This page contains a Flash digital edition of a book.
AL Results for large-molecule HPLC


Large-molecule RP-HPLC separations and loss of retention and efficiency in the hindered diffusion region are referred to as the exclusion effect. Serious performance problems usually result when solute size exceeds about one-tenth of the mean pore diameter.


The 160-Å C4 SPP column in Figure 4 is recommended for solutes up to about 15,000 D, so it would not be expected to perform well for a monoclonal antibody (150 kDa). The peak is too broad and too close to the exclusion limit of this 160-Å column. A calibration plot with a 200-Å SEC column would likely predict this result; this was confirmed by doing the same separation on a 400-Å C4 column. Note that peak width has become much narrower with the larger-pore column, which shows that pore crowding has been greatly reduced for the mAb, allowing better mass transfer and a much sharper peak. Although surface area is much smaller for the larger-pore column, retention is comparable, showing there is much more room in the larger pores for phase access.


Figure 5 shows a monoclonal antibody separation on a 1000-Å SPP C4 column and a 300-Å FPP C4 column. The larger-pore SPP particle exhibits greater retention and higher efficiency for the mAb. This is ex- pected for the 1000-Å particle, because the radius-ratio would be about 0.1 compared to only about 0.3 for the 300-Å particle, and peak broad- ening arises from slower mass transfer, described earlier by Giddings.


Figure 4 – Example of monoclonal antibody that is too large for a 160-Å-pore column, showing peak broadening corrected by changing to a 400-Å-pore column, which greatly improves efficiency and allows important peaks to be detected. Conditions—columns: 2.1 × 150 mm; mobile phase A: water/0.1% difluoroacetic acid; mobile phase B: ACN (acetonitrile)/0.1% difluoroacetic acid; gradient: 27–37% B in 20 min; flow rate: 0.4 mL/min; temperature: 80 °C; injection volume: 2 µL of 0.5 mg/mL Sigma MAb; instrument: Nexera; detection: PDA, 280 nm.


Although calibration curve results are not shown for 300-Å and 1000-Å SEC columns, mAbs should lie in the low-performance region for a 300-Å SEC column and in the high-performance region for a 1000-Å SEC column (see Figure 2).


UNLOCK THE ANSWERS WITH CROSS-PLATFORM ANALYSIS


6(0 %6( LPDJH $FFX72) '$57 +LJK 5HV 0DVV 6SHF $QDO\VLV ZLWK +506 0DVV 6SHF


Questions about… • Morphology?


• Chemical Composition? • Structure? • Or all three?


Fast, accurate imaging and analysis of samples using JEOL Scanning Electron Microscopy, ambient ionization Mass Spectrometry, and NMR Spectroscopy.


Solutions for Inno ation ('6 VSHFWUDO RYHUOD\ & &D 3 PDSV ,7 ,Q7RXFK6FRSH 6(0 NMR analysis


Contact us at salesinfo@jeol.com to arrange a demo 8QNQRZQ WDEOHW DQDO\]HG DQG LGHQWLÀHG DV FKURPLXP SLFROLQDWH


6(0 • 7(0 • 0DVV 6SHF • 105 • (65 • (30$ • $XJHU • ),% • 6DPSOH 3UHS • $SSOLFDWLRQV ([SHUWLVH ,PDJLQJ


e l


K $Qalysis =


<2.2*86+, MHROXVD FRP \RNRJXVKL


AMERICAN LABORATORY 25 JUNE/JULY 2017


Solutions for InnovaStion


W


e


K


i


a


K


t


c


c


a


a


u o


n


y


e w


v


e


?


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52  |  Page 53  |  Page 54  |  Page 55  |  Page 56