Scanning Probe Microscopy in an Ultra-Low Vibration Closed-Cycle Cryostat: Skyrmion Lattice Detection and Tuning Fork Implementation


Francesca Paola Quacquarelli , 1 † * Jorge Puebla , 1 †† Thomas Scheler , 1 Dieter Andres , 1 Christoph Bödefeld , 1 Balázs Sipos , 1 Claudio Dal Savio , 1 Andreas Bauer , 2 Christian Pfl eiderer , 2


Andreas Erb , 3 and Khaled Karrai 1 1 attocube systems AG , Königinstraße 11a , 80539 München , Germany 2 Physik-Department , Technische Universität München , James-Franck-Str. 1 , 85748 Garching , Germany 3 Walther Meissner Institute , Bavarian Academy of Sciences and Humanities , 85748 Garching , Germany


* francescapaola.quacquarelli@gmail.com Introduction


Many-body phenomena such as recently imaged currents in topological insulators [ 1 ] or nanoscale spin textures in chiral magnets, so-called magnetic skyrmions [ 2 ], arise at cryogenic temperatures constraining scanning probe microscopy (SPM) to be low-temperature compatible and to operate in liquid helium based cryostats [ 3 , 4 ]. However, the high costs of helium and its scarcity have become a limiting factor, recently propelling the development of closed-cycle cryostats [ 5 , 6 ]. T ese are independent of a continuous supply of liquid helium but present a challenge to the implementation of SPM owing to their noisy mechanics. Tuning fork-based scanning gate microscopy in a pulse-tube-cooled dilution refrigerator has been recently reported [ 7 ], opening the way to SPM in dry cryostats. T is pioneering achievement required heavy intervention on the cryo-cooling system, sacrifi cing thermal contact and requiring long pre-cooling times. In this article we report on the development of an ultra-low vibration closed-cycle cryostat that allows top-loading of a variety of scanning probe microscopes with rapid turnovers. Because the microscopes do not require spring-mounting to further reduce vibrations, high-resolution confocal light microscopy imaging can be combined with scanning force microscopy. T e ultra-low amplitude of vertical vibrations measured in our system enabled the AFM resolution of 0.39 nm high atomic steps in a SrTiO 3 sample. T is low level of vibration also allowed detection of the skyrmion-lattice phase in a chiral magnet. Furthermore, it opened the way to the implementation for the fi rst time of tuning fork-based shear-force microscopy in a closed-cycle cryostat. Scanning tunneling microscopy measurements recently reported by Leiden Cryogenics BV complete the picture of SPM measurements in a closed-cycle cryostat [ 8 ]. T is article describes the cryo-cooled AFM and gives some example images.


Materials and Methods Cryo-cooling system . Our dry cryo-cooling system


( Figure 1 ) relies on conductive cooling and is based on pulse tube technology [ 6 ]. Here, a cryogenic fl uid at high pressure is pulsed longitudinally through the diff erent stages of the tube


† Now at Institut für Schlaganfall-und Demenzforschung (ISD), Klinikum der Universität München, Feodor-Lynen-Straße 17, D-81377 München, Germany †† Now at Center for Emergent Matter Science, RIKEN, 2-1 Hirosawa, Wako 351-0198, Japan


12


Figure 1 : Closed-cycle cryo-cooling system based on pulse tube technology. The system consists of three main parts: a pulse tube, a rotary valve, and a compressor. A 9 T superconducting magnet, fi xed to the low-temperature stage of the pulse tube, surrounds coaxially the sample space, and it is cooled by thermal contact. A vibration detector in its vacuum insert is top-loaded into the system to characterize vibrations in the sample space.


doi: 10.1017/S1551929515000954 www.microscopy-today.com • 2015 November


at a frequency of 1.4 Hz, reaching temperatures of approxi- mately 3 K. T e cyclic expansion and compression of the gas in the adiabatic chamber where the fl uid is stored produces a stratifi cation of temperature along the tube, so that one end is warmer than the other. T e Cryomech PT-410 2-stage pulse tube cryo-cooler is driven by a compressor, which produces 1 W of cooling power at 4.2 K. A rotary valve regulates the pressure of the gas through the pulse tube. A 9 T supercon- ducting magnet completes the system and is cooled by the pulse tube through thermal contact. Initial cooling of the whole system requires 12 hours; cooling down of the microscope insert from T=300 K to base temperature, namely 4 K, is normally achieved within 1 hour, depending on the mass of the microscope. One hour is also required to warm the system up. T e sample space is immediately accessible aſt er warming up thanks to the top-loading architecture. T e system confi g- uration enables measurement turnovers every 3 hours: sample exchange times of typically a few minutes are included in this estimate. T e SPM experiments and the characterization of the vibrations in the system take place in the 50 mm diameter sample space, consisting of air-tight, removable, thin-walled


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56  |  Page 57  |  Page 58  |  Page 59  |  Page 60  |  Page 61  |  Page 62  |  Page 63  |  Page 64  |  Page 65  |  Page 66  |  Page 67  |  Page 68