plan in advance for what is needed to get approval.”


There is a shift towards regulatory affairs teams entering the development process earlier, materials information being considered before concepts are finalised and greater preference for materials with established clinical histories.


69


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68  |  Page 69  |  Page 70  |  Page 71  |  Page 72  |  Page 73  |  Page 74  |  Page 75  |  Page 76  |  Page 77  |  Page 78  |  Page 79  |  Page 80  |  Page 81  |  Page 82  |  Page 83  |  Page 84  |  Page 85  |  Page 86  |  Page 87  |  Page 88  |  Page 89  |  Page 90  |  Page 91  |  Page 92  |  Page 93  |  Page 94  |  Page 95  |  Page 96  |  Page 97  |  Page 98  |  Page 99  |  Page 100  |  Page 101  |  Page 102