EMC & Circuit Protection
used measures 2.4 metres by 1.5 metres and includes eyelets for vertical mounting. Using two of these panels allowed us to quickly assemble a reliable ground plane system.
Figure 2: Recommended grounding of LISN, the green and yellow earthing wire ensures safety, the bottom plate of the LISN is directly connected to the test ground for RF measurement
Test setup overview
This article explains how to establish a conducted emissions test setup using a bench-top configuration. The goal is to enable engineers to validate product designs before sending them to an accredited lab. The focus here is on commercial mains-powered products. Similar methods apply to industrial products, while military and defence applications require different setups. We follow the CISPR 32 recommended setup for tabletop devices, and set up the test, as shown in Figure 1. Key points include: 1. The use of both horizontal and vertical test ground planes. 2. A table height of exactly 80 cm. The table must be non-conductive. Polyform is ideal; wood is acceptable. 3. The Device Under Test (DUT) must be positioned 40 cm from the vertical ground plane. The mains cable must be exactly 1 metre in length, with 80 cm hanging freely in the air. Any excess cable should
be bundled and supported by a non- conductive material.
We recommend using galvanized steel sheets for the ground planes due to their affordability and good conductivity. However, this approach presents a few challenges: 1. It typically requires multiple large sheets to construct both planes. 2. The sheet edges are sharp, so appropriate safety measures such as gloves are needed. 3. Electrical continuity between overlapping sheets must be ensured using fastening points.
Aluminium is another
option, but its anodized surface can interfere with conductivity between sheets, making it less ideal.
To resolve these issues, we used a roll-up ground plane from Tekbox. The model we
3. It functions as a voltage measurement device. When connected to a spectrum analyser (with 50 ohm input impedance), the results can be read directly in dBμV, which is the standard unit for conducted emissions. The LISN must be bonded to the horizontal test ground plane to establish a consistent EMC measurement reference ground. It must also be earthed to ensure operator safety. This can be seen in Figure 2. In addition, an isolation transformer is recommended when powering the LISN from the building’s mains supply. The LISN tends to draw a significant leakage current, which can trip the RCD (Residual Current Device) in typical building installations.
Figure 3: An EMI receiver is used for the demonstration. A transient limiter is visibly connected to the RF input
www.cieonline.co.uk
Measurement equipment The output of the LISN is connected to an EMI receiver or spectrum analyser. For this demonstration, we used the Tekbox TBMR-110M EMI receiver. Alternative spectrum analysers from Rigol or Siglent are also viable, but to ensure accurate
published by the manufacturer. The profiles matched closely, with slightly higher emission levels recorded in our setup. This is expected, given that filter tolerances in mass-produced units can vary by ±10 per cent. The outcome confirms that our in- house setup delivers results comparable to those from an accredited lab.
Video demonstration
A full video demonstration of the test setup and results is available here:
https://youtu.be/ epEG51hy1CM?si=ma3pxCQd8tuN_gkR
Acknowledgments
The author would like to thank Telonic Instruments (
https://telonic.co.uk ) for providing the test equipment used in this article.
Mach One Design Ltd is a UK-based engineering firm that specializes in EMC consulting, troubleshooting, and training.
http://www.mach1design.co.uk Components in Electronics September 2025 15
LISN and power considerations For mains-powered devices, a 50 microhenry LISN (Line Impedance Stabilization Network) is required. The LISN serves several critical purposes in conducted emissions testing: 1. It provides a defined impedance – typically 50 ohms – within the relevant frequency range for EMC conducted emissions testing, which is from 9 kHz to 30 MHz in this case. This standardization ensures that the test conditions are consistent and comparable. 2. It offers noise decoupling between the DUT and the rest of the mains supply network. This helps ensure that the measured noise originates from the DUT itself and not from other appliances operating in the same building.
and repeatable results, they should be used in conjunction with dedicated EMC software such as EMCview. Otherwise, measurement inaccuracies may occur due to the limitations of basic sweep modes in lower-cost analysers.
A transient limiter must be placed at the RF input of the analyser or receiver to protect against spurious low-frequency transients on the mains supply. Without this protection, the receiver input can be damaged – often without the user realising – leading to inaccurate results. A transient limiter from Tekbox is shown in Figure 3.
DUT and reference results For the test, we used a switched-mode power supply (SMPS) evaluation board from Monolithic Power Systems (MPS). This board was chosen because the manufacturer provides conducted emissions test data, allowing us to cross-validate our results. Using the Tekbox EMI receiver, we compared our measurements to those
Figure 4: Comparison between the manufacturer’s datasheet results and our own conducted emissions measurements
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60