CHROMATOGRAPHY 43
BATTLING AIR BUBBLES
Degassing in HPLC and ion chromatography G
as bubbles are produced in a liquid when there is more available gas than
the liquid can actually absorb. Te amount of gas that a liquid can absorb depends on several factors, such as the pressure and temperature gradients, the nature and type of liquid and the gas. Air bubbles in particular are the enemy of practical HPLC and ion chromatography. Tey can cause shifting retention times, pulsating baselines, unexpected changes in peak areas, irregular peak shapes, pump outages and excessively high back pressures.
Tis means that analytical scientists may be unable to distinguish between trace amounts of analytes and the baseline, therefore inaccuracy and poor precision (especially at low signal levels) can occur. It also prevents the determination of analytes that are close to their limits of detection. In general, air includes 78%
nitrogen and 21% oxygen. Te molecular presence of oxygen can interfere with the detection of some analytes, particularly when electrochemical and fluorescence detection are used. Within HPLC/ion
chromatography systems, air bubbles are normally encountered when the air of mobile phase/s has not been removed, when low pressure mixing gradients are formed, when mobile phase filters become clogged with buffer
A typical Cecil Instruments Q-Adept HPLC system
salts, microbes, dust, etc, and occasionally if the column interface with detector does not have a sufficiently high back pressure. Consequently, to address the first two aforementioned circumstances, analytical scientists need to remove the air (degas) from mobile phases. Mobile phases need to be free of air before they reach the pumps. Several degassing techniques are commonly used. Tey generally fall into two camps: offline or inline/online.
Offline degassing is where the mobile phases are degassed before placement within the main body of the HPLC/ion chromatography system. With inline/online degassing, mobile
phases are continually degassed within the main body of the HPLC/ion chromatography system.
Offline degassing has the major disadvantages that the mobile phases begin to reabsorb air when they are placed within the main body of the HPLC/ion chromatography system and they are time consuming.
Offline degassing techniques include:-
l Helium sparging. Unlike nitrogen, helium is barely soluble in water. So helium gas is bubbled through the
mobile phases. Tis is actually the most effective degassing technique, but it is off-line and there is a current shortage of helium.
l Sonication, by placing containers of mobile phases into ultrasonic water baths and sonicating until the air is deemed to have escaped the containers.
l Filtration under vacuum. l Warming and shaking,
which brings the temperature of cold mobile phase to ambient temperatures.
Inline/online degassing
has the very major advantages of being instant, continuous and time saving. Inline/online degassing involves the placement of a vacuum degassing device, within the body of the HPLC/ ion chromatography system, just before the mobile phases enter pumps. Te vacuum degassing device contains a gas permeable membrane, through which nitrogen and oxygen molecules are vacuum extracted from the mobile phase. Hence the mobile phases are degassed as and when required. Ironically, inline/online
vacuum degassers are theoretically not as effective as helium sparging, but in practice, they are adequate and highly convenient.
For more information visit
www.cecilinstruments.com www.scientistlive.com
Page 1 |
Page 2 |
Page 3 |
Page 4 |
Page 5 |
Page 6 |
Page 7 |
Page 8 |
Page 9 |
Page 10 |
Page 11 |
Page 12 |
Page 13 |
Page 14 |
Page 15 |
Page 16 |
Page 17 |
Page 18 |
Page 19 |
Page 20 |
Page 21 |
Page 22 |
Page 23 |
Page 24 |
Page 25 |
Page 26 |
Page 27 |
Page 28 |
Page 29 |
Page 30 |
Page 31 |
Page 32 |
Page 33 |
Page 34 |
Page 35 |
Page 36 |
Page 37 |
Page 38 |
Page 39 |
Page 40 |
Page 41 |
Page 42 |
Page 43 |
Page 44 |
Page 45 |
Page 46 |
Page 47 |
Page 48 |
Page 49 |
Page 50 |
Page 51 |
Page 52 |
Page 53 |
Page 54 |
Page 55 |
Page 56 |
Page 57 |
Page 58 |
Page 59 |
Page 60 |
Page 61 |
Page 62 |
Page 63 |
Page 64 |
Page 65 |
Page 66 |
Page 67 |
Page 68 |
Page 69 |
Page 70 |
Page 71 |
Page 72 |
Page 73 |
Page 74 |
Page 75 |
Page 76 |
Page 77 |
Page 78 |
Page 79 |
Page 80 |
Page 81 |
Page 82 |
Page 83 |
Page 84 |
Page 85 |
Page 86 |
Page 87 |
Page 88 |
Page 89 |
Page 90 |
Page 91 |
Page 92