USER GUIDE
Matters needing attention when the K-kit is in use
With Newton’s rings
SPECIFICATIONS K-Kit Carrier
K-Kit Die Grid Step Cap Seal Sealed by channel tips
Inspection Before Use With Newton’s rings on the membrane. (Be sure to keep the channel vacuum sealed)
Silicon body of K-kit should be damage-free. K-Kit Stand K-Kit
Each carrier has a K-kit attached on top, protected with a clear cap. (A copper grid is enclosed at the bottom of the carrier.)
Flat membrane
Base
Dimensions Window Length: Open to atmosphere Channel Tips Removal
Be sure to remove both the channel tips before using K-kit.
It should be finished the liquid loading within 0.5 hour, after breaking the channel tips.
With color patterns (After liquid filled)
300µm, Width 25µm
Channel Height (H): 0.2 and 2.0 standard 0.5, 1.0 and 5.0 available
H = 0.2µm, 0.5µm, 1µm, 2µm, 5µm
Liquid Loading
Keep the K-kit steadily touching on liquid for around 1 minute, to allow the filling to complete.
Do not immerse the K-kit in liquid.
Liquid well reserved (glue the openings soon)
Gluing Process
Glue both ends of the channel within 1 minute after liquid loading.
Be sure not to do the channel gluing step, if making for Thin Layer mode of K-kit.
Take care during the gluing step, to avoid the glue flowing into the observation window.
12