mainstream cigarette smoke. It recommends establishing levels for selected mainstream smoke toxicants per milligram of nicotine and prohibiting the sale or import of cigarette brands that have yields above these levels. Te initial regulatory levels recommended for nitrosamines are 0.072µg/mg for 4-(N-nitrosomethylamino)- 1-(3-pyridyl)-1-butanone (NNK) and 0.114µg/mg for N-nitrosonornicotine (NNN).


Detection, measurement and analysis of VNAs have traditionally been accomplished using gas chromatography (GC) and thermal energy analyser (TEA) detection. Tis combined technique allows the detection of VNAs, but poses two key challenges. Te first challenge is that it is not capable of achieving the low detection limits required by current legislation. In addition, once the identification of VNAs has been completed, GC-TEA does not allow further analysis of the tobacco samples for other contaminants.


Te use of GC is preferable due to the significant amount of chromatography information that becomes available when using this technique. Recent technological advancements have demonstrated that by coupling a gas chromatograph to a triple quadrupole mass spectrometer, the effective separation of VNAs can be achieved, while decreasing detection limits, increasing specificity and allowing for the analysis of many other organic contaminants including pesticides. Te technique can be used in conjunction with timed selected reaction monitoring (t-SRM), thereby allowing the operator of the system to easily set up methods and run samples while the instrument automatically determines the optimal time for SRM parameters, even with partial overlap of SRM transitions. Tis innovative technology delivers significantly lower detection limits and limits of quantitation than conventional GC-TEA.


A simple three-step sample www.scientistlive.com


was set at 210°C. t-SRM provided increased precision and sensitivity by only performing SRM transitions near peak elution. Tis reduced the amount of transition overlap, which in turn increased dwell time and thus sensitivity and precision.


Te matrix calibration curves ranged from 2-50ng/mL and 1-25µg/ kg were extracted from matrix (Figs 1-6). All components exhibited a correlation coefficient of better than 0.995 r2 value. Te per cent variance for the points of the matrix calibration curves was less than 25 per cent. Initially, a minimum detection limit was found to be 1ng/ mL, with a1 µL injection volume. A known uncontaminated sample was spiked and run at that level. Te results indicate that a detection limit lower than 1ng/mL, or 0.5µg/kg in sample, could be achieved (Fig. 7).


Figs. 7. Injection of 1 ng/mL VNA in an uncontaminated clean tobacco sample, demonstrating signal-to-noise ratio.


preparation procedure was followed. A 2g sample was initially placed in 0.01 N KOH solution, which was then shaken for 30 minutes followed by centrifugation. A 10mL aliquot of the sample was subsequently extracted using solid phase extraction, going from an aqueous sample to methylene chloride. Te final extract was then concentrated to 1mL. An autosampler (TriPlus, Termo Fisher Scientific) was used to provide the 1µL injection. Te system’s syringe was held in the injector for four seconds before and three seconds after the plunger was depressed.


GC analysis was performed on a gas chromatograph (TRACE GC Ultra, Termo Fisher Scientific) which used a two-step oven ramp. Te oven program started at 45°C for three minutes, with a first ramp at 25°C/min to 130°C and a second


ramp at 12°C/min to 250°C, with a final hold of two minutes. A medium polarity column was used to provide better separations of the nitrosamines. Te splitless injector temperature was set at 140°C and the surge pressure was set at 300kPa for two minutes. Te column flow rate was held constant at 1.5mL/min.


A triple quadrupole mass spectrometer (TSQ Quantum XLS, Termo Fisher Scientific) was optimised for positive chemical ionization (PCI) using methane as the reagent gas. Te optimum methane flow rate was 2.5mL/ min and the source temperature


Reference: 1


Conclusion In order to protect public health and adhere to the strict legislative requirements, the tobacco industry needs a competent method that can facilitate accurate and reliable detection, measurement and analysis of low levels of VNAs in tobacco. Conventional GC-TEA technology has been used for many years for the measurement of VNAs, however this method is specific only to VNAs and does not achieve the low detection limits required by regulations. Triple quadrupole GC-MS/MS, used together with t-SRM, offers an alternative powerful solution for dependable, low-level analysis of VNAs and other contaminants in tobacco. Te technique is highly sensitive and generates calibration curves that exhibit good linearity.


Eric Phillips, Thermo Fisher Scientific,


Austin, TX, USA. www.thermofisher.com. Mary Dennis, Lancaster Laboratories, Inc., Lancaster, PA, USA.


http://lancasterlabs.com WHO Technical Report Series, Te Scientific Basis of


Tobacco Product Regulation, Second Report of a WHO Study Group, www.who.int/tobacco/global_interaction/tobreg/ publications/9789241209519.pdf


Page 1  |  Page 2  |  Page 3  |  Page 4  |  Page 5  |  Page 6  |  Page 7  |  Page 8  |  Page 9  |  Page 10  |  Page 11  |  Page 12  |  Page 13  |  Page 14  |  Page 15  |  Page 16  |  Page 17  |  Page 18  |  Page 19  |  Page 20  |  Page 21  |  Page 22  |  Page 23  |  Page 24  |  Page 25  |  Page 26  |  Page 27  |  Page 28  |  Page 29  |  Page 30  |  Page 31  |  Page 32  |  Page 33  |  Page 34  |  Page 35  |  Page 36  |  Page 37  |  Page 38  |  Page 39  |  Page 40  |  Page 41  |  Page 42  |  Page 43  |  Page 44  |  Page 45  |  Page 46  |  Page 47  |  Page 48  |  Page 49  |  Page 50  |  Page 51  |  Page 52